CardioChekRecalls & reported problems across all models
Henry Schein, Inc. · 3 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 3
We found 3 models of CardioChek by Henry Schein, Inc. in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Jul 22, 2019 and Jan 10, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (5 reports).
3 FDA recall records name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
CardioChek recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Dec 2, 2019Terminated
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
Z-0835-2020 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 2, 2019Terminated
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
Z-0834-2020 · Root cause: Device Design · Recall details · FDA record ↗
- Jun 7, 2019Terminated
The MEMo chips provided with lot P901 may be mislabeled, the MEMo Chip may display Q901 on the analyzer LCD screen, instead of P901
Z-2158-2019 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported CardioChek problems
Device problems
- Electrical /Electronic Property Problem5
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
CardioChek models
3 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CardioChek 5700411ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL1 recall5 reports
- CardioChek 5700534TUBE, COLLECTION, CAPILLARY BLOOD1 recall0 reports
- CardioChek 5700414LDL & VLDL PRECIPITATION, HDL1 recall0 reports
Recent CardioChek adverse event reports
The 5 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJan 10, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER THAT THEIR CARDIOCHEK PLUS ANALYZER IS "HOT WHEN USED". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18484431 ↗MalfunctionJan 7, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT WHEN THEY INSERT BATTERIES, THEIR ANALYZER "HEATS UP TO THE POINT OF SMOKING". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13193856 ↗MalfunctionJul 26, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT AFTER REPLACING THE BATTERIES THE ANALYZER WAS "VERY HOT." THE CUSTOMER ALSO STATED THAT TWO OF THE BATTERIES PLASTIC COATING HAD MELTED. THERE WAS NO ALLEGATION OF PATIENT/USER HARM. THERE WAS NO ALLEGATION OF INCORRECT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12226035 ↗MalfunctionFeb 12, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT THE ANALYZER IS "FRIED". UPON REPLACING THE BATTERIES, THE CUSTOMER STATED "IT HEATED UP AND THE BATTERIES EXPLODED". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WAS NO ALLEGATION OF INCORRECT RESULTS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9699197 ↗MalfunctionJul 22, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER REPORTED THAT THE ANALYZER WAS "INITIALLY GETTING HOT" WHEN THEY WENT BACK TO USE IT AGAIN IT "GOT EXTREMELY HOT". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8812244 ↗
Other Henry Schein, Inc. products
All Henry Schein, Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page