Merge Cardio recallZ-1402-2025 · Merge Healthcare, Inc. · initiated Feb 19, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merge Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
FDA root cause category: Software design
What the company is doing
Merge Healthcare issued an Urgent Field Safety Notice to it's consignees on 02/19/2025 via email. The notice explained the issue, risk to patients, provided a workaround to immediately address the issue, and recommended clinicians check measurements in the final report to ensure that they accurately reflect the patient's condition pending software update. Required Actions: Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Schedule an update to implement the configuration change on your system. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge. Please forward this notice to any individuals within your organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The firm identified a missing software version (11.0.2) and issued the corrected notice on 03/03/2025.
Recall details
- Recall number
- Z-1402-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 19, 2025
- Posted by the FDA
- Mar 20, 2025
- Recalling firm
- Merge Healthcare, Inc.
- Quantity
- 253
- Product code
- LLZ
- FDA submissions
- K192206
- Distribution
- US
Product description
Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.1, 12.1.1, 12.2, 12.3, 12.3.1, 12.3.2, 12.3.3, 12.4, 12.4.1, 12.4.2; Merge Cardio is a software medical device that is an integrated multi-modality image cardiovascular information management system.
Code information (models, lots, serial numbers)
Lot Code: Software versions and UDIs 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206 12.1, UDI: (01)00842000100928(10)12.1(11)210122; 12.1.1, UDI: (01)00842000100928(10)12.1.1(11)211220; 12.2 UDI: (01)00842000100065(10)12.2.0(11)220713; 12.3, UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.1 UDI: (01)00842000101109(10)12.3.0(11)221114; 12.3.2 UDI: (01)00842000101109(10)12.3.2(11)230612; 12.3.3 UDI: (01)00842000101109(10)12.3.3(11)230913; 12.4 UDI: (01)00842000101123(10)12.4.0(11)231220; 12.4.1 UDI: (01)00842000101123(10)12.4.1(11)240522; 12.4.2 UDI: (01)00842000101123(10)12.4.2(11)250206
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Merge Cardio 12.1Merge Healthcare Solutions Inc. · System, Image Processing, Radiological1 recall1 death8 reports
- Merge Cardio 12.3Merge Healthcare Solutions Inc. · System, Image Processing, Radiological1 recall3 reports
- Merge Cardio 12.2Merge Healthcare Solutions Inc. · System, Image Processing, Radiological1 recall1 report
- Merge Cardio 12.4Merge Healthcare Solutions Inc. · System, Image Processing, Radiological1 recall0 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K192206 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- 32mm Glenosphere and Humeral Cup 102-0803Fx Solutions · Shoulder Prosthesis, Reverse Configuration3 reports
- Humelock Reversed System 105-3210Fx Solutions · Shoulder Prosthesis, Reverse Configuration3 reports
- Humeral Cup Standard 312-0703Fx Solutions · Shoulder Prosthesis, Reverse Configuration1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 96344).
- Z-1398-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture arch…
- Z-1399-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information…
- Z-1400-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 11.x, Software Version: 11.0.2, 11.0.3, 11.0.4, 11.1, 11.1.1; The VERICIS is an integrated cardiovascular information system cl…
- Z-1401-2025 · Open, ClassifiedVERICIS, Merge Cardio, Model/Catalog Number: Version 12.X, Software Version: 12.0.1, 12.0.2; The VERICIS is an integrated cardiovascular information system classified as a picture …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.