Skip to content

MSS - Monobloc stem 04.01.10.0978Recalls & reported problems

Medacta International SA · Short Monobloc Stem Broach – Ø15 · Class 1

All models of MSS - Monobloc stem (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Apr 30, 2026Class IIOpen, Classified

    Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

    Z-2618-2026 · Root cause: Device Design · Recall details · FDA record ↗

Other Orthopedic manual surgical instrument devices

All Orthopedic manual surgical instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page