Skip to content

Artis zeegoRecalls & reported problems across all models

Siemens Healthcare GmbH · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one Artis zeego model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
31
Deaths reported
2
Injuries reported
2
Recalls
6

We found 2 models of Artis zeego by Siemens Healthcare GmbH in the FDA's device registry (GUDID). Together they are named in 31 adverse event reports in the FDA's MAUDE database, received between Jul 16, 2020 and Apr 30, 2026 — classified as 2 deaths, 2 injuries, 27 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (9 reports).

6 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Artis zeego recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 21, 2024Open, Classified

    Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

    Z-1598-2024 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  2. Dec 16, 2022Open, Classified

    In rare cases, the system may only boot into backup mode after an abrupt shutdown and not reach full operating mode. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.

    Z-1040-2023 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  3. Aug 10, 2022Open, Classified

    Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.

    Z-1645-2022 · Root cause: Device Design · Recall details · FDA record ↗

  4. May 24, 2022Terminated

    It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.

    Z-1343-2022 · Root cause: Component design/selection · names 2 of these models · Recall details · FDA record ↗

  5. Feb 23, 2022Open, Classified

    If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

    Z-0977-2022 · Root cause: Device Design · Recall details · FDA record ↗

  6. Feb 23, 2022Open, Classified

    If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

    Z-0973-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported Artis zeego problems

Device problems

  • Mechanical Problem9
  • Insufficient Device Problem Information3
  • No Apparent Adverse Event3
  • Output Problem3
  • Radiation Output Problem3
  • Use of Device Problem3
  • Electrical /Electronic Property Problem2
  • Radiation Output Failure2

Patient problems

  • No Clinical Signs, Symptoms or Conditions24
  • Cardiac Arrest2
  • Insufficient Information2
  • Fall1
  • No Known Impact Or Consequence To Patient1
  • Radiodermatitis1
  • Renal Impairment1

Artis zeego models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Artis zeego adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Siemens Healthcare GmbH products

All Siemens Healthcare GmbH devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page