Skip to content

BIOKNOTLESS PANACRYL ETHIBOND recallZ-1678-2023 · DePuy Mitek, Inc., a Johnson & Johnson Co. · initiated Apr 13, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DePuy Mitek, Inc., a Johnson & Johnson Co. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

IFU Warning: Applying bending force and/or bending force with impaction to the BIOKNOTLESS Anchor Inserter may cause a fragment of the inserter tip to break off during use and be retained in the patient. broken BIOKNOTLESS Anchor inserter tip detected during surgery may require removal, potentially causing bone damage and longer surgery time

FDA root cause category: Use error

What the company is doing

DePuy Mitek issued Urgent Medical Device Correction (Notification) Letter on 4/13/23 via First Class Mail to US consignees. Internal and sales/distributors notified via email. International affiliates will be notified via email and will be initiated per local regulations. Letter states reason for recall, health risk and action to take: 1. Product is NOT being removed from the field and does not need to be returned. 2. Familiarize yourself with the content 2. Familiarize yourself with the content of this letter. 3. Review, complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to [email protected] within three (3) business days of receipt of this notification. IMPORTANT: Complete the BRF even if you do not have any of the subject product in your current inventory. 4. Forward this notice to any personnel in your facility who need to be informed. 5. If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 6. Post a copy of this notice in a visible area for awareness and keep a copy for your records. If you have any questions, please contact your local DePuy Synthes Sales Consultant. For Medical Information request, please visit our website: https://www.jnjmedicaldevices.com/mir.

Recall details

Recall number
Z-1678-2023
Status
Open, Classified
Initiated by the company
Apr 13, 2023
Posted by the FDA
Jun 2, 2023
Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Quantity
540 units
Product code
MAI
FDA submissions
K040827, K062170, K070925, K150209
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, Switzerland, China, Czech Republic, Germany, Egypt, Spain, United Kingdom, Hong Kong, Hungary, Ireland, India, Italy, Korea, Netherlands, New Zealand, Slovakia, Thailand.

Product description

DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725

Code information (models, lots, serial numbers)

UDI-DI: 10886705002115 All lot numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

71 devices cleared under K040827, K062170, K070925, K150209 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 92210).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page