CROSSERRecalls & reported problems across all models
Bard Peripheral Vascular, Inc. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 160
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls
- 0
We found 6 models of CROSSER by Bard Peripheral Vascular, Inc. in the FDA's device registry (GUDID). Together they are named in 160 adverse event reports in the FDA's MAUDE database, received between Jun 23, 2016 and Dec 1, 2025 — classified as 9 injuries, 151 malfunctions.
The most frequently reported specific device problem is “Material Separation” (65 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CROSSER problems
Device problems
- Material Separation65
- Detachment of Device or Device Component57
- Device-Device Incompatibility39
- Break27
- Material Twisted/Bent25
- Fracture11
- Material Puncture/Hole11
- Retraction Problem11
Patient problems
- No Consequences Or Impact To Patient82
- No Clinical Signs, Symptoms or Conditions41
- No Patient Involvement34
- Injury4
- Foreign Body In Patient3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Blood Loss1
- Perforation of Vessels1
CROSSER models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CROSSER® CRUO14SACatheter for crossing total occlusions60 reports
- CROSSER® CRUS6ACatheter for crossing total occlusions47 reports
- CROSSER® CRU14SACatheter for crossing total occlusions20 reports
- CROSSER® CRUOS106Catheter for crossing total occlusions18 reports
- CROSSER® CRUS6106Catheter for crossing total occlusions12 reports
- CROSSER® CRUS106Catheter for crossing total occlusions3 reports
Recent CROSSER adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER, THE WIRE PORT ALLEGEDLY DETACHED FROM CATHETER. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23678676 ↗MalfunctionNov 17, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER. DURING THE PROCEDURE, THE WIRE PORT ALLEGEDLY DETACHED FROM CATHETER. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23566003 ↗MalfunctionNov 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER. DURING THE PROCEDURE, THE WIRE PORT ALLEGEDLY DETACHED FROM CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23554107 ↗MalfunctionNov 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER. DURING THE PROCEDURE, THE WIRE PORT ALLEGEDLY DETACHED FROM CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23553249 ↗MalfunctionNov 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER. DURING THE PREPARATION OF THE PROCEDURE, THE WIRE PORT ALLEGEDLY DETACHED FROM CATHETER. IT WAS FURTHER REPORTED THAT THERE WAS NO DAMAGE NOTED ON THE BOX. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23551735 ↗MalfunctionNov 14, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ON (B)(6) 2025, A PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CROSSER CATHETER. DURING THE PROCEDURE, THE WIRE PORT ALLEGEDLY DETACHED FROM CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23551257 ↗MalfunctionAug 25, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ON (B)(6) 2025, A FEMALE PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CTO CROSSER CATHETER. IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, WHILE FLUSHING THE WIRE PORT, THE VALVE WAS ALLEGEDLY COMPLETELY BROKEN OFF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22881929 ↗MalfunctionAug 25, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ON (B)(6) 2025, A FEMALE PATIENT UNDERWENT A RECANALIZATION PROCEDURE USING THE CTO CROSSER CATHETER. IT WAS REPORTED THAT PRIOR TO THE PROCEDURE, WHILE FLUSHING THE WIRE PORT, THE VALVE WAS ALLEGEDLY COMPLETELY BROKEN OFF. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22881926 ↗MalfunctionMar 17, 2025Material Separation; Material Twisted/Bent; Activation Problem; Material Separation; Material Twisted/Bent; Activation Problem
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE SUPERFICIAL FEMORAL ARTERY, THE DEVICE ALLEGEDLY HAD A CHIRP VIBRATION. THE PROCEDURE WAS COMPLETED TREAT WITH ANGIOPLASTY AND STENT. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21615734 ↗MalfunctionJan 24, 2025Material Separation; Device-Device Incompatibility; Material Separation; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE CATHETER WAS ALLEGEDLY STICK TO THE WIRE. IT WAS FURTHER REPORTED THAT THE WIRE ACCESS WAS LOST WHILE TRYING TO RETRIEVE THE DEVICE. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21231529 ↗MalfunctionDec 13, 2023Break; Difficult to Remove; Device-Device Incompatibility; Activation Problem; Break; Difficult to Remove; Device-Device Incompatibility; Activation Problem
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE CATHETER ALLEGEDLY BROKE. IT WAS FURTHER REPORTED THAT THE TIP OF THE DEVICE ALLEGEDLY STOPPED WORKING AND THE CATHETER ALLEGEDLY GOT STUCK IN THE SUPPORT CATHETER. REPORTEDLY BOTH THE CATHETERS WERE REMOVED. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18316956 ↗MalfunctionAug 30, 2023Connection Problem; Detachment of Device or Device Component; Connection Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE, THE CATHETER ALLEGEDLY WOULD NOT CONNECT TO THE TRANSDUCER CABLE. IT WAS FURTHER REPORTED THAT WHEN THEY TRIED TO UNSCREW THE CATHETER FROM THE CABLE TO CATHETER HOUSING CAME APART EXPOSING THE WIRE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17648142 ↗MalfunctionJul 11, 2023Retraction Problem; Stretched; Detachment of Device or Device Component; Retraction Problem; Stretched; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE RECANALIZATION PROCEDURE, THE CATHETER TIP ALLEGEDLY BROKE WHEN THE DEVICE WAS BEING PULLED BACK OUT OF THE PATIENT. IT WAS FURTHER REPORTED THAT THE TIP OF THE CATHETER WAS UNABLE TO BE RETRIEVED AS THE VESSEL WAS COMPLETELY OCCLUDED WHERE THE TIP WAS BROKEN OFF. REPORTEDLY, THE CATHETER TIP WAS LEFT IN THE PATIENT. FURTHERMORE, NO ADDITIONAL TREATMENT WAS DONE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 17300213 ↗MalfunctionMay 30, 2023Break; Retraction Problem; Failure to Advance; Device-Device Incompatibility; Material Twisted/Bent; Break; Retraction Problem; Failure to Advance; Device-Device Incompatibility; Material Twisted/Bent
IT WAS REPORTED THAT DURING A RECANALIZATION PROCEDURE IN THE LEFT SUPERFICIAL ARTERY VIA RIGHT GROIN ACCESS, THE CATHETER ALLEGEDLY BECAME STUCK IN THE LESION. IT WAS FURTHER REPORTED THAT THE CATHETER WAS ALLEGEDLY COMPLETELY BROKEN AND STUCK INSIDE THE SUPPORT CATHETER. FURTHERMORE, THE CATHETER ALLEGEDLY HAD DIFFICULTY IN REMOVING FROM THE PATIENT. REPORTEDLY, THE PHYSICIAN WAS ABLE TO REMOVE THE DEVICE FROM THE PATIENT WITHOUT RIPPING THE VESSEL APART. THE CROSSER IS STILL STUCK INSIDE THE SUPPORT CATHETER. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17018301 ↗MalfunctionApr 4, 2023Break; Crack; Leak/Splash; Break; Crack; Leak/Splash
IT WAS REPORTED THAT PRIOR TO A RECANALIZATION PROCEDURE, THE CATHETER WAS ALLEGEDLY BROKEN WHEN IT WAS OPENED. IT WAS FURTHER REPORTED WHEN THE CATHETER WAS FLUSHED, THERE WAS SOME FLUID COMING OUT OF THE SIDE OF THE CATHETER JUST ABOUT AN INCH FROM THE PORT. REPORTEDLY UPON CLOSER OBSERVATION, THERE WAS A SMALL CRACK IN THE SIDE OF THE CATHETER THAT WAS LEAKING FLUID. THE PROCEDURE WAS COMPLETED BY USING ANOTHER DEVICE. THERE WAS NO PATIENT CONTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16672349 ↗
Other Bard Peripheral Vascular, Inc. products
- MAX-CORE11 models · 8,845 reports · 1 recall
- DENALI Vena Cava Filter2 models · 1,021 reports
- MAGNUM19 models · 571 reports · 1 recall
- Conquest32 models · 560 reports
- Ultraverse RX56 models · 536 reports
- Fluency Endovascular Stent Graft51 models · 532 reports
- Atlas Gold33 models · 427 reports
- Ultraverse 035 PTA Dilatation Catheter92 models · 393 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page