FreeStyle Libre 2Recalls & reported problems across all models
Abbott Diabetes Care Inc · 12 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one FreeStyle Libre 2 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 30,420
- Deaths reported
- 2
- Injuries reported
- 5,600
- Recalls
- 2
We found 12 models of FreeStyle Libre 2 by Abbott Diabetes Care Inc in the FDA's device registry (GUDID). Together they are named in 30,420 adverse event reports in the FDA's MAUDE database, received between Sep 18, 2020 and Sep 30, 2026 — classified as 2 deaths, 5,600 injuries, 24,818 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (19,887 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
FreeStyle Libre 2 recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jul 12, 2023Open, Classified
If blood glucose monitoring system users attempt to start a new sensor, when the old sensor wasn't used for the full 14-day wear time, then the reader may display "Incompatible Sensor" message. If users don't have the previous sensor, or FreeStyle Libre 2 app, or the sensor's full wear time hasn't ended, users may be unable to start a new sensor, which could result in no or delayed glucose results
Z-2414-2023 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗
- Feb 13, 2023Open, Classified
Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.
Z-1275-2023 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗
Most reported FreeStyle Libre 2 problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings19,887
- Unable to Obtain Readings2,838
- Device Displays Incorrect Message2,737
- Product Quality Problem2,705
- Device Alarm System1,486
- Detachment of Device or Device Component1,458
- Low Readings851
- High Readings438
Patient problems
- No Clinical Signs, Symptoms or Conditions23,973
- Loss of consciousness2,627
- Hypoglycemia2,535
- Hyperglycemia983
- Convulsion/Seizure863
- Device Embedded In Tissue or Plaque811
- Dizziness591
- Insufficient Information569
FreeStyle Libre 2 models
12 models, most reported first. Each model has its own page with its reports, recalls and device details.
- FreeStyle Libre 2 71992Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems1 death23,063 reports
- FreeStyle Libre 2 71953System, test, blood glucose, over the counter2 recalls1 death3,286 reports
- FreeStyle Libre 2 Plus 78747Integrated Continuous Glucose Monitoring system, factory calibrated1,833 reports
- FreeStyle Libre 2 Plus 78747Integrated Continuous Glucose Monitoring system, factory calibrated1,337 reports
- FreeStyle Libre 2 72000Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems453 reports
- FreeStyle Libre 2 71952System, test, blood glucose, over the counter1 recall324 reports
- FreeStyle Libre 2 71951Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems2 recalls72 reports
- FreeStyle Libre 2 App 71926-01Integrated continuous glucose monitoring system, factory calibrated, not for use with automated insulin delivery systems21 reports
- FreeStyle Libre 2 Plus 78749Integrated Continuous Glucose Monitoring system, factory calibrated17 reports
- FreeStyle Libre 2 71953System, Test, Blood Glucose, Over The Counter6 reports
- FreeStyle Libre 2 Plus 78749Integrated Continuous Glucose Monitoring system, factory calibrated5 reports
- FreeStyle Libre 2 App 71857-01System, test, blood glucose, over the counter3 reports
Recent FreeStyle Libre 2 adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Detachment of Device or Device Component
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50955132 ↗MalfunctionSep 30, 2026Detachment of Device or Device Component
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 50953814 ↗InjurySep 30, 2026Unable to Obtain Readings; Device Displays Incorrect Message
Patient problems: Hypoglycemia
MAUDE report 50953813 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 46 mg/dL compared to readings of 249 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50949349 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 200 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into C Zone, showing difference in values to be clinically significant. The length of sensor wear was 4 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946446 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50.00 mg/dL compared to readings of 210.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944676 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 66.00 mg/dL,68.00 mg/dL compared to readings of 172.00 mg/dL,202.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 8 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944671 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 103.00 mg/dL compared to readings of 249.00 mg/dL respectively obtained on a Meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 5 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944336 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 64.00 mg/dL compared to readings of 172.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943968 ↗MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943513 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50.00 mg/dL and 58.00 mg/dL compared to readings of 258.00 mg/dL and 240.00 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 3 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935928 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 57 mg/dL compared to readings of 220 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 7 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931744 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 50 mg/dL and 53 mg/dL compared to readings of 177 mg/dL and 107 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 1 day. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50931740 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50930492 ↗MalfunctionSep 29, 2026Incorrect, Inadequate or Imprecise Result or Readings
A customer reported receiving erroneous glucose results from an Abbott Diabetes Care device. The customer received sensor scan results of 53 mg/dL compared to readings of 162 mg/dL respectively obtained on a meter and the results, when plotted on a Parkes Error Grid, fall into the C Zone, showing the difference in values to be clinically significant. The length of sensor wear was 10 days. There was no report of death, serious injury, or mistreatment associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50929666 ↗
Other Abbott Diabetes Care Inc products
- FreeStyle Libre 314 models · 69,860 reports · 6 recalls
- FreeStyle Libre 3 Plus5 models · 52,808 reports · 2 recalls
- FreeStyle Libre 14 day5 models · 16,749 reports · 1 recall
- FreeStyle Lite9 models · 3,388 reports
- FreeStyle Libre 2 Plus4 models · 3,192 reports
- FreeStyle Freedom Lite5 models · 2,436 reports
- FreeStyle Libre7 models · 705 reports · 1 recall
- Precision Xtra10 models · 391 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered, combining the brand names used for this product line.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page