EndurityRecalls & reported problems across all models
St. Jude Medical, Inc. · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names at least one Endurity model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 279
- Deaths reported
- 3
- Injuries reported
- 127
- Recalls
- 1
We found 2 models of Endurity by St. Jude Medical, Inc. in the FDA's device registry (GUDID). Together they are named in 279 adverse event reports in the FDA's MAUDE database, received between Jun 22, 2018 and Sep 17, 2026 — classified as 3 deaths, 127 injuries, 149 malfunctions.
The most frequently reported specific device problem is “Defective Device” (28 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Endurity recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 15, 2021Class IOpen, Classified
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
Z-1531-2021 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
Most reported Endurity problems
Device problems
- Adverse Event Without Identified Device or Use Problem74
- Defective Device28
- Premature Discharge of Battery27
- Over-Sensing19
- Incorrect Measurement18
- Failure to Interrogate16
- Signal Artifact/Noise16
- Failure to Disconnect15
Patient problems
- No Clinical Signs, Symptoms or Conditions137
- No Known Impact Or Consequence To Patient44
- Insufficient Information28
- Unspecified Infection28
- Pocket Erosion17
- Arrhythmia10
- Discomfort9
- No Consequences Or Impact To Patient8
Endurity models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Endurity™ PM2160Pulse generator, permanent, implantable1 recall2 deaths237 reports
- Endurity™ PM1160Pulse generator, permanent, implantable1 recall1 death42 reports
Recent Endurity adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Discomfort; Pocket Erosion; Unspecified Infection
MAUDE report 50797165 ↗MalfunctionSep 14, 2026No Pacing
It was reported that the patient presented remotely via Merlin.net. Upon review, the patient experienced episodes of bradycardia and fainting. The physician confirmed that the pacemaker was functioning perfectly; however, the intervention and patient condition were unknown.
Patient problems: Arrhythmia; Syncope/Fainting
MAUDE report 50750107 ↗InjurySep 11, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient presented for a post-operative follow-up. It was noted that the patient repeatedly developed pocket infections associated with redness, swelling, and ulceration. The wound remained open, with the pacemaker exposed, despite two debridement procedures. A small portion of heart tissue was observed on the lead. The physician believed that the infection was device and procedure related. The entire system, including the pacemaker, and the right ventricular lead, was explanted and replaced with a leadless pacemaker. The patient remained in stable condition.
Patient problems: Pocket Erosion; Unspecified Infection; Wound Dehiscence
MAUDE report 50725406 ↗InjurySep 1, 2026No Pacing
It was reported that the patient was hospitalized due to repeated fainting episodes. It was also noted that the Subsequent 24-hour Holter monitoring revealed pauses lasting more than 10 seconds. The pacemaker was explanted and replaced. The patient was in stable condition.
Patient problems: Syncope/Fainting
MAUDE report 50536396 ↗MalfunctionJun 4, 2026Unexpected Mode Switch
IT WAS REPORTED THAT THE PACEMAKER WAS FOUND TO HAVE REVERTED INTO BACKUP AAI MODE DURING CIQ TESTING. THE DEVICE WAS SUBSEQUENTLY RESTORED FROM BACKUP MODE WITH ASSISTANCE FROM TECHNICAL SERVICES (TS). THE PACEMAKER WAS ULTIMATELY NOT USED AND WAS DEACTIVATED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25402652 ↗MalfunctionMay 12, 2026No Apparent Adverse Event
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 25152320 ↗MalfunctionApr 17, 2026Loose or Intermittent Connection; Failure to Disconnect
IT WAS REPORTED THAT DURING A PROCEDURE, THE SET-SCREW OF THE PACEMAKER (PM) WAS STRIPPED. THE PHYSICIAN WAS UNABLE TO SUCCESSFULLY LOOSEN THE SET-SCREW TO REMOVE THE LEAD FROM THE HEADER OF THE PM. THE PM WAS EXPLANTED AND REPLACED ON (B)(6) 2026. THE PATIENT WAS STABLE THROUGHOUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24908859 ↗MalfunctionApr 16, 2026Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE PATIENT'S PACEMAKER EXHIBITED FAR R OVER-SENSING WITH P-WAVES OCCURRING WITHIN THE REFRACTORY PERIOD. NO INTERVENTION WAS PERFORMED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24892778 ↗MalfunctionApr 13, 2026Loose or Intermittent Connection; Failure to Disconnect
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A GENERATOR REPLACEMENT PROCEDURE DUE TO REACHING THE ELECTIVE REPLACEMENT INDICATOR (ERI). DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO CONNECT THE CHRONIC RIGHT ATRIAL (RA) LEAD TO THE PACEMAKER, AN UNSPECIFIED FITTING ISSUE WAS ENCOUNTERED IN THE PACEMAKER HEADER DURING THE SCREWING PROCESS. THE RA LEAD ALSO COULD NOT BE REMOVED FROM THE PACEMAKER HEADER. AS A RESULT, THE RA LEAD WAS CUT AND CAPPED, AND THE PACEMAKER WAS NOT USED. THE PHYSICIAN ELECTED TO USE A SINGLE-CHAMBER PACEMAKER TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24854700 ↗MalfunctionFeb 3, 2026Incorrect Measurement
DURING AN IN CLINIC FOLLOW UP IN (B)(6) 2025 THE DEVICE SHOWED 6 MONTHS OF BATTERY REMAINING. ANOTHER IN CLINIC FOLLOW UP WAS PERFORMED ON (B)(6) 2026 AND SHOWED 11.8 MONTHS OF BATTERY REMAINING. TECHNICAL SUPPORT WAS CONTACTED AND A DIAGNOSTIC ANOMALY WAS SUSPECTED AS THE 11.8 MONTH REMAINING BATTERY ESTIMATION WAS DEEMED CORRECT. FURTHER INVESTIGATION WAS RECOMMENDED BY TECHNICAL SUPPORT. NO INTERVENTION HAS BEEN PERFORMED. THE PATIENT WAS STABLE AND WILL CONTINUE BEING MONITORED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24254089 ↗InjuryJan 22, 2026Failure to Interrogate; No Pacing
IT WAS REPORTED THAT THE PATIENT HAD PRESENTED TO THE CLINIC FOR A ROUTINE FOLLOW-UP. IT WAS NOTED THAT THE PACEMAKER HAD FAILED TO BE INTERROGATED VIA INDUCTIVE TELEMETRY. THERE WAS NO LV OUTPUT, AND PACING COULD NOT BE INDUCED WITH MAGNET APPLICATION. THE PACEMAKER WAS SUBSEQUENTLY EXPLANTED AND REPLACED. THE PATIENT REMAINED IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24157770 ↗InjuryDec 29, 2025Failure to Interrogate; No Pacing; Failure to Interrogate; No Pacing
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR IN CLINIC FOLLOW-UP WITH VENTRICULAR BRADYCARDIA. DEVICE INTERROGATION WAS ATTEMPTED HOWEVER, THE PACEMAKER (PM) FAILED TO BE INTERROGATED. ON (B)(6) 2025, AN ELECTROCARDIOGRAM (ECG) WAS PERFORMED, NOTING THAT THE PM WAS NOT PROVIDING PACING. THE PHYSICIAN ELECTED TO EXPLANT AND REPLACE THE PM. THE PATIENT CONDITION WAS STABLE.
Patient problems: Arrhythmia
MAUDE report 23904098 ↗InjuryDec 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC WITH REDNESS, SWELLING, DISCHARGE DUE TO INFECTION AND EROSION AT THE DEVICE SITE. THE INFECTION WAS UNCLEAR EITHER IT IS ASSOCIATED WITH ANY ABBOTT PRODUCT AND/OR PROCEDURE. THE PACEMAKER AND RIGHT VENTRICULAR LEAD WERE EXPLANTED AND REPLACED. THE PATIENT WAS IN STABLE CONDITION THROUGHOUT THE PROCEDURE.
Patient problems: Pocket Erosion; Unspecified Infection
MAUDE report 23753797 ↗MalfunctionNov 21, 2025No Apparent Adverse Event; No Apparent Adverse Event
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 23623563 ↗MalfunctionNov 11, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT A DIAGNOSTICS ANOMALY WAS NOTED ON THE PATIENT'S PACEMAKER. ATRIAL FIBRILLATION EPISODES WERE NOT BEING RECORDED AS EXPECTED. NO INTERVENTION WAS PERFORMED. THE PATIENT CONDITION WAS NOT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523110 ↗
Other St. Jude Medical, Inc. products
- Tendril15 models · 49,498 reports
- Octrode10 models · 36,740 reports
- Proclaim12 models · 35,038 reports · 17 recalls
- Durata28 models · 19,805 reports
- Assurity MRI2 models · 17,838 reports · 1 recall
- Gallant8 models · 12,885 reports · 5 recalls
- Quartet10 models · 11,001 reports
- MitraClip19 models · 9,056 reports · 2 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.