Artis zeegoRecalls & reported problems across all models
Siemens Healthcare GmbH · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
5 open recalls name at least one Artis zeego model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 31
- Deaths reported
- 2
- Injuries reported
- 2
- Recalls
- 6
We found 2 models of Artis zeego by Siemens Healthcare GmbH in the FDA's device registry (GUDID). Together they are named in 31 adverse event reports in the FDA's MAUDE database, received between Jul 16, 2020 and Apr 30, 2026 — classified as 2 deaths, 2 injuries, 27 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (9 reports).
6 FDA recall records name at least one of these models by its device identifier, 5 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Artis zeego recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Mar 21, 2024Open, Classified
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Z-1598-2024 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗
- Dec 16, 2022Open, Classified
In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
Z-1040-2023 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗
- Aug 10, 2022Open, Classified
Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.
Z-1645-2022 · Root cause: Device Design · Recall details · FDA record ↗
- May 24, 2022Terminated
It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.
Z-1343-2022 · Root cause: Component design/selection · names 2 of these models · Recall details · FDA record ↗
- Feb 23, 2022Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Z-0977-2022 · Root cause: Device Design · Recall details · FDA record ↗
- Feb 23, 2022Open, Classified
If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".
Z-0973-2022 · Root cause: Device Design · Recall details · FDA record ↗
Most reported Artis zeego problems
Device problems
- Mechanical Problem9
- Insufficient Device Problem Information3
- No Apparent Adverse Event3
- Output Problem3
- Radiation Output Problem3
- Use of Device Problem3
- Electrical /Electronic Property Problem2
- Radiation Output Failure2
Patient problems
- No Clinical Signs, Symptoms or Conditions24
- Cardiac Arrest2
- Insufficient Information2
- Fall1
- No Known Impact Or Consequence To Patient1
- Radiodermatitis1
- Renal Impairment1
Artis zeego models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Artis zeego 10848283Interventional fluoroscopic x-ray system5 recalls1 death7 reports
- Artis zeego 10280959Interventional fluoroscopic x-ray system4 recalls1 death24 reports
Recent Artis zeego adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2026Use of Device Problem; Device Handling Problem
SIEMENS WAS INFORMED ABOUT AN ADVERSE EVENT THAT OCCURRED WHILE OPERATING THE ARTIS Q ZEEGO SYSTEM. THE PATIENT, WHO WAS PLACED ON THE TABLE IN THE HYBRID OPERATING THEATRE, FELL FROM THE TABLE. ACCORDING TO THE INFORMATION PROVIDED, THE INCIDENT WAS DESCRIBED AS AN ASSISTED FALL OR A SLIDE. THE PATIENT LATER PASSED AWAY; HOWEVER, IT WAS CONFIRMED THAT THE FALL DID NOT CONTRIBUTE TO THE DEATH. NO CAUSAL RELATIONSHIP BETWEEN DEVICE PERFORMANCE AND THE PATIENT¿S DEATH WAS IDENTIFIED.
Patient problems: Fall
MAUDE report 25043552 ↗MalfunctionAug 7, 2025Radiation Overexposure; Radiation Overexposure
SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WHILE OPERATING THE ARTIS ZEEGO UNIT. SIEMENS RECEIVED INFORMATION THAT THE DOSE ON AN INTERVENTION RADIOLOGIST'S BADGE WAS SIGNIFICANTLY HIGHER IN (B)(6) 2024 THAN IN OTHER MONTHS. NO INFORMATION ON THE RECORDED DOSE WAS PROVIDED. NO HEALTH CONSEQUENCES WERE REPORTED FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22740064 ↗MalfunctionSep 23, 2024Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS Q ZEEGO SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, A SYSTEM MALFUNCTION AND SYNCHRONIZATION FAILURE TO THE DATABASE OCCURRED. ADDITIONAL INFORMATION WAS PROVIDED THAT A PATIENT MAY HAVE SUFFERED AN INJURY TO THE KIDNEY. DETAILED CLARIFICATIONS OF THIS EVENT ARE CURRENTLY ONGOING. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: Insufficient Information; Renal Impairment
MAUDE report 20278185 ↗MalfunctionJul 25, 2024Use of Device Problem; No Apparent Adverse Event; Use of Device Problem; No Apparent Adverse Event
SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED ON THE ARTIS ZEEGO UNIT. DURING AN EMERGENCY PATIENT PROCEDURE, THE AUTOMATIC ROBOT DRIVE COULD NOT BE ACTIVATED RESTRICTING TABLE MOTION. THE PATIENT HAD TO BE RELOCATED, AND THE PROCEDURE WAS COMPLETED ON AN ALTERNATE UNIT. SIEMENS IS UNAWARE OF ANY ADVERSE HEALTH IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19828643 ↗MalfunctionFeb 26, 2024Unintended Movement; Unintended Movement
SIEMENS BECAME AWARE OF A MALFUNCTION THAT OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. THE USER REPORTED AN UNINTENDED ROBOT MOVEMENT DURING SYSTEM USAGE. NO COLLISION WITH OTHER OBJECTS HAS BEEN REPORTED. WE HAVE NO INDICATIONS OF ADVERSE EFFECTS ON THE HEALTH STATUS OF PATIENTS, USERS OR THIRD PARTIES. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION REGARDING THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18778153 ↗InjuryFeb 19, 2024Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WHILE OPERATING THE ARTS Q ZEEGO SYSTEM DURING AN EMERGENCY PROCEDURE. IT WAS STATED BY THE USER THAT DURING THE SYSTEM START-UP, ALL SYSTEM MOVEMENTS WERE BLOCKED AND AN ERROR MESSAGE ¿COLLISION CONTROL DEACTIVATED¿ WAS DISPLAYED TO THE USER. DUE TO THIS ISSUE THE EMERGENCY PATIENT PROCEDURE THAT WAS PLANNED ON THE CONCERNED UNIT HAD TO BE PERFORMED WITH AN ALTERNATE MOBILE C-ARM SYSTEM. ADDITIONAL INFORMATION STATING THAT THE PATIENT PASSED AWAY WAS PROVIDED TO SIEMENS. THE RELATIONSHIP BETWEEN THE SYSTEM BEHAVIOR AND THE PATIENT'S DEATH IS UNKNOWN AS OF THE DATE OF THIS REPORT. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION FOR THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 18731301 ↗MalfunctionNov 17, 2023Unintended Collision; Unintended Collision
SIEMENS BECAME AWARE OF AN INCIDENT THAT OCCURRED WHILE OPERATING THE ARTIS Q ZEEGO SYSTEM. IT WAS REPORTED THAT THE C-ARM CRASHED INTO THE STAND. WE HAVE NO INDICATIONS OF ANY ADVERSE EFFECTS ON THE HEALTH STATUS OF THE INVOLVED PERSONS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18160761 ↗MalfunctionJul 19, 2022Output Problem; Radiation Output Problem; Radiation Output Failure; Output Problem; Radiation Output Problem; Radiation Output Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. DURING SYSTEM START-UP FOR AN EMERGENCY PROCEDURE, THE USER REPORTED THAT NO X-RAY WAS AVAILABLE WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT. THE REPORTED EVENT OCCURRED IN SWEDEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15052942 ↗MalfunctionJul 18, 2022Output Problem; Radiation Output Problem; Radiation Output Failure; Output Problem; Radiation Output Problem; Radiation Output Failure
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT NO X-RAY WAS AVAILABLE AND AN EXPOSURE WAS NOT POSSIBLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT. THE REPORTED EVENT OCCURRED IN NORWAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15045074 ↗MalfunctionJun 22, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT NO X-RAY WAS AVAILABLE AND AN EXPOSURE WAS NOT POSSIBLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14771731 ↗MalfunctionJun 21, 2022Mechanical Jam; Mechanical Jam
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT THE SYSTEM WAS DRIVING WITH SLOW SPEED. AS A RESULT, FULL RESTART OF THE SYSTEM CAUSED A DELAY OF 20 MINUTES. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14759310 ↗DeathApr 27, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE ARTIS Q ZEEGO SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, SYSTEM MOVEMENT WAS BLOCKED. ADDITIONAL INFORMATION WAS PROVIDED THAT THE PATIENT PASSED AWAY AND THE USER DID NOT USE PATIENT RESCUE ON THE ARTIS Q ZEEGO. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: Cardiac Arrest
MAUDE report 14217070 ↗DeathJan 24, 2022Optical Problem; Optical Problem
IT WAS REPORTED TO SIEMENS THAT AN ADVERSE EVENT OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. THE CUSTOMER REPORTED NO IMAGE ON THE LARGE DISPLAY AFTER SYSTEM START UP. FURTHER INFORMATION WAS PROVIDED THAT THE PATIENT PASSED AWAY. AT THIS TIME, THERE IS NO CAUSAL RELATIONSHIP BETWEEN THE DEATH OF THE PATIENT AND THE SYSTEM BEHAVIOR. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: Cardiac Arrest
MAUDE report 13333216 ↗MalfunctionOct 19, 2021Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE ARTIS ZEEGO SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE USER REPORTED THAT NO MOVEMENT WAS POSSIBLE. THE PROCEDURE WAS CONTINUED AND COMPLETED ON AN ALTERNATE SYSTEM. WE ARE UNAWARE OF ANY IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12655374 ↗MalfunctionSep 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE SYNGO.X WORKPLACE ON AN ARTIS ZEEGO SYSTEM. THE USER REPORTED THAT THE CT PORTION OF THE ARTIS ZEEGO X-RAY SYSTEM WAS NOT WORKING. THE OR TEAM HAD TO TURN THE MACHINE OFF AND ON THREE TIMES BEFORE THE MACHINE WOULD TURN ON. THE USER REPORTED THAT THIS HAS HAPPENED OVER THE PAST FEW DAYS. THERE IS NO REPORT OF IMPACT TO THE STATE OF HEALTH OF ANY PATIENT OR USER INVOLVED. SIEMENS HAS REQUESTED ADDITIONAL INFORMATION IN ORDER TO CONDUCT AN INVESTIGATION OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12445503 ↗
Other Siemens Healthcare GmbH products
- MAGNETOM Verio2 models · 2 reports · 4 recalls
- MAGNETOM Skyra2 models · 10 reports · 3 recalls
- MAGNETOM Cima.X2 models · 0 reports · 3 recalls
- MAGNETOM Vida2 models · 12 reports · 2 recalls
- MAGNETOM Lumina2 models · 3 reports · 2 recalls
- MAGNETOM Prisma2 models · 2 reports · 2 recalls
- MAGNETOM Sola2 models · 16 reports · 1 recall
- Cios Alpha2 models · 15 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page