Product/Part: MitraClip NTR Clip Delivery System… recallZ-0116-2023 · Abbott Vascular · initiated Sep 8, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Vascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking component material properties and certain use conditions. Unsuccessful clip procedure can lead to recurrence of severe mitral regurgitation, device embolization, or tissue injury.
FDA root cause category: Process control
What the company is doing
On 9/8/22, correction notices were distributed to customers who were asked to do the following: 1) Read and share the information in the notice with all device users. 2) continue to follow the steps of the IFU (eifu.abbottvascular.com) that is also summarized in the correction notice. It details the most relevant steps to aid in optimal function of the Clip locking mechanism, thereby reducing the probability of EFAA failure and COWL to occur. 3) Complete and return the Acknowledgement Form. 4) If you have transferred affected products to other centers, inform those centers of this notice. The recalling firm is working on producing new lots with updated manufacturing processing and raw material to mitigate the EFAA/COWL events. They are also using the correction notice to emphasize the most relevant sections of the IFU that can reduce occurrence. Customers with questions can contact Customer Service at 1800-544-1664.
Recall details
- Recall number
- Z-0116-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 8, 2022
- Posted by the FDA
- Oct 20, 2022
- Recalling firm
- Abbott Vascular
- Quantity
- 20,045
- Product code
- NKM
- Distribution
- Worldwide distribution - US Nationwide including in the states of PR, WA, NE, MO, IA, KS, TN, MN, ND, MD, VA, MA, PA, DC, LA, KY, CA, NV, IL, WI, TX, CT, VT, AL, AR, GA, MS, MI, OH, FL, NC, NJ, CO, NY, DE, OR, SC, IN, HI, AZ, NH, ME, WV, OK, ID, UT, MT, NM, RI, AK, SD and the countries of AE, AR, AT, AU, BE, BG, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, EC, EE, EG, ES, FI, FR, GB, GR, HK, HU, ID, IE, IL, IN, IR, IT, JO, JP, KR, KW, LB, LU, MT, MX, MY, NL, NO, NZ, OM, PE, PH, PK, PL, PT, QA, RE, RO, RU, SA, SE, SG, SI, SK, TH, TN, TR, TW.
Product description
Product/Part: MitraClip NTR Clip Delivery System US/CDS0601-NTR, MitraClip NTR Clip Delivery System CE/CDS0602-NTR, MitraClip XTR Clip Delivery System CE/CDS0602-XTR, MitraClip G4 Clip Delivery System NT CE/CDS0702-NT, MitraClip G4 Clip Delivery System NTW CE/CDS0702-NTW, MitraClip G4 Clip Delivery System XT CE/CDS0702-XT, MitraClip G4 Clip Delivery System XTW CE/CDS0702-XTW, MitraClip G4 Clip Delivery Sys NT ROW/CDS0705-NT, MitraClip G4 Clip Delivery Sys NTW ROW/CDS0705-NTW, MitraClip G4 Clip Delivery Sys XT ROW/CDS0705-XT, MitraClip G4 Clip Delivery Sys XTW ROW/CDS0705-XTW, TriClip XT Clip Delivery Sys CE/TCDS0202-XT, TriClip NT Clip Delivery Sys CE/TCDS0202-NT, TriClip XT Clip Delivery Sys ROW/TCDS0205-XT, TriClip NT Clip Delivery Sys ROW/TCDS0205-NT, TriClip G4 NT Delivery System CE/TCDS0302-NT, TriClip G4 NTW Delivery System CE/TCDS0302-NTW, TriClip G4 XT Delivery System CE/TCDS0302-XT, TriClip G4 XTW Delivery System CE/TCDS0302-XTW, TriClip G4 NT Delivery System ROW/TCDS0305-NT, TriClip G4 NTW Delivery System ROW/TCDS0305-NTW, TriClip G4 XT Delivery System ROW/TCDS0305-XT, TriClip G4 XTW Delivery System ROW/TCDS0305-XTW, TriClip NT Clip Delivery Sys CE/TCDS0202-NT, TriClip NT Clip Delivery Sys ROW/TCDS0205-NT. For cardiac procedures.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Part/UDI-GTIN: CDS0601-NTR/08717648226342, CDS0602-NTR/08717648226359, CDS0602-XTR/08717648226373, CDS0702-NT/08717648230950, CDS0702-NTW/08717648230974, CDS0702-XT/08717648230998, CDS0702-XTW/08717648231018, CDS0705-NT/08717648288616, CDS0705-NTW/08717648288630, CDS0705-XT/08717648288623, CDS0705-XTW/08717648288647, TCDS0202-XT/08717648229978, TCDS0202-NT/08717648229985, TCDS0205-XT/08717648288708, TCDS0205-NT/08717648288692, TCDS0302-NT/08717648334238, TCDS0302-NTW/08717648334252, TCDS0302-XT/08717648334245, TCDS0302-XTW/08717648334269, TCDS0305-NT/08717648334313, TCDS0305-NTW/08717648334337, TCDS0305-XT/08717648334320, TCDS0305-XTW/08717648334344, TCDS0202-NT/08717648229985, TCDS0205-NT/08717648288692 All lots.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- MITRACLIP CDS0601-NTRAbbott Vascular Inc. · Mitral valve repair devices1 recall21 deaths641 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90840).
- Z-0115-2023 · Open, ClassifiedProduct/Part: MitraClip XTR Clip Delivery System US/CDS0601-XTR, MitraClip G4 Clip Delivery System NT US/CDS0701-NT, MitraClip G4 Clip Delivery System NTW US/CDS0701-NTW, MitraClip…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page