CurrentRecalls & reported problems across all models
St. Jude Medical, Inc. · 11 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 393
- Deaths reported
- 18
- Injuries reported
- 213
- Recalls
- 0
We found 11 models of Current by St. Jude Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 393 adverse event reports in the FDA's MAUDE database, received between Jun 11, 2018 and Sep 28, 2026 — classified as 18 deaths, 213 injuries, 162 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (68 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Current problems
Device problems
- Premature Discharge of Battery68
- Inappropriate or Unexpected Reset49
- Adverse Event Without Identified Device or Use Problem43
- Inappropriate/Inadequate Shock/Stimulation43
- Failure to Charge37
- Over-Sensing33
- Delayed Charge Time21
- Failure to Interrogate21
Patient problems
- No Known Impact Or Consequence To Patient185
- No Clinical Signs, Symptoms or Conditions112
- Therapeutic Effects, Unexpected35
- Death15
- No Information15
- No Consequences Or Impact To Patient11
- Unspecified Infection9
- Arrhythmia6
Current models
11 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Current™ CD1211-36QPulse generator, permanent, implantable4 deaths121 reports
- Current™ CD1211-36Pulse generator, permanent, implantable2 deaths76 reports
- Current® VR RFPulse generator, permanent, implantable2 deaths52 reports
- Current™ CD2211-36QPulse generator, permanent, implantable3 deaths52 reports
- Current™ CD2211-36Pulse generator, permanent, implantable2 deaths42 reports
- Current™ PFA-GEN-01Percutaneous Cardiac Ablation Catheter for Treatment of Atrial Fibrillation with Irreversible Electroporation22 reports
- Current® DR RFPulse generator, permanent, implantable4 deaths18 reports
- Current® 2107-36Pulse generator, permanent, implantable1 death5 reports
- Current® 1107-36Pulse generator, permanent, implantable3 reports
- Current® DR RFPulse generator, permanent, implantable1 report
- Current® VR RFPulse generator, permanent, implantable1 report
Recent Current adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Device Emits Odor; Noise, Audible
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917650 ↗MalfunctionSep 18, 2026Noise, Audible
This report is to advise of thermal damage identified during device evaluation. During a pulsed field ablation atrial fibrillation procedure, a loud bang was heard from the Current PFA generator, and a PFA generator hardware error appeared on the screen. It was noted that mapping of atrial tachycardia in the left atrium had begun, and the capacitors were charged and ready for PFA delivery, but no PFA was delivered when the loud bang occurred.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817779 ↗MalfunctionSep 17, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50808695 ↗MalfunctionSep 11, 2026Output Problem; Noise, Audible
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50739588 ↗MalfunctionAug 20, 2026Thermal Decomposition of Device
This report is to advise of thermal damage found during investigation.During a procedure with TactiFlex Duo catheter, the catheter was successfully connected to the Current generator and placed in the heart. The energy was charged to 100%. Before the first ablation, a loud sound was heard from the generator and an error message appeared which had a restart button. The exact message was not clear but it mentioned "low voltage detected". The generator was restarted two times with no resolution and another error appeared saying "Hardware issue detected". The generator was replaced and the procedure was completed with no adverse consequences to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50344456 ↗MalfunctionAug 13, 2026Thermal Decomposition of Device; Communication or Transmission Problem
This report is to advise on thermal damage found during analysis.During an AF procedure using Rf and PFA (TactiFlex Duo), it was noted that the following error message was displayed on the Current generator; "Hard drive issue". The TactiFlex Duo catheter was replaced but the issue persisted. The procedure was completed using the TactiCath. There were no patient consequences.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50236650 ↗MalfunctionJul 28, 2026Thermal Decomposition of Device
This report is to advise on thermal damage found during analysis. During an Endo-Epicardial VT procedure, both the TactiFlex Duo catheter and Advisor HD Grid catheter were used. Before ablation was started, it was made sure that everything was connected correctly. At that time, the Current PFA generator made a big sound and no longer functioned, despite it being restarted twice. The issue was resolved by using a different Current PFA generator to continue the case. The procedure was completed with no adverse consequences to the patient.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50060457 ↗MalfunctionJul 8, 2026Thermal Decomposition of Device
A REPORT IS NEEDED TO ADVISE OF THERMAL DAMAGE FOUND IN THE DEVICE DURING ANALYSIS. AFTER THE CASE, A LOUD POP SOUND CAME FROM THE CURRENT GENERATOR. AFTER, AN ERROR MESSAGE WAS DISPLAYED ON BOTH THE ENSITE X AND GENERATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25749210 ↗MalfunctionJun 10, 2026Thermal Decomposition of Device
THIS REPORT IS TO ADVISE OF THERMAL DAMAGE FOUND IN THE DEVICE DURING ANALYSIS. DURING A PROCEDURE, WHEN DELIVERING PFA APPLICATION A LOUD POP WAS HEARD FROM THE GENERATOR AND A GENERATOR FAULT MESSAGE DISPLAYED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25453098 ↗MalfunctionJun 9, 2026Mechanical Problem; Noise, Audible
DURING A PFA PROCEDURE, A "NOISE" WAS HEARD FROM THE CURRENT PFA GENERATOR AND A MESSAGE "VOLT GENERATOR HARDWARE ERROR. PLEASE REBOOT" WAS DISPLAYED ON THE ENSITE SYSTEM. TROUBLESHOOTING INCLUDED MULTIPLE REBOOTS OF THE GENERATOR WITH AND WITHOUT DEVICES OR CABLES CONNECTED, AS WELL AS TRYING DIFFERENT POWER OUTLETS, BUT THE ISSUE PERSISTED. IT WAS NOTED THAT THE STANDBY BUTTON WAS NOT ON DURING BOOTUP. THE ISSUE COULD NOT BE RESOLVED AND THE PROCEDURE WAS CANCELLED. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25448770 ↗MalfunctionMay 18, 2026Communication or Transmission Problem
THIS REPORT IS TO ADVISE OF THERMAL DECOMPOSITION THAT WAS NOTED DURING ANALYSIS. DURING A PFA PROCEDURE, WHEN CONNECTING THE TWO TOUCHSCREENS AND ADDING THE BOOM MONITOR VIA AN HDMI CABLE, IT WAS NOTED THAT THE SCREENS REMAINED BLACK. TYPICALLY, AFTER CONNECTING THE HDMI CABLE, THE TOUCHSCREENS WOULD GO BLACK FOR A SECOND BEFORE PROPERLY BOOTING UP. ON THE ENSITE X GUI, IT WAS SEEN THAT THE CURRENT PFA GENERATOR WAS STILL CONNECTED BUT, DESPITE MULTIPLE REBOOTS WITH DIFFERENT CABLE CONFIGURATIONS, NO IMAGES WERE ABLE TO POPULATE ON THE GENERATOR AFTER THE INITIAL ERROR. SINCE THE GENERATOR SCREEN WOULD NOT TURN ON, THE CASE WAS CONTINUED WITH A TACTIFLEX CATHETER. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25200556 ↗MalfunctionMay 15, 2026Mechanical Problem; Noise, Audible
DURING THE ATRIAL FIBRILLATION PFA PROCEDURE, ON THE 17TH PFA APPLICATION, THE CURRENT PFA GENERATOR PRODUCED A LOUD POPPING SOUND, AND A HARDWARE ERROR APPEARED ON THE MONITOR. THE SYSTEM PROMPTED US TO RESTART THE COMPUTER. HOWEVER, AFTER RESTARTING, A CASE COULD NOT BE STARTED, AND THE SYSTEM CONTINUOUSLY PROMPTED US TO REBOOT. EVEN AFTER DISCONNECTING ALL CABLES, AND REBOOTING THE SYSTEM AGAIN, THE ISSUE PERSISTED. PHYSICIAN DECIDED TO END THE PROCEDURE AND RE-DO IF THE ARRHYTHMIA RECUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25187208 ↗MalfunctionMay 9, 2026Thermal Decomposition of Device
THIS REPORT IS TO ADVISE OF THERMAL DAMAGE FOUND ON THE DEVICE DURING ANALYSIS. DURING AN ATRIAL FIBRILLATION PROCEDURE, WHEN USING THE VOLT 2.0 CATHETER IN THE LEFT ATRIUM, IT WAS NOTED THAT THE CURRENT PFA GENERATOR HAD RETURNED TO DEFAULT AFTER AN AUDIBLE SOUND WAS HEARD. SINCE A STUDY WAS UNABLE TO BE RAN ON THE CURRENT PFA GENERATOR AFTER, THE ISSUE WAS RESOLVED BY CHANGING TO A DIFFERENT SYSTEM TO CONTINUE THE CASE. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25132842 ↗MalfunctionMay 7, 2026Communication or Transmission Problem
DURING A DUAL PROCEDURE FOR AF, THE PROCEDURE WAS NOT COMPLETED. IT WAS NOTED THAT THE COOL POINT PUMP REMAINED AT 13 ML/MIN AFTER THE LESION. IN ORDER TO RESOLVE THE ISSUE, ANOTHER LESION WAS PERFORMED, WHICH STOPPED THE COOL POINT PUMP, AND THEN WHEN ANOTHER LESION WAS PERFORMED WITH THE CATHETER, THE PUMP STARTED FUNCTIONING AGAIN. THE ISSUE OCCURRED MULTIPLE TIMES THROUGHOUT THE PROCEDURE. ADDITIONALLY, IT WAS NOTED THAT AFTER ANOTHER LESION, THE MESSAGE "PEDAL DEPRESSED" WAS DISPLAYED, EVEN THOUGH THE PHYSICIAN'S FOOT WAS NOT ON IT. AFTER THIS, ENERGY WAS SWITCHED FROM RF TO PFA TO PERFORM THE NEXT LESION. THE PROCEDURE WAS NOT COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25115994 ↗MalfunctionApr 29, 2026Device Emits Odor; Noise, Audible
THIS REPORT IS TO ADVISE OF THERMAL DAMAGE THAT WAS NOTED DURING ANALYSIS. DURING THE PFA PROCEDURE WITHOUT DISCONNECTING OR MOVING ANYTHING, THE CURRENT PFA GENERATOR, SUDDENLY MADE A LOUD SOUND AND TURNED OFF. WHEN THIS HAPPENED, UPON TURNING IT BACK ON, THE SCREEN TO ACCESS THE CASE WAS GRAYED OUT AND INACCESSIBLE. THE GENERATOR WAS RESTARTED SEVERAL TIMES, CONNECTING AND DISCONNECTING EVERYTHING, BUT THE ISSUE WAS NOT RESOLVED. THE PROCEDURE WAS CONTINUED USING CRYOABLATION FROM A NON-ABBOTT SYSTEM WITHOUT ANY FURTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25028840 ↗
Other St. Jude Medical, Inc. products
- Tendril15 models · 49,498 reports
- Octrode10 models · 36,740 reports
- Proclaim12 models · 35,038 reports · 17 recalls
- Durata28 models · 19,805 reports
- Assurity MRI2 models · 17,838 reports · 1 recall
- Gallant8 models · 12,885 reports · 5 recalls
- Quartet10 models · 11,001 reports
- MitraClip19 models · 9,056 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.