CoLagRecalls & reported problems across all models
In2Bones USA, LLC · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names at least one CoLag model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 2 models of CoLag by In2Bones USA, LLC in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between Jan 10, 2022 and Mar 19, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
CoLag recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 9, 2026Open, Classified
Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.
Z-0157-2027 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported CoLag problems
Device problems
- Fracture2
Patient problems
- Bone Fracture(s)1
- No Clinical Signs, Symptoms or Conditions1
CoLag models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CoLag™ P65 ST862Screw, Fixation, Bone1 recall0 reports
- CoLag P06 S0004Orthopedic manual surgical instrument2 reports
Recent CoLag adverse event reports
The 2 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2025Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF A CANNULATED DRILL BROKE OFF WHILE DRILLING. THE SURGEON WAS UNABLE TO FULLY REMOVE THE DRILL AND THE PATIENT'S MEDIAL MALLEOULUS BROKE WHILE ATTEMPTING TO REMOVE THE DRILL. THE SURGERY WAS ABLE TO BE COMPLETED SUCCESSFULLY WITH K-WIRES. THE DHR REVIEW CONFIRMED THAT THE SINGLE-USE, STERILE DRILL WAS MANUFACTURED TO SPECIFICATIONS.
Patient problems: Bone Fracture(s)
MAUDE report 21644151 ↗MalfunctionJan 10, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A MIDFOOT FUSION PROCEDURE, THE 2.7 MM CANNULATED DRILL WAS PLACED OVER THE 0.062 " X 6 " GUIDE WIRE FROM THE SINGLE-USE INSTRUMENT KIT. THE DISTAL PORTION OF THE DRILL AND WIRE BROKE OFF AND WERE LEFT INSIDE THE PATIENT. THEY WERE UNABLE TO BE RETRIEVED. DEVICE HISTORY RECORDS WERE REVIEWED, AND THEY SHOWED THE DEVICES WERE MANUFACTURED TO SPECIFICATIONS. THE INSTRUMENTATION WAS DISCARDED POST-OP, SO THE PRODUCT WAS NOT AVAILABLE FOR EVALUATION. THE SURGEON STATED THERE WAS NO INTERFERENCE OR ANOMALIES FELT DURING THE DRILLING OR USE OF THE WIRE. AT THE TIME OF THE INCIDENT, NO OTHER HARDWARE WAS IN THE PATIENT. IT WAS ESTIMATED BY THE SURGEON THE DRILL WAS ABOUT 1 CM DEEP IN THE BONE AT THE TIME OF THE FRACTURE. NO KNOWN PATIENT ADVERSE EFFECTS HAVE BEEN REPORTED TO DATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13212646 ↗
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page