Merge CardioRecalls & reported problems across all models

Merge Healthcare Solutions Inc. · 6 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

4 open recalls name at least one Merge Cardio model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
34
Deaths reported
1
Injuries reported
0
Recalls
5

We found 6 models of Merge Cardio by Merge Healthcare Solutions Inc. in the FDA's device registry (GUDID). Together they are named in 34 adverse event reports in the FDA's MAUDE database, received between Aug 12, 2018 and Mar 20, 2025 — classified as 1 death, 33 malfunctions.

The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (12 reports).

5 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Merge Cardio recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 19, 2025Open, Classified

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

    Z-1402-2025 · Root cause: Software design · names 3 of these models · Recall details · FDA record ↗

  2. Feb 19, 2025Open, Classified

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

    Z-1401-2025 · Root cause: Software design · Recall details · FDA record ↗

  3. Feb 19, 2025Open, Classified

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

    Z-1400-2025 · Root cause: Software design · Recall details · FDA record ↗

  4. Feb 19, 2025Open, Classified

    When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

    Z-1399-2025 · Root cause: Software design · Recall details · FDA record ↗

  5. May 8, 2020Terminated

    An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative.

    Z-2403-2020 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗

Most reported Merge Cardio problems

Device problems

  • Application Program Problem: Parameter Calculation Error12
  • Data Problem11
  • Application Program Problem6
  • Computer Software Problem6
  • Loss of Data2
  • Patient Data Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Known Impact Or Consequence To Patient13
  • Insufficient Information1
  • Unspecified Heart Problem1

Merge Cardio models

6 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Merge Cardio adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Merge Healthcare Solutions Inc. products

All Merge Healthcare Solutions Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page