BIOKNOTLESS PANACRYL ETHIBOND 212725Recalls & reported problems
Depuy Mitek, LLC · BIOKNOTLESS Plus Anchor w/PANACRYL PLA Absorbable Anchor, Size 1 (4 metric) PANACRYL braided absorbable suture anchor loop, Size 2/0 green ETHIBOND braided polyester suture utility loop. · Class 2 · Implant
All models of BIOKNOTLESS PANACRYL ETHIBOND (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886705002115), received between Jul 24, 2018 and Jul 24, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
6 other recalls cover devices cleared under the same FDA submission (K062170, K150209); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Apr 13, 2023Class IIOpen, Classified
IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time
Z-1678-2023 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2018Break; Break
THE SURGEON WAS DOING A BANKART REPAIR AND WHEN TRYING TO INSERT THE TIP OF THE ANCHOR INTO THE HOLE ONE SIDE OF THE TIP BROKE OFF. RETRIEVED THE PIECE OF THE ANCHOR THAT BROKE. OPENED A NEW ANCHOR AND COMPLETED THE CASE WITH NO PROBLEMS. THE PATIENT, A FEMALE ESTIMATED TO BE IN HER UPPER 20'S TO EARLY 30'S, HAD AVERAGE BONE QUALITY. THE PATIENT HAD OTHER SURGERIES IN HER PAST, A PRIOR SHOULDER SCOPE THAT DID NOT INVOLVE ANY ANCHORS, AND A BACK SURGERY. THIS SURGERY REMAINED ARTHROSCOPIC AFTER THE COMPLAINT DEVICE BROKE. THE PROCEDURE WAS EXTENDED 3 TO 5 MINUTES DUE TO THE ANCHOR BREAKING. THE COMPLAINT DEVICE WAS REMOVED AND DISCARDED. THE SALES REP STATED THAT SHE BELIEVES THE SURGEON WAS OFF AXIS DURING INSERTION AT THE TIME THE TIP OF THE ANCHOR BROKE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7715145 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062170, K150209. One submission often covers many models, so check the product and lot details in each record.
- Z-1679-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
- Z-1680-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
- Z-1677-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
- Z-1681-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
- Z-0316-2012 · Oct 31, 2011 · TerminatedDePuy Mitek Lupine Loop Plus w/ Orthocord DS Catalog #210709 The LUPINE BR Anchor System is indicated for use in soft tissue to bone fixation in association with adequate pos
- Z-0315-2012 · Oct 31, 2011 · TerminatedDePuy Mitek Lupine Loop Plus w/ Orthocord Catalog #210708 The LUPINE BR Anchor System is indicated for use in soft tissue to bone fixation in association with adequate posto
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page