CONTOUR® NEXT 9697Recalls & reported problems

Ascensia Diabetes Care US Inc. · Blood Glucose Monitoring System. 1 kit. For self-testing, monitoring and management of diabetes. This kit contains the following items: CONTOUR® Next Blood Glucose Meter; 10 CONTOUR® NEXT Blood Glucose Test Strips; User Guide; Quick Reference Guide; Lancing Device; 10 Colored Lancets (provided sterile); Logbook; Carrying Case; CONTOUR® NEXT Control Solution; Warranty Card. · Class 1

All models of CONTOUR NEXT (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Ascensia Contour Next 9697 kit is a home blood glucose meter sold with test strips, lancets and control solution for people managing diabetes. Most of the 56 reports describe readings not being saved in the meter, especially after a battery change, along with some high, low or inaccurate results. Two reports describe injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 00301939697010), received between Dec 6, 2017 and Feb 4, 2025 — classified as 2 injuries, 54 malfunctions.

The most frequently reported specific device problem is “Missing Test Results” (43 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Missing Test Results43
  • High Readings7
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Low Readings2
  • No Display/Image1
  • Nonstandard Device1

Patient problems

  • No Known Impact Or Consequence To Patient30
  • No Clinical Signs, Symptoms or Conditions23
  • Diarrhea1
  • Dizziness1
  • Head Injury1
  • Hyperglycemia1
  • Loss of consciousness1
  • Seizures1

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THE CUSTOMER HAD CHANGED THE METER'S BATTERIES 3 TO 4 DAYS BEFORE THE EVENT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21305625 ↗
  • MalfunctionJun 3, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH HIS CONTOUR NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT ONE OF HIS BLOOD GLUCOSE READINGS WAS NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19447952 ↗
  • MalfunctionApr 15, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THEIR CONTOUR NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT THEIR BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19109616 ↗
  • MalfunctionApr 15, 2024Missing Test Results; Missing Test Results

    AN ADVOCATE CALLED ON BEHALF OF THE CUSTOMER TO SEEK ASSISTANCE WITH THEIR CONTOUR NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT THEIR BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19105217 ↗
  • MalfunctionSep 1, 2023High Readings; High Readings

    THE CUSTOMER REPORTED THAT WITHIN 10 MINUTES SHE OBTAINED BLOOD GLUCOSE READINGS OF OVER 400 MG/DL AND 109 MG/DL WITH THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17670651 ↗
  • MalfunctionDec 2, 2022Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15905089 ↗
  • MalfunctionNov 2, 2022Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THE CUSTOMER ALSO STATED THAT THE YEAR ON THE DATE ASSOCIATED WITH THE READINGS SWITCHED FROM 2021 TO 2011 TO 2022. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15713889 ↗
  • MalfunctionNov 1, 2022Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT AFTER SHE CHANGED THE BATTERY ON THE CONTOUR NEXT METER, THE TIME AND DATE ON THE METER WERE SET INCORRECTLY AND HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15710555 ↗
  • MalfunctionFeb 10, 2022Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED TO REPORT THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13505614 ↗
  • MalfunctionOct 1, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. DURING TROUBLESHOOTING, ANOTHER TESTING WAS PERFORMED, WHICH WAS NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12561795 ↗
  • MalfunctionSep 7, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. DURING TROUBLESHOOTING, IT WAS FOUND THAT THE DATE WAS INCORRECTLY SET ON THE METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12433003 ↗
  • MalfunctionAug 2, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12259221 ↗
  • MalfunctionJul 1, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12098734 ↗
  • MalfunctionJun 3, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11929468 ↗
  • MalfunctionApr 21, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11701168 ↗
  • MalfunctionApr 21, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11701159 ↗
  • MalfunctionApr 13, 2021Nonstandard Device; Nonstandard Device

    THE CUSTOMER FROM CANADA REPORTED THAT HER CONTOUR NEXT ONE METER WAS READING IN MG/DL INSTEAD OF MMOL/L. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11655605 ↗
  • MalfunctionMar 3, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11405013 ↗
  • MalfunctionMar 3, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11405010 ↗
  • MalfunctionFeb 3, 2021Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11273738 ↗
  • MalfunctionDec 11, 2020Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT ONE OF HIS BLOOD GLUCOSE READINGS WAS NOT STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10992750 ↗
  • MalfunctionNov 3, 2020Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10777302 ↗
  • MalfunctionSep 18, 2020Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS BETWEEN (B)(6) 2020 TO (B)(6) 2020 WERE NOT STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10553832 ↗
  • MalfunctionSep 4, 2020Missing Test Results; Missing Test Results

    AN ADVOCATE REPORTED THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS SINCE (B)(6) 2020 WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10495130 ↗
  • MalfunctionSep 4, 2020Missing Test Results; Missing Test Results

    AN ADVOCATE REPORTED THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS FROM (B)(6) 2020 WERE NOT STORED IN THE MEMORY OF THE CONTOUR NEXT METER. DURING THE CALL, THE CUSTOMER PERFORMED A BLOOD GLUCOSE TEST AND THE READING WAS PROPERLY STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10495129 ↗

Other Single (Specified) Analyte Controls (Assayed And Unassayed) devices

All Single (Specified) Analyte Controls (Assayed And Unassayed) devices →