Skip to content

CardioChek recallZ-0835-2020 · Polymer Technology Systems, Inc. · initiated Dec 2, 2019

Class II Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Polymer Technology Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

FDA root cause category: Device Design

What the company is doing

Urgent Medical Device Correction notification letters dated 11/15/19 were sent to customers. Actions to be taken by the Customer/Distributor/User: 1. Immediately check your inventories and apply the sticker/labels as directed. If you are an end user: Please adhere one label to the outside of the battery door compartment of your analyzer and apply the other label to the analyzer user guide. If you are a distributor: Please remove the tape liner on the back of the large sticker leaving both small labels intact for the end user to apply. Apply the large sticker to the top panel of the analyzer box, and press firmly. 2. If you have further distributed this product, immediately identify the users and notify them at once of this product correction. Forward this notice, stickers, and response form to all customers/users who did, or who may have, received, this product. 3. Once you have taken the appropriate action, please fill out the attached response form and return it to PTS Diagnostics via the information below. It is important that you complete and return the response form, even if you do not have affected product on hand. 4. Please contact PTS for additional stickers and a PDF version of this letter and response form, if needed. Thank you for your prompt action, and we apologize for any inconvenience this issue may have caused you and your facility. If you have any questions, please contact Customer Service, Monday through Friday, 8:00 AM to 5:00 PM, Eastern Time at +1 319-870-5610 or 877-870-5610 (U.S. toll-free), or via email at [email protected].

Recall details

Recall number
Z-0835-2020
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 2, 2019
Recalling firm
Polymer Technology Systems, Inc.
Quantity
44,448 total
Product code
NBW
FDA submissions
K140068
Distribution
The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

Product description

Henry Schein CardioChek Plus Professional Analyzer, New Analyzer, boxed, with carrying case, Reference Number 4060 HS, 570-0411

Code information (models, lots, serial numbers)

All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00304040037713

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 84274).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page