Skip to content

CoLag recallZ-0157-2027 · In2bones USA, LLC · initiated Sep 9, 2026

Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from In2bones USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to labeling error, bone screws specified as fully threaded, are actually partially threaded screws.

FDA root cause category: Error in labeling

What the company is doing

On September 9, 2026, In2Bones, a CONMED Corporation, issued a Urgent: Medical Device Recall Notification via email. In2Bones ask consignees to take the following actions: 1. Please review, immediately identify and quarantine all affected devices within your facility. Discontinue use and distribution of affected devices. 2. Share this notice with all departments within your facility and all other facilities that may have received affected products from you. 3. Please complete the medical device recall response form and return it with the devices to: CONMED Corporation, RGA#: RGA-RA-CS-XXX,7750A Trinity Road, Cordova, TN 38018, Attn: Cameron Ealy, Return via: UPS Account #: 9A0303 (no charge to your facility) 3a. If you DO NOT HAVE any affected devices to return, please complete the medical device recall response form, indicating you have no devices and return by email to: [email protected] 4. Please do not return open or used devices.

Recall details

Recall number
Z-0157-2027
Recall class
Not yet classified by the FDA
Status
Open, Classified
Initiated by the company
Sep 9, 2026
Posted by the FDA
Oct 14, 2026
Recalling firm
In2bones USA, LLC
Quantity
29 units
Product code
HWC
FDA submissions
K170518
Distribution
US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.

Product description

REF: P65 ST862 CoLag SCREW, FT, 6.7X62.5 mm

Code information (models, lots, serial numbers)

UDI: 00817906023992/Lot: GM008380; GM007868

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K170518 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page