Skip to content

CardioChek 5700411Recalls & reported problems

Henry Schein, Inc. · CardioChek Plus Analyzer · Class 1

All models of CardioChek (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00304040037713), received between Jul 22, 2019 and Jan 10, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (5 reports).

1 FDA recall record name this device's identifier.

3 other recalls cover devices cleared under the same FDA submission (K140068); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 2, 2019Class IITerminated

    One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

    Z-0835-2020 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Electrical /Electronic Property Problem5

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 10, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER THAT THEIR CARDIOCHEK PLUS ANALYZER IS "HOT WHEN USED". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18484431 ↗
  • MalfunctionJan 7, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT WHEN THEY INSERT BATTERIES, THEIR ANALYZER "HEATS UP TO THE POINT OF SMOKING". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13193856 ↗
  • MalfunctionJul 26, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT AFTER REPLACING THE BATTERIES THE ANALYZER WAS "VERY HOT." THE CUSTOMER ALSO STATED THAT TWO OF THE BATTERIES PLASTIC COATING HAD MELTED. THERE WAS NO ALLEGATION OF PATIENT/USER HARM. THERE WAS NO ALLEGATION OF INCORRECT RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12226035 ↗
  • MalfunctionFeb 12, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT THE ANALYZER IS "FRIED". UPON REPLACING THE BATTERIES, THE CUSTOMER STATED "IT HEATED UP AND THE BATTERIES EXPLODED". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WAS NO ALLEGATION OF INCORRECT RESULTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9699197 ↗
  • MalfunctionJul 22, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    THE CUSTOMER REPORTED THAT THE ANALYZER WAS "INITIALLY GETTING HOT" WHEN THEY WENT BACK TO USE IT AGAIN IT "GOT EXTREMELY HOT". THERE WERE NO ALLEGATIONS OF PATIENT/USER HARM. THERE WERE NO ALLEGATIONS OF INCORRECT RESULTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8812244 ↗

FDA clearance and classification

Cleared through Special 510(k) K140068 on May 22, 2015 (decision: Substantially Equivalent). Submitted by Polymer Technology Systems, Inc. as “Cardiochek Plus Test System, Cardiochek Home Test System”. The FDA review took 494 days. FDA record ↗

The FDA regulates product code CHH (enzymatic esterase--oxidase, cholesterol) under 21 CFR 862.1175 as a Class I device. 21 CFR 862.1175 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K140068. One submission often covers many models, so check the product and lot details in each record.

  • Z-0833-2020 · Dec 2, 2019 · Terminated
    PTS Diagnostics CardioChek Plus Professional Analyzer, Refurbished analyzer, without box or carrying case, Reference Numbers 2702
  • Z-0832-2020 · Dec 2, 2019 · Terminated
    PTS Diagnostics CardioChek Plus Professional Analyzer, New analyzer, boxed, with carrying case, Reference Numbers 2700
  • Z-0834-2020 · Dec 2, 2019 · Terminated
    Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036

Common questions about the CardioChek 5700411

Has the CardioChek 5700411 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0835-2020 (Class II), began on Dec 2, 2019: One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the CardioChek 5700411?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Jul 22, 2019 and Jan 10, 2024. In those reports, the most common specific device problem is “Electrical /Electronic Property Problem” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CardioChek 5700411?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CardioChek 5700411 FDA approved?

It was cleared through Special 510(k) K140068 on May 22, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CardioChek 5700411, and is it still sold?

The FDA's device database lists Henry Schein, Inc. as the company responsible for it (brand CardioChek, model 5700411). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL devices

All ENZYMATIC ESTERASE--OXIDASE, CHOLESTEROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page