QUADROX-i recallZ-0437-2019 · Maquet Cardiovascular Us Sales, Llc · initiated Oct 25, 2018

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm, GETTINGE, sent an "URGENT MEDICAL DEVICE RECALL" notification letters dated 10/25/18 to customers. The letters described the product, problem, and actions to be taken. The customers were instructed to do the following: -Please immediately examine your inventory to determine if you have any of the QUAD ROX-i Neonatal Oxygenator with the product codes/lot numbers listed in this notice. -If you wish to use the product, please remove and do not use the enclosed reducing screw adapter 1/4" x 3/16" (Ref. No. 70104.8593). -If you do not wish to use the product as packaged, please contact Maquet/Ge tinge Customer Support at +l (888) 627-8383 (press option 2, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) to obtain instructions on returning the affected product for a full credit. Please note that this product is no longer being offered in the United States, and therefore product replacement is not available. -Please complete and sign the attached MEDICAL DEVICE RECALL- RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Maquet/Ge tinge by e-mailing a scanned copy to [email protected] or by f axing the form to +l (800) 859-8371. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. If you have any questions, please contact your Maquet Cardiopulmonary /Ge tinge representative or call the Maquet Cardiopulmonary /Ge tinge Customer Support at +l (888) 627-8383 (press option 2, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).

Recall details

Recall number
Z-0437-2019
Status
Terminated
Initiated by the company
Oct 25, 2018
Posted by the FDA
Nov 28, 2018
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Quantity
490 total in US
Product code
DTZ
FDA submissions
K102464, K141432
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakistan, Portugal, Reunion, Saudi Arabia, Sweden, South Africa, Poland, Serbia, Russia, Turkey, Ukraine, China, India, Japan, Republic of Korea (South Korea), Malaysia, Thailand, and Taiwan.

Product description

QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Code information (models, lots, serial numbers)

UDI: 04037691773414; Lots: 92209374, 92211752, 92215106, 92222131, 92223169, 92228805, 92234286, 92236876

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 81382).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page