Skip to content

HeartWare Controller DC AdapterRecalls & reported problems across all models

Heartware Inc. · 6 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one HeartWare Controller DC Adapter model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
59
Deaths reported
0
Injuries reported
0
Recalls
1

We found 6 models of HeartWare Controller DC Adapter by Heartware Inc. in the FDA's device registry (GUDID). Together they are named in 59 adverse event reports in the FDA's MAUDE database, received between Aug 10, 2017 and Sep 5, 2023 — classified as 59 malfunctions.

The most frequently reported specific device problem is “Power Problem” (35 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

HeartWare Controller DC Adapter recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 2, 2018Open, Classified

    Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.

    Z-1903-2018 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

Most reported HeartWare Controller DC Adapter problems

Device problems

  • Power Problem35
  • Connection Problem13
  • Material Integrity Problem12
  • Material Split, Cut or Torn3
  • Component Missing1
  • Device Operates Differently Than Expected1
  • Disconnection1
  • Display or Visual Feedback Problem1

Patient problems

  • No Known Impact Or Consequence To Patient25
  • No Clinical Signs, Symptoms or Conditions17
  • No Consequences Or Impact To Patient17
  • Dizziness1
  • No Patient Involvement1

HeartWare Controller DC Adapter models

6 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent HeartWare Controller DC Adapter adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Heartware Inc. products

All Heartware Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page