Skip to content

iMRX SystemRecalls & reported problems across all models

IMRIS Imaging, Inc. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one iMRX System model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls
1

We found 2 models of iMRX System by IMRIS Imaging, Inc. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Aug 24, 2024 and Aug 24, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Output Problem” (1 report).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

iMRX System recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 30, 2025Open, Classified

    There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

    Z-0161-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

Most reported iMRX System problems

Device problems

  • Output Problem1
  • Unexpected Shutdown1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

iMRX System models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent iMRX System adverse event reports

The 1 most recent reports across all models, in the reporter's own words as published by the FDA.

Other IMRIS Imaging, Inc. products

All IMRIS Imaging, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page