Revolution recallZ-2149-2026 · GE Medical Systems, LLC · initiated Mar 26, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
FDA root cause category: Software design
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/26/26 was sent to customers. Actions to be taken by Customer/User Pending corrections by GE HealthCare, you can continue to use your system by following security and cybersecurity best practices. Before the interim solution (see Product Correction section below) is applied, please follow the instructions below as identified in the respective CT system User Manuals: CAUTION: 3D or slab reconstructions provide additional supplemental information, complementing diagnosis that should be based on classical techniques. WARNING: Do not use 3D or slab views only to perform any measurements (distance, angle, region of interest, report cursor, area, volume, etc.). Always check measurement points position and refer to 2D baseline views (acquisition images or reformatted images of minimal thickness) to confirm measurements. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via (https://gehealthcare-svc.my.site.com/publicForm/s/?formId=aGjUr000002jnD7) or print, fill out manually, scan and email to [email protected] Product Correction GE HealthCare will correct all affected products at no cost to you in two steps. A GE HealthCare representative will contact you to arrange for an interim solution to deactivate AW Server authentication through the CT system on the EHL-based Smart Subscription, and later, to install a CT software update. Note: With the interim solution, all alternative authentication mechanisms (direct application launch from CT console user interface, login using users defined in AW Server user database, federation with hospital user directory through LDAP) will remain available. Contact Information If you have any questions or concerns regarding this notification, please contact GE HealthCare
Recall details
- Recall number
- Z-2149-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 26, 2026
- Posted by the FDA
- May 8, 2026
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 25 units
- Product code
- JAK
- FDA submissions
- K213938
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Revolution Ascend, computed tomography, Model Numbers 6969000-100 and 6969000-300
Code information (models, lots, serial numbers)
UDI-DI: 00840682146173; Serial Numbers: AS10G2200136YC AS10G2200101YC AS10G2200068YC AS10G2200035YC AS10G2300093YC AS10G2400094YC AS10G2400079YC AS10G2100008YC AS10G2500002YC CBDWG2400019HM CBDWG2300081HM AS10G2400056YC AS10G2400081YC AS10G2400087YC CBDWG2400117HM CBDWG2400085HM CBDWG2500003HM CBDWG2400140HM CBDWG2400126HM AS10G2200085YC AS10G2300057YC CBDWG2500011HM CBDWG2300056HM CBDWG2300041HM AS10G2300041YC
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Revolution AscendGE Healthcare Japan Corporation · System, x-ray, tomography, computed3 recalls0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K213938 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Revolution Ascend EliteGE Healthcare Japan Corporation · System, x-ray, tomography, computed1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98738).
- Z-2150-2026 · Open, ClassifiedRevolution CT, x-ray computed tomography, Model Numbers 5590000 and 5590000-6
- Z-2151-2026 · Open, ClassifiedRevolution Apex Elite, X-ray Computed Tomography, Model Number 5590000-23
- Z-2152-2026 · Open, ClassifiedRevolution Apex Plus, X-ray Computed Tomography, Model Number 5590000-25
- Z-2153-2026 · Open, ClassifiedRevolution CT ES, X-ray Computed Tomography, Model Number 5590000-15
- Z-2154-2026 · Open, ClassifiedRevolution Apex, X-ray Computed Tomography, Model Number 5590000-20
- Z-2155-2026 · Open, ClassifiedRevolution Apex Select, X-ray Computed Tomography, Model Number 5590000-27
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page