ASAHI Gladius Mongo14 ESRecalls & reported problems across all models
Asahi Intecc CO., Ltd. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls
- 0
We found 3 models of ASAHI Gladius Mongo14 ES by Asahi Intecc CO., Ltd. in the FDA's device registry (GUDID). Together they are named in 33 adverse event reports in the FDA's MAUDE database, received between May 14, 2019 and Sep 10, 2026 — classified as 20 injuries, 13 malfunctions.
The most frequently reported specific device problem is “Material Separation” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported ASAHI Gladius Mongo14 ES problems
Device problems
- Material Separation9
- Detachment of Device or Device Component8
- Fracture6
- Peeled/Delaminated6
- Break5
- Material Split, Cut or Torn4
- Material Fragmentation3
- Device Dislodged or Dislocated1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Foreign Body In Patient10
- Device Embedded In Tissue or Plaque8
- Insufficient Information5
- Vessel Or Plaque, Device Embedded In1
ASAHI Gladius Mongo14 ES models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ASAHI Gladius Mongo14 ES AP14R324SWire, guide, catheter20 reports
- ASAHI Gladius Mongo14 ES AP14R324PWire, guide, catheter12 reports
- ASAHI Gladius Mongo14 ES AP14R024SWire, guide, catheter1 report
Recent ASAHI Gladius Mongo14 ES adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 10, 2026Material Separation
It was reported that an Asahi Gladius Mongo 14 ES guide wire was used during a plain old balloon angioplasty (POBA) to treat a heavily calcified, heavily tortuous chronic total occlusion (CTO) in the mid anterior tibial artery (ATA). The ATA was totally occluded throughout its entire length from the proximal segment. The Gladius Mongo 14 ES guide wire was advanced into the proximal ATA. While advancing the tip of the guide wire to make a small J shape, it got caught and entangled, forming a knot in one of the side branches of the ATA branching off the proximal segment. Subsequently, the entangled portion of the guide wire could not be unknotted and withdrawn. The tip of the Gladius Mongo 14 ES guide wire broke off within the small side branch of the ATA. After the tip fracture, the proximal segment of the Gladius Mongo 14 ES guide wire was successfully withdrawn from the patient anatomy. The separated wire tip remained in the patient anatomy and could not be retrieved. The physician commented that the tip of the guide wire became entangled and formed a knot as a result ofrapid advancement and rotation of the wire within an extremely narrow side branch of the ATA. The size of the kn
Patient problems: Foreign Body In Patient
MAUDE report 50722714 ↗InjurySep 9, 2026Material Separation
It was reported that an Asahi Gladius Mongo 14 ES guide wire was used during a plain old balloon angioplasty (POBA) to treat a heavily calcified, heavily tortuous chronic total occlusion (CTO) in the mid anterior tibial artery (ATA). The ATA was totally occluded throughout its entire length from the proximal segment. The Gladius Mongo 14 ES guide wire was advanced into the proximal ATA. While advancing the tip of the guide wire to make a small J shape, it got caught and entangled, forming a knot in one of the side branches of the ATA branching off the proximal segment. Subsequently, the entangled portion of the guide wire could not be unknotted and withdrawn. The tip of the Gladius Mongo 14 ES guide wire broke off within the small side branch of the ATA. After the tip fracture, the proximal segment of the Gladius Mongo 14 ES guide wire was successfully withdrawn from the patient anatomy. The separated wire tip remained in the patient anatomy and could not be retrieved.The physician commented that the tip of the guide wire became entangled and formed a knot as a result of rapid advancement and rotation of the wire within an extremely narrow side branch of the ATA. The size of the kn
Patient problems: Foreign Body In Patient
MAUDE report 50683808 ↗InjuryJul 21, 2026Peeled/Delaminated
IT WAS REPORTED THAT A PERCUTANEOUS PERIPHERAL INTERVENTION (PPI) WAS PERFORMED FOR A HEAVILY CALCIFIED LESION AT THE ORIGIN OF THE PERONEAL ARTERY. THE LESION WAS CROSSED WITH AN ASAHI GLADIUS MONGO 14 ES GUIDE WIRE. AFTER THE GLADIUS MONGO 14 ES GUIDE WIRE WAS REMOVED FOR WIRE EXCHANGE, PEELING OF THE POLYMER JACKET WAS OBSERVED, AND THE NEW GUIDE WIRE COULD NOT BE ADVANCED. THE DEVICE SYSTEM WAS THEN REMOVED AND THE PROCEDURE WAS RESTARTED. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT AND THE PATIENT WAS STABLE AFTER THE PROCEDURE. MFR REPORT #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25891167 ↗InjuryJul 21, 2026Peeled/Delaminated
IT WAS REPORTED THAT AN ASAHI GLADIUS MONGO 14 ES GUIDE WIRE WAS USED DURING A PERCUTANEOUS PERIPHERAL INTERVENTION (PPI) FOR AN UNSPECIFIED LESION. AFTER THE GLADIUS MONGO 14 ES GUIDE WIRE WAS REMOVED FOR WIRE EXCHANGE, PEELING OF THE POLYMER JACKET WAS OBSERVED, AND THE NEW GUIDE WIRE COULD NOT BE ADVANCED. THE DEVICE SYSTEM WAS THEN REMOVED AND THE PROCEDURE WAS RESTARTED. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT AND THE PATIENT WAS STABLE AFTER THE PROCEDURE. MFR REPORT #: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25890913 ↗InjuryJul 17, 2026Peeled/Delaminated
IT WAS REPORTED THAT A PERCUTANEOUS PERIPHERAL INTERVENTION (PPI) WAS PERFORMED FOR A HEAVILY CALCIFIED LESION AT THE ORIGIN OF THE PERONEAL ARTERY. THE LESION WAS CROSSED WITH AN ASAHI GLADIUS MONGO 14 ES GUIDE WIRE. AFTER THE GLADIUS MONGO 14 ES GUIDE WIRE WAS REMOVED FOR WIRE EXCHANGE, PEELING OF THE POLYMER JACKET WAS OBSERVED, AND THE NEW GUIDE WIRE COULD NOT BE ADVANCED. THE DEVICE SYSTEM WAS THEN REMOVED AND THE PROCEDURE WAS RESTARTED. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT AND THE PATIENT WAS STABLE AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25851291 ↗InjuryJun 13, 2025Material Separation; Material Separation
IT WAS REPORTED THAT AN ASAHI GLADIUS MG14 ES GUIDE WIRE WAS USED FOR REVASCULARIZATION OF A MILDLY CALCIFIED 100% OCCLUDED LESION IN THE ULNAR ARTERY. THE GUIDE WIRE WAS EXCESSIVELY TORQUED AND KNOTTED UP IN THE DISTAL ULNAR ARTERY. DURING REMOVAL ATTEMPTS, A DISTAL COIL SEGMENT WAS DETACHED FROM THE PROXIMAL SIDE AND LEFT IN THE PATIENT ANATOMY. THE PHYSICIAN CONCLUDED THAT THE WIRE FRAGMENT COULD BE LEFT IN SITU WITHOUT ISSUE. IT WAS INFORMED THAT THE PATIENT WAS DISCHARGED, WHOSE CONDITION WAS THE SAME AS THAT BEFORE THE PROCEDURE ALTHOUGH THE INTENDED REVASCULARIZATION WAS NOT ACHIEVED.
Patient problems: Foreign Body In Patient
MAUDE report 22200441 ↗MalfunctionMay 27, 2025Material Fragmentation; Material Fragmentation
MONGO USED VIA ULNAR ARTERY TO ATTEMPT REVASCULARIZATION; WIRE WAS HEAVILY TORQUED AND "KNOTTED UP" IN THE DISTAL ULNAR ARTERY. WHEN ATTEMPTING TO REMOVE THE WIRE FROM THE ANATOMY, PART OF DISTAL COIL DETACHED FROM THE PROXIMAL COIL & WAS LEFT IN THE BODY.
Patient problems: Foreign Body In Patient
MAUDE report 22092511 ↗InjuryOct 17, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PERIPHERAL PERCUTANEOUS INTERVENTION (PPI) WAS PERFORMED FOR A HEAVILY CALCIFIED, MODERATELY TORTUOUS, AND 90-99% OCCLUDED LESION IN THE ANTERIOR TIBIAL ARTERY (ATA). WHEN AN ATTEMPT WAS MADE WITH AN ASAHI GLADIUS MONGO 14 ES GUIDE WIRE, INSERTED IN AN ASAHI CROSSLEAD PERIPHERAL SUPPORT CATHETER, TO PERFORM PEDAL LOOP REVASCULARIZATION. DURING PROCEDURE, THE SUBJECT GLADIUS MONGO 14 ES GUIDE WIRE WAS CAUGHT IN THE DISTAL TIBIAL ARTERY (TA). AFTER SEVERAL WITHDRAWAL ATTEMPTS WITH DIFFICULTY, THE GUIDE WIRE WAS REMOVED BUT FOUND ITS DISTAL SEGMENT DETACHED. AN UNSPECIFIED MICRO SNARE WAS USED AND THE WIRE FRAGMENT WAS SUCCESSFULLY RETRIEVED. IT WAS INFORMED THAT THERE WERE NO ADVERSE PATIENT EFFECTS AND THE PROCEDURE WAS TERMINATED. THE PATIENT WAS REPORTEDLY DISCHARGED AFTER HEMOSTASIS WAS ACHIEVED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20478017 ↗MalfunctionOct 3, 2024Break; Break
WHILE WORKING ON VERY CALCIFIC AND DIFFUSELY DISEASED ANTERIOR TIBIAL ARTERY, THE MONGO ES .014 WIRE BECAME TRAPPED IN THE DISTAL AT BY THE PEDAL LOOP WHILE IN THE .014 170CM CROSSWALK. IT WAS NOTED THAT WHEN THE WIRE WAS FINALLY ABLE TO BE REMOVED, THE DISTAL TIP OF THE WIRE WAS DISCONNECTED FROM THE REST OF THE WIRE. A MICRO SNARE WAS USED TO SUCCESSFULLY REMOVE THE BROKEN PIECE OF THE WIRE. NO ISSUES WERE NOTED, AND CASE WAS TERMINATED. PATIENT WAS RECOVERED IN THE RECOVERY AREA UNTIL HEMOSTASIS OF FEMORAL SITE WAS SEALED, THEN DISCHARGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20364338 ↗MalfunctionSep 12, 2024Material Fragmentation; Material Fragmentation
MD KNUCKLED WIRE WITH IN RCA CTO AND UPON REMOVING JACKETED PORTION CAME OFF CORE WIRE. THE JACKETED PORTION BROKE OFF IN THE SUBINTIMAL SPACE AND WAS UNRETRIEVABLE. THERE WAS NO SIDE EFFECTS. THEY WERE UNABLE TO CROSS DUE TO TOUGH CTO AND THE AMOUNT OF RADIATION, SO THE SECOND PROCEDURE WAS PERFORMED LATER. IN THE SECOND PROCEDURE, THE CTO WAS CROSSED FROM RETROGRADE APPROACH AND STENTED.
Patient problems: Foreign Body In Patient
MAUDE report 20207610 ↗MalfunctionSep 6, 2024Break; Break
DR (B)(6) WAS TRYING TO NAVIGATE AN EXTREMELY CALCIFIED AT CTO WITH NO BLOOD FLOW. HE TRIED TRAVERSING AROUND THE SEGMENT AND WIRE BECAME STUCK AND WHEN HE TRIED TO ADVANCE ABOUT A 1CM PORTION OF THE DISTAL TIP CAME OFF. BECAUSE OF WHERE THE TIP BROKE AND NO FLOW IT WAS DETERMINED BY PHYSICIAN THAT LEAVING WHERE IT WAS WOULD NOT POSE ANY SORT OF RISK TO PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 20158906 ↗InjurySep 1, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PERIPHERAL PERCUTANEOUS INTERVENTION (PPI) WAS PERFORMED FOR A HEAVILY CALCIFIED AND MODERATELY TORTUOUS CHRONIC TOTAL OCCLUSION (CTO) IN THE TIBIAL ARTERY (TA). WHEN AN ATTEMPT WAS MADE WITH AN ASAHI GLADIUS MONGO 14 ES GUIDE WIRE TO CROSS THE LESION, THE GUIDE WIRE WAS CAUGHT BY THE LESION. AS FURTHER ATTEMPTS WERE MADE TO ADVANCE THE GUIDE WIRE, IT BROKE OFF FOR APPROXIMATELY 1CM FROM THE WIRE TIP. HAVING CONSIDERED THAT THERE HAD BEEN NO FLOW IN THE SEGMENT WHERE THE WIRE WAS DETACHED, THE PHYSICIAN CONCLUDED THAT THE FRAGMENT COULD BE LEFT IN SITU AS THERE WOULD BE NO RISK TO THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT PROBLEM, AND THE PATIENT WAS DISCHARGED IN A FEW HOURS AFTER THE PROCEDURE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 20115710 ↗InjuryJul 18, 2024Material Separation; Material Separation
IT WAS REPORTED THAT A PERCUTANEOUS PERIPHERAL INTERVENTION (PPI) WAS PERFORMED TO TREAT THE RIGHT LOWER EXTREMITY FOR A PATIENT WHO HAD PERIPHERAL ARTERY DISEASE (PAD) AS WELL AS A HISTORY OF BELOW-THE-KNEE AMPUTATION OF THE LEFT LOWER EXTREMITY. ANGIOGRAPHY SHOWED MULTIPLE CALCIFIED LESIONS THROUGHOUT THE RIGHT SUPERFICIAL FEMORAL ARTERY (SFA), POPLITEAL ARTERY, RIGHT TIBIOPERONEAL TRUNK EXTENDING THROUGHOUT RIGHT POST TIBIAL ARTERY (PTA). AN UNSPECIFIED SHEATH WAS INSERTED INTO THE RIGHT SFA VIA CONTRALATERAL APPROACH AND AN ASAHI GLADIUS MONGO14 ES GUIDE WIRE WAS USED AS THE INITIAL PERIPHERAL GUIDE WIRE. THE GUIDE WIRE WAS SUCCESSFULLY ADVANCED THROUGH SFA, POPLITEAL ARTERY AND TIBIOPERONEAL TRUNK TO THE PTA ; HOWEVER, THE GUIDE WIRE COULD NOT BE ANY FURTHER AFTERWARDS. IN ATTEMPTS TO CROSS THE LESION WITH ANOTHER GUIDE WIRE, AN ASAHI CROSSWALK MICROCATHETER WAS ADVANCED AND THE GLADIUS MONGO14 ES GUIDE WIRE WAS THEN REMOVED. UPON REVIEW OF THE REMOVED GLADIUS MONGO14 ES GUIDE WIRE, THE TIP OF THE GUIDE WIRE WAS FOUND DETACHED. ANGIOGRAPHY SHOWED THAT THE DETACHED WIRE TIP WAS LEFT IN THE DISTAL PTA AT THE BEGINNING OF THE CHRONIC TOTAL OCCLUSION (CTO). THE PHYSICIAN DETERMINE
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 19769155 ↗MalfunctionJul 11, 2024Fracture; Fracture
SFA, POPLITEAL, RT TIBIOPERONEAL TRUNK EXTENDING THROUGHOUT RT POST TIBIAL ARTERY. CONTRALATERAL PLACEMENT OF SHEATH TO RT SFA COMPLETED AND MONGO14 ES USED AS INITIAL PERIPHERAL GUIDE WIRE. AFTER SUCCESSFULLY NAVIGATING THROUGH SFA, POPLITEAL AND TIBIOPERONEAL TRUNK AND MOST OF THE RT POST TIB. WIRE ADVANCEMENT WAS STOPPED, CROSSWALK ADVANCED AND MONGO WIRE WAS REMOVED WITH THE INTENT TO TRY ANOTHER WIRE TO ADVANCE. UPON REVIEW OF WIRE POST REMOVAL, THE TIP OF THE WIRE EXCLUDING THE CORE WAS STRIPPED FROM THE WIRE. UPON ANGIOGRAPHIC REVIEW, THE TIP OF WIRE WAS LEFT IN THE DISTAL POSTERIOR TIBIAL ARTERY AT THE BEGINNING OF THE CTO. THE PHYSICIAN AND I DISCUSSED THE FINDING AND HE STATED THAT IT WOULD CAUSE NO FURTHER HARM TO THE PATIENT BECAUSE OF THE EXISTING CTO.
Patient problems: Foreign Body In Patient
MAUDE report 19727324 ↗InjuryApr 15, 2024Material Fragmentation; Material Fragmentation
ACCESSED THE ANTERIOR TIBIAL THEN ATTEMPTED TO CROSS THE OCCLUDED PROXIMAL AND MID ANTERIOR TIBIAL ARTERY OCCLUSION. AT PROXIMAL OCCLUSION AREA, WHEN TRYING TO WIRE INTO THE DISTAL SEGMENT OF THE POPLITEAL ARTERY, WE WERE SUBINTIMAL TRYING TO NAVIGATE BACK INTO THE NATIVE AND THE WIRE TIP BROKE OFF WHEN TRYING TO REMOVE WIRE FROM CATHETER.
Patient problems: Foreign Body In Patient
MAUDE report 19105409 ↗
Other Asahi Intecc CO., Ltd. products
All Asahi Intecc CO., Ltd. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.