MSS - Monobloc stem recallZ-2618-2026 · Medacta Usa Inc · initiated Apr 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medacta Usa Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Surgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.

FDA root cause category: Device Design

What the company is doing

On 4/30/2026, field safety notices were emailed to customers who were asked to do the following: No pre-emptive interventions are recommended, however monitoring of the range of motion and stability should be evaluated during patient follow-up visits as per the standard of care. " Return all units of the lot numbers listed to firm. " Forward this notice to all people and organizations potentially affected. Firm Customer Service can be reached at 312-878-2381.

Recall details

Recall number
Z-2618-2026
Status
Open, Classified
Initiated by the company
Apr 30, 2026
Posted by the FDA
Jun 29, 2026
Recalling firm
Medacta Usa Inc
Quantity
50
Product code
PHX
FDA submissions
K250644
Distribution
US Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.

Product description

Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System

Code information (models, lots, serial numbers)

REF/UDI/Lot: 04.01.10.0970/07630542768910/2585086, 04.01.10.0971/07630542768927/2585087, 04.01.10.0972/07630542768934/2585088, 04.01.10.0973/07630542768941/2585089, 04.01.10.0974/07630542768958/2585090, 04.01.10.0975/07630542768965/2585091, 04.01.10.0976/07630542768972/2585092, 04.01.10.0977/07630542768989/2585093, 04.01.10.0978/07630542768996/2585094, 04.01.10.0979/07630542769009/2585095

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines