FreeStyle Libre 2 PlusRecalls & reported problems across all models
Abbott Diabetes Care Inc · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3,192
- Deaths reported
- 0
- Injuries reported
- 145
- Recalls
- 0
We found 4 models of FreeStyle Libre 2 Plus by Abbott Diabetes Care Inc in the FDA's device registry (GUDID). Together they are named in 3,192 adverse event reports in the FDA's MAUDE database, received between Feb 20, 2024 and Sep 30, 2026 — classified as 145 injuries, 3,047 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2,735 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported FreeStyle Libre 2 Plus problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2,735
- Detachment of Device or Device Component183
- Low Readings113
- Device Alarm System78
- Unable to Obtain Readings28
- Device Displays Incorrect Message26
- High Readings25
- Adverse Event Without Identified Device or Use Problem15
Patient problems
- No Clinical Signs, Symptoms or Conditions2,878
- Device Embedded In Tissue or Plaque177
- Hyperglycemia46
- Loss of consciousness38
- Hypoglycemia36
- Dizziness24
- Insufficient Information18
- Fatigue9
FreeStyle Libre 2 Plus models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- FreeStyle Libre 2 Plus 78747Integrated Continuous Glucose Monitoring system, factory calibrated1,833 reports
- FreeStyle Libre 2 Plus 78747Integrated Continuous Glucose Monitoring system, factory calibrated1,337 reports
- FreeStyle Libre 2 Plus 78749Integrated Continuous Glucose Monitoring system, factory calibrated17 reports
- FreeStyle Libre 2 Plus 78749Integrated Continuous Glucose Monitoring system, factory calibrated5 reports
Recent FreeStyle Libre 2 Plus adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Detachment of Device or Device Component
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50953814 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 46 mg/dL compared to readings of 249 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949349 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50.00 mg/dL compared to readings of 210.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944676 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 66.00 mg/dL,68.00 mg/dL compared to readings of 172.00 mg/dL,202.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 8 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944671 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 103.00 mg/dL compared to readings of 249.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944336 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943513 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50.00 mg/dL and 58.00 mg/dL compared to readings of 258.00 mg/dL and 240.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935928 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 57 mg/dL compared to readings of 220 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931744 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL and 53 mg/dL compared to readings of 177 mg/dL and 107 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931740 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50930492 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 162 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929666 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 41 mg/dL compared to readings of 269 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 12 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929468 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 67 mg/dL compared to readings of 200 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929326 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 70.00 mg/dL compared to readings of 300.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929099 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929098 ↗
Other Abbott Diabetes Care Inc products
- FreeStyle Libre 314 models · 69,860 reports · 6 recalls
- FreeStyle Libre 3 Plus5 models · 52,808 reports · 2 recalls
- FreeStyle Libre 212 models · 30,420 reports · 2 recalls
- FreeStyle Libre 14 day5 models · 16,749 reports · 1 recall
- FreeStyle Lite9 models · 3,388 reports
- FreeStyle Freedom Lite5 models · 2,436 reports
- FreeStyle Libre7 models · 705 reports · 1 recall
- Precision Xtra10 models · 391 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page