CONTOUR® NEXT 7377Recalls & reported problems

Ascensia Diabetes Care US Inc. · Blood Glucose Monitoring System. 1 kit. For self-testing, monitoring and management of blood glucose. In this kit you will find: CONTOUR® NEXT Blood Glucose Meter; User Guide and Quick Reference Guide; Lancing Device; 10 Colored Lancets (provided sterile); Logbook and Carrying Case; Warranty Card. · Class 1

All models of CONTOUR NEXT (13)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ascensia Contour Next is a home blood glucose meter for checking blood sugar. Its 95 reports, from 2018 to 2025, mostly describe readings that were not saved in the meter's memory, sometimes after a battery change, and some describe incorrect or high readings. Several reports say a replacement meter kit was sent. The reports include 7 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
95
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 95 adverse event reports that list this exact device (DI 00301937377013), received between Jan 8, 2018 and Dec 12, 2025 — classified as 7 injuries, 88 malfunctions.

The most frequently reported specific device problem is “Missing Test Results” (78 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Missing Test Results78
  • Incorrect, Inadequate or Imprecise Result or Readings10
  • High Readings6
  • No Device Output1

Patient problems

  • No Clinical Signs, Symptoms or Conditions52
  • No Known Impact Or Consequence To Patient35
  • Hypoglycemia3
  • Dizziness2
  • Hyperglycemia2
  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 95 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT ONE OF THE CUSTOMER'S BLOOD GLUCOSE READINGS WAS NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23791082 ↗
  • MalfunctionJun 9, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT AFTER REPLACING THE BATTERY OF THE CONTOUR® NEXT METER, HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22168986 ↗
  • MalfunctionJun 3, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT AFTER REPLACING THE BATTERY OF THE CONTOUR® NEXT METER, HIS READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22133917 ↗
  • MalfunctionApr 3, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH THE CONTOUR® NEXT METER. DURING THE TROUBLESHOOTING, IT WAS FOUND THAT THE CUSTOMER'S BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21768124 ↗
  • MalfunctionMar 13, 2025Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR® NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21596233 ↗
  • MalfunctionSep 16, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION, AS THE CUSTOMER THREW IT AWAY. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20226712 ↗
  • MalfunctionJul 1, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED ASCENSIA CUSTOMER SERVICE TO SEEK ASSISTANCE WITH HER CONTOUR NEXT METER. DURING TROUBLESHOOTING, IT WAS FOUND THAT ONE OF THE CUSTOMER'S BLOOD GLUCOSE READINGS WAS NOT STORED IN THE METER'S MEMORY. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643432 ↗
  • MalfunctionJun 20, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THE DATE AND TIME SET ON THE METER WERE INCORRECT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19576319 ↗
  • MalfunctionMar 25, 2024Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THE DATE AND TIME SET ON THE METER WERE INCORRECT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18967892 ↗
  • MalfunctionJan 18, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT HE RAN THREE BLOOD GLUCOSE TESTS WITH THE CONTOUR NEXT METER AND EACH READING WAS DIFFERENT BY OVER 100 MG/DL. NO SPECIFIC READINGS WERE PROVIDED. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18536505 ↗
  • MalfunctionSep 29, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17842531 ↗
  • InjuryAug 15, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT HE TOOK MORE THAN USUAL INSULIN BASED ON THE BLOOD GLUCOSE RESULT OBTAINED ON THE CONTOUR NEXT METER AND HAD TO GO TO THE HOSPITAL. NO FURTHER INFORMATION WAS PROVIDED BY THE CUSTOMER. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION.A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 17548369 ↗
  • MalfunctionAug 7, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17470432 ↗
  • MalfunctionMay 15, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16930876 ↗
  • InjuryMay 15, 2023No Device Output; No Device Output

    THE CUSTOMER REPORTED THAT SHE WAS UNABLE TO PERFORM BLOOD GLUCOSE TEST WITH THE CONTOUR NEXT METER AS THE METER WAS DEAD. SHE WAS EXPERIENCING SYMPTOMS OF HYPERGLYCEMIA WHICH WERE DESCRIBED AS BEING DIZZY AND THIRSTY. THE CUSTOMER WENT TO A PHYSICIAN'S OFFICE WHERE A BLOOD GLUCOSE TEST WAS PERFORMED AND A READING OF 175 MG/DL WAS OBTAINED. THE CUSTOMER WAS GIVEN AN INJECTION (UNKNOWN MEDICATION). AFTER, SHE WENT HOME AND RAN ANOTHER BLOOD GLUCOSE TEST AND OBTAINED A READING OF 139 MG/DL. THE CUSTOMER RECOVERED WELL AFTER THE TREATMENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: Dizziness; Hyperglycemia

    MAUDE report 16930872 ↗
  • MalfunctionMay 9, 2023Missing Test Results; Missing Test Results

    AN ADVOCATE CALLED ON BEHALF OF HER HUSBAND TO REPORT THAT THEIR BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER AFTER SHE PLACED THE METER THROUGH AN X-RAY SCANNER IN THE AIRPORT. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16894591 ↗
  • MalfunctionMay 9, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER CALLED TO REPORT THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16894590 ↗
  • MalfunctionApr 18, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16761457 ↗
  • MalfunctionMar 20, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT ONE OF HIS BLOOD GLUCOSE READINGS WAS NOT STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16577188 ↗
  • MalfunctionFeb 6, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT ONE OF HER BLOOD GLUCOSE READINGS WAS NOT STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16304668 ↗
  • InjuryFeb 6, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED THAT SHE WAS OBTAINING INACCURATE READINGS WITH THE CONTOUR NEXT METER. THE CUSTOMER STATED THAT THE READINGS WERE NOT IN ALIGNMENT WITH HER SYMPTOMS. FOR EXAMPLE, IF THE METER GAVE HIGH READINGS, THE CUSTOMER WAS EXPERIENCING SYMPTOMS OF HYPOGLYCEMIA, WHILE WHEN THE METER GAVE LOW READINGS, SHE WAS EXPERIENCING SYMPTOMS OF HYPERGLYCEMIA. NO SPECIFIC READINGS WERE PROVIDED. THE CUSTOMER STATED THAT SHE SOUGHT MEDICAL ATTENTION. NO FURTHER DETAILS WERE PROVIDED. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION. THE METER IS NO EXPECTED TO BE RETURNED FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: Hyperglycemia; Hypoglycemia

    MAUDE report 16304664 ↗
  • MalfunctionFeb 1, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16277667 ↗
  • MalfunctionJan 23, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT ONE OF HER BLOOD GLUCOSE READINGS WAS NOT STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16224583 ↗
  • MalfunctionJan 11, 2023Missing Test Results; Missing Test Results

    THE CUSTOMER REPORTED THAT HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR NEXT METER. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. A REPLACEMENT METER KIT WAS SENT TO THE CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16137302 ↗
  • MalfunctionNov 17, 2022High Readings; High Readings

    AN ADVOCATE CALLED ON BEHALF OF THE CUSTOMER TO REPORT THAT THE CUSTOMER TOOK 16 UNITS OF INSULIN. THE CUSTOMER WAS FEELING WEAK AND ATE 2 BARS OF CHOCOLATE AND FELT SAME AFTERWARDS. THE CUSTOMER WENT TO THE PHYSICIAN'S OFFICE WHERE HE OBTAINED BLOOD GLUCOSE READINGS OF 83 MG/DL WITH THE CONTOUR NEXT METER AND 52 MG/DL WITH THE LABORATORY TESTING WITHIN 6 MINUTES. EVEN THOUGH, A LOW READING WAS OBTAINED WITH THE LABORATORY TESTING, THE PHYSICIAN ADVISED THE CUSTOMER TO INCREASE THE INSULIN DOSAGE BASED ON METER'S READINGS. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE ADVOCATE WAS ADVISED TO RETURN THE DEVICE FOR EVALUATION. REPLACEMENT METER KIT AND TEST STRIPS WERE SENT TO THE CUSTOMER. SINCE THE STRIP INFORMATION WAS NOT PROVIDED, THIS REPORT WILL BE SUBMITTED UNDER THE METER INFORMATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 15813191 ↗

FDA clearance and classification

Cleared through Special 510(k) K191286 on Nov 8, 2019 (decision: Substantially Equivalent). Submitted by Ascensia Diabetes Care as “CONTOUR® NEXT Blood Glucose Monitoring System, CONTOUR® NEXT USB Blood Glucose Monitoring System, CONTOUR® NEXT ONE Blood Glucose Monitoring System, CONTOUR® NEXT EZ Blood Glucose Monitoring System, CONTOUR® NEXT LINK Wireless Blood Glucose Monitoring System”. The FDA review took 179 days. FDA record ↗

Cleared through Special 510(k) K160430 on Jun 24, 2016 (decision: Substantially Equivalent). Submitted by Ascensia Diabetes Care U.S., Inc. as “Contour Next Link Wireless Blood Glucose Monitoring System, Contour Next Blood Glucose Monitoring System”. The FDA review took 129 days. FDA record ↗

Cleared through Special 510(k) K121190 on Jul 26, 2012 (decision: Substantially Equivalent). Submitted by Bayer Healthcare, LLC as “Contour Next Blood Glucose Meter”. The FDA review took 98 days. FDA record ↗

The FDA regulates product code JJX (single (specified) analyte controls (assayed and unassayed)) under 21 CFR 862.1660 as a Class I device. 21 CFR 862.1660 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CONTOUR® NEXT 7377

Has the CONTOUR® NEXT 7377 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00301937377013), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CONTOUR® NEXT 7377?

The FDA's MAUDE database lists 95 adverse event reports for this device, received between Jan 8, 2018 and Dec 12, 2025. In those reports, the most common specific device problems are “Missing Test Results” (78), “Incorrect, Inadequate or Imprecise Result or Readings” (10) and “High Readings” (6); the most common patient problems are “Hypoglycemia” (3), “Dizziness” (2) and “Hyperglycemia” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CONTOUR® NEXT 7377?

None of the reports are classified as deaths, 7 as injuries and 88 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CONTOUR® NEXT 7377 FDA approved?

It was cleared through Special 510(k) K121190 on Jul 26, 2012; its record lists 3 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the CONTOUR® NEXT 7377, and is it still sold?

The FDA's device database lists Ascensia Diabetes Care US Inc. as the company responsible for it (brand CONTOUR® NEXT, model 7377). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Single (Specified) Analyte Controls (Assayed And Unassayed) devices

All Single (Specified) Analyte Controls (Assayed And Unassayed) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page