The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00810021866086), received between Jan 10, 2022 and Mar 19, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Bone Fracture(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF A CANNULATED DRILL BROKE OFF WHILE DRILLING. THE SURGEON WAS UNABLE TO FULLY REMOVE THE DRILL AND THE PATIENT'S MEDIAL MALLEOULUS BROKE WHILE ATTEMPTING TO REMOVE THE DRILL. THE SURGERY WAS ABLE TO BE COMPLETED SUCCESSFULLY WITH K-WIRES. THE DHR REVIEW CONFIRMED THAT THE SINGLE-USE, STERILE DRILL WAS MANUFACTURED TO SPECIFICATIONS.

    Patient problems: Bone Fracture(s)

    MAUDE report 21644151 ↗
  • MalfunctionJan 10, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A MIDFOOT FUSION PROCEDURE, THE 2.7 MM CANNULATED DRILL WAS PLACED OVER THE 0.062 " X 6 " GUIDE WIRE FROM THE SINGLE-USE INSTRUMENT KIT. THE DISTAL PORTION OF THE DRILL AND WIRE BROKE OFF AND WERE LEFT INSIDE THE PATIENT. THEY WERE UNABLE TO BE RETRIEVED. DEVICE HISTORY RECORDS WERE REVIEWED, AND THEY SHOWED THE DEVICES WERE MANUFACTURED TO SPECIFICATIONS. THE INSTRUMENTATION WAS DISCARDED POST-OP, SO THE PRODUCT WAS NOT AVAILABLE FOR EVALUATION. THE SURGEON STATED THERE WAS NO INTERFERENCE OR ANOMALIES FELT DURING THE DRILLING OR USE OF THE WIRE. AT THE TIME OF THE INCIDENT, NO OTHER HARDWARE WAS IN THE PATIENT. IT WAS ESTIMATED BY THE SURGEON THE DRILL WAS ABOUT 1 CM DEEP IN THE BONE AT THE TIME OF THE FRACTURE. NO KNOWN PATIENT ADVERSE EFFECTS HAVE BEEN REPORTED TO DATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions