CONTOUR NEXTRecalls & reported problems across all models
Ascensia Diabetes Care US Inc. · 13 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 971
- Deaths reported
- 0
- Injuries reported
- 143
- Recalls
- 0
We found 13 models of CONTOUR NEXT by Ascensia Diabetes Care US Inc. in the FDA's device registry (GUDID). Together they are named in 971 adverse event reports in the FDA's MAUDE database, received between Jun 2, 2017 and Sep 25, 2026 — classified as 143 injuries, 828 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (402 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CONTOUR NEXT problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings402
- High Readings294
- Missing Test Results128
- Low Readings87
- Unsealed Device Packaging31
- Device Operates Differently Than Expected5
- Missing Information5
- No Device Output4
Patient problems
- No Known Impact Or Consequence To Patient450
- No Clinical Signs, Symptoms or Conditions312
- Hypoglycemia83
- Dizziness53
- Loss of consciousness48
- Shaking/Tremors33
- Sweating32
- Confusion/ Disorientation23
CONTOUR NEXT models
13 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CONTOUR® NEXT 7312Artificial Pancreas Device System, Threshold Suspend283 reports
- CONTOUR® NEXT 7314Single (Specified) Analyte Controls (Assayed And Unassayed)227 reports
- CONTOUR® NEXT 7311Glucose Dehydrogenase, Glucose101 reports
- CONTOUR® NEXT 7377Single (Specified) Analyte Controls (Assayed And Unassayed)95 reports
- CONTOUR® NEXT 7308Glucose Dehydrogenase, Glucose93 reports
- CONTOUR® NEXT 9697Single (Specified) Analyte Controls (Assayed And Unassayed)56 reports
- CONTOUR® NEXT 7309Artificial Pancreas Device System, Threshold Suspend37 reports
- CONTOUR® NEXT 7278Glucose Dehydrogenase, Glucose24 reports
- CONTOUR® NEXT 7313Glucose Dehydrogenase, Glucose21 reports
- CONTOUR® NEXT 7277Glucose Dehydrogenase, Glucose15 reports
- CONTOUR® NEXT 7310Glucose Dehydrogenase, Glucose9 reports
- CONTOUR® NEXT 7379Single (Specified) Analyte Controls (Assayed And Unassayed)6 reports
- CONTOUR® NEXT 7315Single (Specified) Analyte Controls (Assayed And Unassayed)4 reports
Recent CONTOUR NEXT adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Low Readings
The customer reported obtaining the following blood glucose results using a Contour® Next Link meter: 151 mg/dL at 5:38 p.m., 150 mg/dL at 5:50 p.m., 130 mg/dL at 6:08 p.m., and 93 mg/dL at 7:00 p.m. The customer also reported obtaining blood glucose results using a Contour® Next One meter of 107 mg/dL at 6:25 p.m., 92 mg/dL at 6:55 p.m., and 129 mg/dL at 7:08 p.m. The customer reported using both Contour® Next Link and Contour® Next One meters with Contour® Next test strips from lots 4JHEG01A and 4KHEG01C. The customer was unable to determine which test strip lot or meter-strip combination was associated with any individual blood glucose result. Based on the reported glucose readings, the customer administered 4 units of insulin. The customer subsequently reported symptoms attributed to hyperglycemia, described as feeling "out of his mind," disoriented, and collapsing. The customer was admitted to a hospital for two days and was treated with insulin and his usual medications. Additional medication details and dosages were not provided. Following treatment, the customer recovered and was discharged from the hospital.The customer was advised to return the device for evaluation.Repla
Patient problems: Confusion/ Disorientation; Hyperglycemia
MAUDE report 50898500 ↗InjurySep 25, 2026Low Readings
Patient problems: Confusion/ Disorientation; Hyperglycemia
MAUDE report 50898428 ↗MalfunctionApr 10, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT GEN METER. DURING TROUBLESHOOTING, A CONTROL TEST WAS RUN AND A READING OF 196 MG/DL WAS OBTAINED. THE METER DID NOT AUTOMATICALLY MARK IT AS A CONTROL TEST, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER¿S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24845330 ↗MalfunctionFeb 2, 2026Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER CONTACTED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT GEN METER, AS HE PERFORMED A CONTROL TEST AND THE READING WAS AUTOMATICALLY MARKED AS A BLOOD RESULT IN THE METER'S MEMORY. SINCE THE CUSTOMER USED INCORRECT TESTING TECHNIQUE, DURING TROUBLESHOOTING, ANOTHER CONTROL TEST WAS PERFORMED USING THE CORRECT TESTING TECHNIQUE AND A READING OF 130 MG/DL WAS OBTAINED. THE METER DID NOT AUTOMATICALLY MARK IT AS A CONTROL RESULT, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24236185 ↗MalfunctionDec 12, 2025Missing Test Results; Missing Test Results
THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT ONE OF THE CUSTOMER'S BLOOD GLUCOSE READINGS WAS NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23791082 ↗InjuryOct 17, 2025High Readings; High Readings
THE CUSTOMER RETURNED THE SUSPECTED CONTOUR® NEXT EZ METER FOR EVALUATION. A REVIEW OF THE METER'S LOGBOOK SHOWED THE FOLLOWING BLOOD GLUCOSE READINGS RECORDED ON THE DATE OF THE EVENT (B)(6) 2025): 171 MG/DL AT 11:27 A.M., 115 MG/DL AT 12:25 P.M., 43 MG/DL AT 1:37 P.M., 110 MG/DL AT 7:59 P.M., 86 MG/DL AT 9:08 P.M., AND 60 MG/DL AT 11:29 P.M.
Patient problems: Dizziness
MAUDE report 23324187 ↗MalfunctionSep 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT HE PERFORMED A CONTROL TEST WITH THE CONTOUR® NEXT GEN METER AND OBTAINED A READING OF 176 MG/DL AT 9:38 A.M. THE METER DID NOT AUTOMATICALLY MARK IT AS A CONTROL TEST, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038438 ↗MalfunctionSep 12, 2025Unsealed Device Packaging; Unsealed Device Packaging
THE CUSTOMER OPENED A BOX OF CONTOUR® NEXT TEST STRIPS AND FOUND THAT THE BOTTLE WAS ALREADY OPEN. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038436 ↗MalfunctionSep 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT GEN METER. DURING TROUBLESHOOTING, A CONTROL TEST WAS RUN AND A READING OF 186 MG/DL AT 4:34 P.M. WAS OBTAINED. THE METER DID NOT AUTOMATICALLY MARK IT AS A CONTROL TEST, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER REFUSED TO RETURN THE SUSPECTED DEVICE FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23038435 ↗MalfunctionSep 10, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT HE PERFORMED CONTROL TESTS WITH THE CONTOUR® NEXT ONE METER AND OBTAINED READINGS OF 224 MG/DL AT 4:45 P.M. AND 204 MG/DL AT 4:46 P.M. THE METER DID NOT AUTOMATICALLY MARK THE READINGS AS CONTROL TESTS, AND SO THE READINGS WILL BE DISPLAYED AS BLOOD RESULTS WHEN ACCESSING THE METER'S MEMORY. DURING TROUBLESHOOTING, ADDITIONAL THREE CONTROL TESTS WERE PERFORMED AND READINGS OF 124 MG/DL AT 5:18 P.M., 124 MG/DL AT 5:22 P.M. AND 127 MG/DL AT 5:24 P.M. WERE OBTAINED, WHICH WERE PROPERLY MARKED AS CONTROL TESTS. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23016613 ↗MalfunctionSep 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT SHE PERFORMED A CONTROL TEST WITH THE CONTOUR® NEXT GEN METER AND OBTAINED A READING OF 205 MG/DL AT 12:56 P.M. THE METER DID NOT AUTOMATICALLY MARK IT AS A CONTROL TEST, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER REFUSED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22993008 ↗MalfunctionSep 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT SHE PERFORMED CONTROL TESTS WITH THE CONTOUR® NEXT ONE METER AND OBTAINED READINGS OF 182 MG/DL AT 11:07 A.M., 142 MG/DL AT 11:11 A.M. AND 206 MG/DL AT 11:19 A.M. THE METER DID NOT AUTOMATICALLY MARK THE READINGS OF 182 MG/DL AND 206 MG/DL AS CONTROL TESTS, AND SO THE READINGS WILL BE DISPLAYED AS BLOOD RESULTS WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22993000 ↗MalfunctionSep 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT GEN METER. DURING TROUBLESHOOTING, TWO CONTROL TESTS WERE RUN AND READINGS OF 158 MG/DL AND 338 MG/DL WERE OBTAINED. THE METER DID NOT AUTOMATICALLY MARK THE READING OF 338 MG/DL AS A CONTROL TEST, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22992999 ↗MalfunctionSep 8, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT SHE PERFORMED A CONTROL TEST WITH THE CONTOUR® NEXT GEN METER AND OBTAINED A READING OF 156 MG/DL AT 7:47 A.M. THE METER DID NOT AUTOMATICALLY MARK IT AS A CONTROL TEST, AND SO THE READING WILL BE DISPLAYED AS A BLOOD RESULT WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22992998 ↗MalfunctionAug 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED THAT SHE PERFORMED CONTROL TESTS WITH THE CONTOUR® NEXT ONE METER AND OBTAINED READINGS OF 79 MG/DL AT 5:15 A.M. AND 72 MG/DL AT 5:17 A.M. THE METER DID NOT AUTOMATICALLY MARK THEM AS CONTROL TESTS, AND SO THE READINGS WILL BE DISPLAYED AS BLOOD RESULTS WHEN ACCESSING THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22796328 ↗
Other Ascensia Diabetes Care US Inc. products
All Ascensia Diabetes Care US Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page