Skip to content

MSS - Monobloc stemRecalls & reported problems across all models

Medacta International SA · 10 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names at least one MSS - Monobloc stem model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls
1

We found 10 models of MSS - Monobloc stem by Medacta International SA in the FDA's device registry (GUDID). None of them is named in an adverse event report in our data.

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

MSS - Monobloc stem recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Apr 30, 2026Class IIOpen, Classified

    Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

    Z-2618-2026 · Root cause: Device Design · names 10 of these models · Recall details · FDA record ↗

MSS - Monobloc stem models

10 models, most reported first. Each model has its own page with its reports, recalls and device details.

Other Medacta International SA products

All Medacta International SA devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page