Revolution AscendRecalls & reported problems

GE Healthcare Japan Corporation · NGX Gantry common top asm with UDI · Class 2

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

1 other recall cover devices cleared under the same FDA submission (K203169, K213938); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 26, 2026Open, Classified

    GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.

    Z-2149-2026 · Root cause: Software design · FDA record ↗

  2. Nov 5, 2024Open, Classified

    GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

    Z-0605-2025 · Root cause: Component design/selection · FDA record ↗

  3. Jun 2, 2022Open, Classified

    As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

    Z-1511-2022 · Root cause: Error in labeling · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K203169, K213938. One submission often covers many models, so check the product and lot details in each record.

  • Z-2023-2021 · Jun 30, 2021 · Open, Classified
    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Ci

Other System, x-ray, tomography, computed devices

All System, x-ray, tomography, computed devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.