Skip to content

RevolutionRecalls & reported problems across all models

GE Healthcare Japan Corporation · 3 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

5 open recalls name at least one Revolution model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls
5

We found 3 models of Revolution by GE Healthcare Japan Corporation in the FDA's device registry (GUDID). None of them is named in an adverse event report in our data.

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Revolution recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 26, 2026Class IIOpen, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2149-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Nov 5, 2024Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0608-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

  3. Nov 5, 2024Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0606-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

  4. Nov 5, 2024Class IIOpen, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0605-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

  5. Jun 2, 2022Class IIOpen, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · names 2 of these models · Recall details · FDA record ↗

Revolution models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Sources and updates