Merge Cardio recallZ-1399-2025 · Merge Healthcare, Inc. · initiated Feb 19, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merge Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
FDA root cause category: Software design
What the company is doing
Merge Healthcare issued an Urgent Field Safety Notice to it's consignees on 02/19/2025 via email. The notice explained the issue, risk to patients, provided a workaround to immediately address the issue, and recommended clinicians check measurements in the final report to ensure that they accurately reflect the patient's condition pending software update. Required Actions: Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Schedule an update to implement the configuration change on your system. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge. Please forward this notice to any individuals within your organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The firm identified a missing software version (11.0.2) and issued the corrected notice on 03/03/2025.
Recall details
- Recall number
- Z-1399-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 19, 2025
- Posted by the FDA
- Mar 20, 2025
- Recalling firm
- Merge Healthcare, Inc.
- Quantity
- 11
- Product code
- LLZ
- FDA submissions
- K051649
- Distribution
- US
Product description
VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
Code information (models, lots, serial numbers)
Software Versions and Part Numbers: 10.0: Part Numbers: 88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00, 88-00532-00, 88-00533-00, 70-00008-00; 10.1.1.1: Part Numbers: 88-01094-00; 10.1.2: Part Numbers: 88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00; 10.2.0: Part Numbers: 88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00, 88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00, 10.2 P2: Part Numbers: 88-01203-00 10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Merge Cardio 10.3Merge Healthcare Solutions Inc. · System, Image Processing, Radiological2 recalls13 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K051649 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Merge Cardio V11.1Merge Healthcare Solutions Inc. · System, Image Processing, Radiological2 recalls14 reports
- Merge Cardio 11.0Merge Healthcare Solutions Inc. · System, Image Processing, Radiological2 recalls8 reports
- Merge Cardio 12.1Merge Healthcare Solutions Inc. · System, Image Processing, Radiological1 recall1 death8 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96344).
- Z-1398-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture arch…
- Z-1400-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system cl…
- Z-1401-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture …
- Z-1402-2025 · Open, ClassifiedMerge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical devic…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page