ASAHI SION blueRecalls & reported problems across all models
Asahi Intecc CO., Ltd. · 5 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 53
- Deaths reported
- 0
- Injuries reported
- 30
- Recalls
- 0
We found 5 models of ASAHI SION blue by Asahi Intecc CO., Ltd. in the FDA's device registry (GUDID). Together they are named in 53 adverse event reports in the FDA's MAUDE database, received between Aug 18, 2017 and Jan 26, 2026 — classified as 30 injuries, 23 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported ASAHI SION blue problems
Device problems
- Detachment of Device or Device Component14
- Fracture10
- Unraveled Material7
- Break6
- Material Separation6
- Difficult to Remove5
- Material Fragmentation3
- Stretched3
Patient problems
- Foreign Body In Patient16
- No Clinical Signs, Symptoms or Conditions15
- No Consequences Or Impact To Patient8
- Device Embedded In Tissue or Plaque6
- Insufficient Information6
- Vessel Or Plaque, Device Embedded In3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- No Known Impact Or Consequence To Patient1
ASAHI SION blue models
5 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ASAHI SION blue AHW14R104SWire, guide, catheter29 reports
- ASAHI SION blue AHW14R004SWIRE, GUIDE, CATHETER16 reports
- ASAHI SION blue AHW14R004JWIRE, GUIDE, CATHETER5 reports
- ASAHI SION blue AHW14R304SWIRE, GUIDE, CATHETER2 reports
- ASAHI SION blue AHW14R104JWire, guide, catheter1 report
Recent ASAHI SION blue adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryJan 26, 2026Material Separation
USER FACILITY MEDSUN MANDATORY AND VOLUNTARY REPORT (UF REPORT #: (B)(4)) WAS RECEIVED VIA THE US DISTRIBUTOR ON JANUARY 5, 2026. IT WAS REPORTED THAT DURING A PROCEDURE FOR THE CORONARY ARTERY, AN ASAHI SION BLUE GUIDE WIRE FRACTURED. THE FRACTURED SEGMENT WAS RETRIEVED FROM THE CORONARY ARTERY. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT AND THE PATIENT WAS FINE WITHOUT ANY PROBLEMS AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24175961 ↗MalfunctionJan 15, 2026Material Fragmentation
THERE WAS NO HARM TO THE PATIENT. ASAHI SION BLUE GUIDEWIRE SHEARED OFF DURING THE CORONARY CASE. THE SHEARED PORTION OF THE WIRE WAS REMOVED FROM THE CORONARY ARTERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24088691 ↗MalfunctionSep 19, 2025Material Fragmentation; Unraveled Material; Material Fragmentation; Unraveled Material
SION BLUE PLACED IN LAD, 2ND SION BLUE TO DIAG, BALLOON OF OSTIAL DIAG INTO LAD, STENT PLACED IN OSTIAL DIAG INTO LAD, POST DILATION OF OSTIAL DIAGONAL, PLAQUE SHIFT IN LAD, STENT PLACED IN LAD WITH PROXIMAL EDGE OF THE STENT SLIGHTLY OVERLAPPING OSTIAL DIAG, POST DILATE STENT OF LAD, FINAL PICTURE, LAD WIRE WITHDRAWN WITHOUT INCIDENT, DIAG WIRE PULLED BACK AND TIP SEPARATION WAS NOTED, MICROCATHETER WAS TRAPPED IN AND ADVANCED CAREFULLY OVER THE UNRAVELED WIRE, DISTAL TIP WAS UNABLE TO BE FULLY CAPTURED, WIRE AND MICROCATHETER WERE PULLED OUT AS A UNIT. WIRE WAS WITHDRAWN IN ITS ENTIRETY. NO WIRE PIECES WERE LEFT BEHIND. NO HARM TO THE PATIENT. DISCHARGED AS NORMAL. NOTES: IT DOES NOT APPEAR THAT THE AFFECTED WIRE WAS EVER JAILED BEHIND A STENT OR HAD TO RE-CROSS A STENT. THE TIP POSITION OF THE DIAG WIRE NEVER CHANGED ONCE IT WAS INITIALLY POSITIONED INTO THE DIAG. THE AFFECTED WIRE APPEARED TO HAVE A SMALL LOOP IN IT FOR MUCH OF THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23097361 ↗InjurySep 17, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PLAIN OLD BALLOON ANGIOPLASTY (POBA) WAS PERFORMED TO TREAT A MODERATELY CALCIFIED AND MILDLY TORTUOUS LESION IN THE LEFT ANTERIOR DESCENDING ARTERY (LAD). AN ASAHI SION BLUE GUIDE WIRE WAS ADVANCED TO THE DISTAL LAD, WHILE A SECOND ASAHI SION BLUE GUIDE WIRE WAS ADVANCED TO THE LEFT CIRCUMFLEX ARTERY (LCX). A 2.5X12MM ONYX STENT (MEDTRONIC) WAS PLACED IN THE OBTUSE MARGINAL BRANCH (OM) AND BALLOON DILATATION WAS PERFORMED. AN ADDITIONAL 3.0X8MM ONYX STENT WAS DEPLOYED TO THE PROXIMAL OM. A 3.5X15MM XIENCE DRUG ELUTING STENT (DES) (ABBOTT) WAS DEPLOYED IN THE LAD, AND BALLOON DILATATION WAS PERFORMED. WHEN AN ATTEMPT WAS MADE TO REMOVE THE TWO SION BLUE GUIDE WIRES SIMULTANEOUSLY IN ORDER TO RE-POSITION THE SUBJECT SION BLUE GUIDE WIRE THAT HAD BEEN ADVANCED IN THE LAD TOWARD THE LCX, THE SUBJECT SION BLUE GUIDE WIRE WAS CAUGHT AND THE WIRE TIP WAS DETACHED. THE GUIDE WIRE WAS REPLACED WITH A DIFFERENT ONE AND RE-POSITIONED IN THE LAD. TWO BALLOON CATHETERS WERE USED FOR DILATATION TO SUCCESSFULLY REOPEN THE STENT. THE FINAL STENT OPTIMIZATION WAS PERFORMED WITH POST-DILATATION WITH AN UNIDENTIFIED BALLOON CATHETER. THE MID LAD LESION WAS NOT TREATED. THE PHY
Patient problems: Foreign Body In Patient
MAUDE report 23070614 ↗InjuryAug 27, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT STENT DEPLOYMENT WAS SUCCESSFULLY PERFORMED DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO THE MILDLY CALCIFIED AND 90-99% STENOSED UNSPECIFIED SEGMENT OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD). WHEN AN SECOND ASAHI SION BLUE GUIDE WIRE WAS ADVANCED TO MONITOR WITH THE INTRAVASCULAR ULTRASOUND SYSTEM (IVUS), THE SION BLUE GUIDE WIRE WAS CAUGHT IN THE FIRST DIAGONAL BRANCH (D1). A MICROCATHETER AND BALLOONS WERE PLACED TO REMOVE THE GUIDE WIRE, WITHOUT SUCCESS. THE STUCK SION BLUE GUIDE WIRE CAME OUT, LEAVING THE COATING AND TIP SEGMENT IN THE PATIENT ANATOMY. THE THIRD SUBJECT SION BLUE GUIDE WIRE WAS ADVANCED TO STENT THE WIRE FRAGMENT BUT PERFORATED IN THE D1, CAUSING CARDIAC TAMPONADE. PERICARDIAL DRAINAGE WAS PERFORMED, AND A BALLOON WAS DILATED OVER THE PERFORATION SITE. THE PATIENT WAS TAKEN TO A DIFFERENT HOSPITAL, WHERE A STENT WAS DEPLOYED OVER THE WIRE FRAGMENT. IT WAS REPORTED THAT THE PATIENT HAS BEEN ALREADY DISCHARGED.
Patient problems: Perforation of Vessels
MAUDE report 22901514 ↗InjuryAug 27, 2025Material Separation; Material Separation
IT WAS REPORTED THAT STENT DEPLOYMENT WAS SUCCESSFULLY PERFORMED DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO THE MILDLY CALCIFIED AND 90-99% STENOSED UNSPECIFIED SEGMENT OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD). WHEN THE SECOND ASAHI SION BLUE GUIDE WIRE WAS ADVANCED TO MONITOR WITH THE INTRAVASCULAR ULTRASOUND SYSTEM (IVUS), THE SUBJECT SION BLUE GUIDE WIRE WAS CAUGHT IN THE FIRST DIAGONAL BRANCH (D1). A MICROCATHETER AND BALLOONS WERE PLACED TO REMOVE THE GUIDE WIRE, WITHOUT SUCCESS. THE STUCK SION BLUE GUIDE WIRE CAME OUT, LEAVING THE COATING AND TIP SEGMENT IN THE PATIENT ANATOMY. A THIRD SION BLUE GUIDE WIRE WAS ADVANCED TO STENT THE WIRE FRAGMENT BUT PERFORATED IN THE D1, CAUSING CARDIAC TAMPONADE. PERICARDIAL DRAINAGE WAS PERFORMED, AND A BALLOON WAS DILATED OVER THE PERFORATION SITE. THE PATIENT WAS TAKEN TO A DIFFERENT HOSPITAL, WHERE A STENT WAS DEPLOYED OVER THE WIRE FRAGMENT. IT WAS REPORTED THAT THE PATIENT HAS BEEN ALREADY DISCHARGED.
Patient problems: Foreign Body In Patient
MAUDE report 22901513 ↗InjuryAug 4, 2025Difficult to Remove; Material Split, Cut or Torn; Difficult to Remove; Material Split, Cut or Torn
IT WAS REPORTED THAT A PERCUTANEOUS CORONARY INTERVENTION (PCI) WAS PERFORMED TO TREAT A HEAVILY CALCIFIED AND MODERATELY TORTUOUS 90-99% STENOSED TRIPLE VESSEL DISEASE. THE LEFT ANTERIOR DESCENDING ARTERY (LAD) WAS WIRED WITH AN ASAHI SION BLUE GUIDE WIRE AND THE HEAVILY CALCIFIED LESION IN THE LEFT CIRCUMFLEX ARTERY (LCX) WAS WIRED WITH AN ASAHI SUOH 03 GUIDE WIRE. THE LCX WAS PRE-DILATED WITH AN ORBUSNEICH SAPPHIRE BALLOON CATHETER (0.8 MM), AN UNSPECIFIED NON-COMPLIANT BALLOON (2.5 MM) AND THEN A SPECTRANETICS ANGIOSCULPT BALLOON CATHETER (2.5 MM). THE LAD WAS PRE-DILATED WITH AN UNSPECIFIED NON-COMPLIANT BALLOON (3.0 MM) AND AN ANGIOSCULPT BALLOON CATHETER (3.0 MM). TWO OVERLAPPING STENTS WERE THEN DEPLOYED FROM THE MID LAD BACK TO THE LEFT MAIN STEM (LMS), AND MODIFIED T STENTING WAS PERFORMED TO THE LCX. TWO OVERLAPPING STENTS WERE ALSO DEPLOYED FROM THE DISTAL LCX TO THE OSTIUM WITH KISSING BALLOON TECHNIQUE USING TWO UNSPECIFIED BALLOON CATHETERS (3.0 MM AND 3.5 MM) AND PROXIMAL OPTIMIZATION TECHNIQUE USING AN UNSPECIFIED NON-COMPLIANT BALLOON (5 MM) TO THE LMS. THE TIP OF THE SION BLUE GOT TRAPPED IN THE DISTAL LCX AFTER JAILED WITH LMS STENT, WHICH WAS REMOVED USING A TE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22697157 ↗InjuryMay 15, 2025Material Separation; Material Separation
IT WAS REPORTED THAT TWO ASAHI SION BLUE GUIDE WIRES WERE USED DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) FOR A MODERATELY CALCIFIED 75% OR LESS STENOSIS FROM THE LEFT MAIN TRUNK TO THE LEFT ANTERIOR DESCENDING ARTERY (LAD). ONE SION BLUE GUIDE WIRE WAS INSERTED INTO THE LAD AND THE OTHER SION BLUE GUIDE WIRE WAS INSERTED INTO THE LCX. THERE WAS A DISSECTION IN THE OSTIAL LAD AND THE GUIDE WIRE WAS UNCOILED AND FRACTURED. THE WIRE FRAGMENT EXTENDED INTO THE AORTA. LIFELONG DUAL ANTIPLATELET THERAPY WAS CONTINUED. IT WAS INFORMED THAT THE PATIENT HAD NO CONSEQUENCES AND THE BLOOD BLOW WAS FINE AFTER THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 22025461 ↗InjuryApr 25, 2025Insufficient Device Problem Information; Material Split, Cut or Torn; Insufficient Device Problem Information; Material Split, Cut or Torn
IT WAS REPORTED THAT AN ASAHI SION BLUE GUIDE WIRE WAS USED DURING A PLAIN OLD BALLOON ANGIOPLASTY (POBA) TO TREAT A HEAVILY CALCIFIED LESION IN THE MAIN TRUCK OF THE LEFT ANTERIOR DESCENDING ARTERY (LAD) EXTENDING TO THE DIAGONAL BRANCH. DURING REMOVAL OF THE GUIDE WIRE AT THE END OF THE PROCEDURE, RESISTANCE WAS MET AT THE DISTAL SEGMENT AND THE GUIDE WIRE COULD NOT BE REMOVED. AN UNSPECIFIED MICROCATHETER WAS USED TO REMOVE THE GUIDE WIRE AS AN ADDITIONAL TREATMENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY OR ANY OTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21907034 ↗MalfunctionMar 7, 2025Material Fragmentation; Material Fragmentation
TWO SION BLUE WIRES WERE USED. THE WIRE THAT WAS PLACED IN THE DISTAL SEGMENT OF THE LAD WAS LEFT PROLAPSED FOR ABOUT 15 MINUTES. WHEN PULLING THE WIRE BACK IT ENCOUNTERED CALCIUM IN THE PROXIMAL SEGMENT OF THE LAD AND THE TIP BROKE OFF. A STENT WAS PLACED TO TRAP THE FRAGMENT.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 21554264 ↗InjuryNov 29, 2024Material Separation; Material Separation
IT WAS REPORTED THAT THE TIP OF AN ASAHI SION BLUE GUIDE WIRE HAD FRACTURED DURING A COMPLEX PERCUTANEOUS CORONARY INTERVENTION (PCI) FOR THE HEAVILY DISEASED AND HEAVILY CALCIFIED LEFT CIRCUMFLEX ARTERY (LCX). THE OSTIAL LEFT ANTERIOR DESCENDING ARTERY (LAD) AND THE OSTIAL LCX WAS CALCIFIED. THE SION BLUE GUIDE WIRE WAS SHAPED AND INSERTED INTO THE PATIENT ANATOMY. IT WAS DIFFICULT TO ADVANCE THE SION BLUE GUIDE WIRE INTO THE LCX DUE TO THE TIGHT CURVATURE OF THE VESSEL. THE PHYSICIAN ADVANCED THE SION BLUE GUIDE WIRE INTO THE RELATIVELY EASILY ACCESSIBLE LAD AND ADJUSTED THE TIP CURVE. WHEN AN ATTEMPT WAS MADE TO DELIVER THE SION BLUE GUIDE WIRE TO THE LCX BY PULLING THE GUIDE WIRE BACK, THE WIRE TIP WAS FRACTURED IN THE LAD. THE LENGTH OF THE WIRE FRAGMENT WAS ESTIMATED AT LESS THAN 2CM AND CLOSE TO 1CM. AN UNSPECIFIED STENT WAS DEPLOYED TO TRAP THE WIRE FRAGMENT. THE PATIENT WAS IN CARDIOGENIC SHOCK, AND DURING SUBSEQUENT REVASCULARIZATION SUFFERED CARDIAC ARREST AND DIED. THE PHYSICIAN COMMENTED THAT THE CARDIAC ARREST WAS NOT RELATED DIRECTLY TO THE WIRE FRACTURE, BUT DUE TO A NON-ASAHI SHOCKWAVE CATHETER.
Patient problems: Foreign Body In Patient
MAUDE report 20808846 ↗InjuryNov 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A PLAIN OLD BALLOON ANGIOPLASTY (POBA) WAS PERFORMED TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED 75-90% STENOSIS IN THE RIGHT CORONARY ARTERY (RCA). AN ASAHISION BLUE GUIDE WIRE WAS DELIVERED TO AN UNSPECIFIED SIDE BRANCH FOR PROTECTION PURPOSE AND A NON-ASAHI STENT WAS DEPLOYED IN THE RCA. THE ANGLE BETWEEN THE RCA AND THE SIDE BRANCH WAS ABOUT 90? AND THE SION BLUE WAS 90? CURVED. DURING REMOVAL OF THE SION BLUE, PART OF THE GUIDE WIRE REMAINED BETWEEN THE VESSEL WALL AND THE STENT. THE REMOVED SION BLUE WAS FOUND FRAYED. AS THE RETAINED WIRE FRAGMENT WAS VERY THIN AND WAS TRAPPED BEHIND THE STENT, THE PHYSICIAN DECIDED TO LEAVE IT IN SITU. THE GUIDE WIRE WAS THEN REPLACED TO RESUME THE PROCEDURE. THE PATIENT WAS REPORTEDLY FINE AFTER THE PROCEDURE.
Patient problems: Foreign Body In Patient
MAUDE report 18127508 ↗InjuryApr 30, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE PROCEDURE WAS A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO TREAT A 18MM LONG, 75% STENOSIS IN THE PROXIMAL LEFT ANTERIOR DESCENDING ARTERY (LAD). DURING THE PCI, THE DISTAL TIP OF AN ASAHI SION BLUE GUIDE WIRE FRACTURED AND WAS RETAINED IN THE PATIENT, REQUIRING SURGICAL INTERVENTION. THE GUIDE WIRE WAS PARKED IN THE DIAGONAL ARTERY TO PROTECT VESSEL WHILE TREATING THE PROXIMAL LAD, ALLOWING ACCESS TO THE DIAGONAL ARTERY IN THE EVENT THE DIAGONAL ARTERY OCCLUDED DUE TO PLAQUE SHIFT FROM STENT DEPLOYMENT IN THE LAD. AFTER STENT DEPLOYMENT, THE WIRE IN THE DIAGONAL ARTERY WAS REMOVED, BUT THE DISTAL TIP OF THE WIRE FRACTURED DURING THE PROCESS. ALTHOUGH ATTEMPTS WERE MADE TO RETRIEVE IT USING A PERCUTANEOUS TECHNIQUE, THEY WERE UNSUCCESSFUL AND IT REQUIRED SURGICAL REMOVAL. THE PATIENT WAS STABLE AFTER THE PROCEDURE AND DISCHARGED FROM HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16841470 ↗InjuryApr 21, 2023Fracture; Fracture
THE DISTAL TIP OF THE WIRE FRACTURED AND WAS RETAINED IN THE PATIENT, REQUIRING SURGICAL INTERVENTION. THIS IS ALL THE INFORMATION I HAVE AT THE MOMENT. CUSTOMER STATED THAT THE PRODUCT IS AVAILABE FOR RETURN.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body In Patient
MAUDE report 16792396 ↗InjuryMar 23, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT TWO ASAHI SION BLUE GUIDE WIRES WERE USED IN A PERCUTANEOUS CORONARY INTERVENTION (PCI) TO TREAT AN UNSPECIFIED VESSEL AND THE TIP OF ONE OF THE SION BLUE HAD SEPARATED. AFTER STENTING THE TARGET LESION, THE SION BLUE WAS JAILED AND GOT TRAPPED IN STENT STRUTS. WHEN PULLING FAIRLY HARD TO RELEASE THE GUIDE WIRE, THE WIRE TIP BECAME UNRAVELED AND WAS LEFT IN VESSEL. UNRAVELED COIL FROM THE WIRE FRAGMENT WAS HANGING OUT INTO THE AORTA. ATTEMPTS TO RETRIEVE THE FRAGMENT WITH SNARE WAS NOT SUCCESSFUL. THE PATIENT WAS TREATED WITH USUAL DAPT.
Patient problems: Foreign Body In Patient
MAUDE report 16596435 ↗
Other Asahi Intecc CO., Ltd. products
All Asahi Intecc CO., Ltd. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page