Skip to content

BIOKNOTLESS PANACRYL ETHIBONDRecalls & reported problems across all models

Depuy Mitek, LLC · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one BIOKNOTLESS PANACRYL ETHIBOND model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
35
Deaths reported
0
Injuries reported
1
Recalls
2

We found 3 models of BIOKNOTLESS PANACRYL ETHIBOND by Depuy Mitek, LLC in the FDA's device registry (GUDID). Together they are named in 35 adverse event reports in the FDA's MAUDE database, received between Jun 19, 2018 and Jun 14, 2023 — classified as 1 injury, 34 malfunctions.

The most frequently reported specific device problem is “Break” (30 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

BIOKNOTLESS PANACRYL ETHIBOND recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Apr 13, 2023Open, Classified

    IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

    Z-1681-2023 · Root cause: Use error · Recall details · FDA record ↗

  2. Apr 13, 2023Open, Classified

    IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

    Z-1678-2023 · Root cause: Use error · Recall details · FDA record ↗

Most reported BIOKNOTLESS PANACRYL ETHIBOND problems

Device problems

  • Break30
  • Migration3
  • Material Frayed1
  • Material Twisted/Bent1

Patient problems

  • Foreign Body In Patient22
  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient5
  • No Known Impact Or Consequence To Patient2
  • Not Applicable1
  • Unspecified Tissue Injury1

BIOKNOTLESS PANACRYL ETHIBOND models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent BIOKNOTLESS PANACRYL ETHIBOND adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Depuy Mitek, LLC products

All Depuy Mitek, LLC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page