FreeStyle Libre 2 Plus 78747 (UDI 00357599835002)Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Libre 2 Plus Sensor · Class 2
Quick summary
This Abbott FreeStyle Libre 2 Plus sensor is the subject of 1,029 adverse event reports, which mention no deaths and 46 injuries. Most reports describe glucose readings that were incorrect or imprecise, and a few mention alarms or readings that could not be obtained. Reports rose from 203 in the previous 12 months to 817 in the latest 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,337
- Deaths reported
- 0
- Injuries reported
- 46
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,337 adverse event reports that list this exact device (DI 00357599835002), received between Feb 20, 2024 and Sep 29, 2026 — classified as 46 injuries, 1,291 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1,204 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1,204
- Device Alarm System46
- Detachment of Device or Device Component30
- Low Readings25
- Unable to Obtain Readings12
- Device Displays Incorrect Message10
- Delivered as Unsterile Product7
- High Readings7
Patient problems
- No Clinical Signs, Symptoms or Conditions1,263
- Device Embedded In Tissue or Plaque29
- Loss of consciousness16
- Hyperglycemia15
- Hypoglycemia15
- Insufficient Information6
- Dizziness4
- Fatigue3
Recent adverse event reports
The 25 most recent of 1,337 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL and 53 mg/dL compared to readings of 177 mg/dL and 107 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931740 ↗MalfunctionSep 28, 2026Low Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926206 ↗MalfunctionSep 24, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of380mg/dL,285mg/dL 400mg/dL compared to glucose values of 145mg/dL,100mg/dL ,185mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 2 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884011 ↗MalfunctionSep 23, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 401 mg/dL, 401 mg/dL compared to glucose values of 140 mg/dL, 150 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 15 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868041 ↗MalfunctionSep 22, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 26 mmol/L compared to readings of 2.30 mmol/L respectively obtained on a Competitor Brand Meter and the results, when plotted on a Parkes Error Grid, fall into the E Zone, showing the difference in values to be clinically significant. The length of sensor wear was X days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50853508 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 359mg/dL compared to glucose values of 149mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 Days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50840615 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 61mg/dL compared to glucose values of 202mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear is an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50839729 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 69 mg/dL compared to glucose values of 180 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50837228 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 68 mg/dL compared to glucose values of 255 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50836352 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 88 mg/dL compared to glucose values of 264 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50836294 ↗MalfunctionSep 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 120 mg/dL compared to glucose values of 74 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear is an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50835357 ↗MalfunctionSep 18, 2026Detachment of Device or Device Component
A customer reported that an ADC device component detached and remained in skin. There was no indication that the customer required intervention from a healthcare professional to remove the component. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50818058 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 501 mg/dL compared to glucose values of 77 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 13 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50818057 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 55 mg/dL , 185 mg/dL , 168 mg/dL compared to glucose values of 200 mg/dL , 251 mg/dL , 251 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817856 ↗MalfunctionSep 18, 2026Detachment of Device or Device Component
A customer reported that an ADC device component detached and remained in skin. There was no indication that the customer required intervention from a healthcare professional to remove the component. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50817415 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 165 mg/dL compared to glucose values of 85 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817273 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 54 mg/dL, 114 mg/dL compared to glucose values of 200 mg/dL, 198 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817114 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 56 mg/dL compared to glucose values of 181 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50815969 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to glucose values of 210 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50815441 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 52 mg/dL compared to glucose values of 163 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50814319 ↗MalfunctionSep 18, 2026Device Alarm System
A customer reported an unexpected high/low glucose alarm with the ADC device, in use with the application. There was no indication that the customer experienced an adverse event or required intervention as a result of the reported issue. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50814236 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 298mg/dL compared to glucose values of 148mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear is an unknown number of days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50814141 ↗MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 39 mg/dL, 47 mg/dL compared to glucose values of 309 mg/dL, 298 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50813111 ↗MalfunctionSep 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 300 mg/dL compared to glucose values of 89 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the D Zone, showing the difference in values to be clinically significant. The length of sensor wear was 9 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50800586 ↗MalfunctionSep 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 55 mg/dL compared to glucose values of 174 mg/dL respectively obtained on the comparison device, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50800585 ↗
Other Integrated Continuous Glucose Monitoring system, factory calibrated devices
- Dexcom G6 Transmitter 9445-24Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated240,475 reports
- Dexcom G6 Transmitter 9445-02Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 deaths229,493 reports
- Dexcom G6 Sensor Applicator Assembly 9500-45Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall4 deaths114,678 reports
- US G6 Transmitter (Retail, Receiver Optional) STT-OE-002Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated76,977 reports
- Dexcom G7 GSS (161) 9500-161Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated70,146 reports
- Dexcom G7 iOS App SW12300Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated3 recalls1 death55,219 reports
- Dexcom G7 GSS (161) STP-AT-012Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated1 recall5 deaths48,180 reports
- G6 Medicare Transmitter STT-OM-001Dexcom, Inc. · Integrated Continuous Glucose Monitoring System, Factory Calibrated38,339 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.