SAFELINERecalls & reported problems across all models
B. Braun Medical Inc. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one SAFELINE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 6 models of SAFELINE by B. Braun Medical Inc. in the FDA's device registry (GUDID). Together they are named in 15 adverse event reports in the FDA's MAUDE database, received between Nov 23, 2016 and Sep 4, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
SAFELINE recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Oct 29, 2025Open, Classified
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Z-0713-2026 · Root cause: Process control · names 3 of these models · Recall details · FDA record ↗
Most reported SAFELINE problems
Device problems
- Fluid/Blood Leak6
- Leak/Splash6
- Contamination /Decontamination Problem1
- Device Contamination with Chemical or Other Material1
- Device Operates Differently Than Expected1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Hemorrhage/Blood Loss/Bleeding3
- No Consequences Or Impact To Patient3
- No Patient Involvement2
SAFELINE models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- SAFELINE® NF1290Set, administration, intravascular1 recall1 report
- SAFELINE® 352604Set, administration, intravascular1 recall0 reports
- SAFELINE® NF1251Set, administration, intravascular1 recall0 reports
- SAFELINE® NF1330Set, administration, intravascular6 reports
- SAFELINE® NF1320Set, administration, intravascular4 reports
- SAFELINE® NF9100Set, administration, intravascular4 reports
Recent SAFELINE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Leak/Splash
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50611610 ↗MalfunctionSep 4, 2026Leak/Splash
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50611607 ↗MalfunctionSep 2, 2026Leak/Splash
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50571350 ↗MalfunctionSep 2, 2021Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 2: END USER REPORTS, "WE ARE HAVING ISSUES WITH OUR T-PORT CONNECTORS FOR IV TUBING. THEY ARE CRACKING AND LEAKING. SAMPLES HAVE BLOOD ON THEM FROM THE LEAKAGE. THIS CAN LEAD TO POTENTIAL INFECTION AND ALSO LOSS OF IVF TO THE PATIENT." NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12419046 ↗MalfunctionSep 2, 2021Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 1: END USER REPORTS, "WE ARE HAVING ISSUES WITH OUR T-PORT CONNECTORS FOR IV TUBING. THEY ARE CRACKING AND LEAKING. SAMPLES HAVE BLOOD ON THEM FROM THE LEAKAGE. THIS CAN LEAD TO POTENTIAL INFECTION AND ALSO LOSS OF IVF TO THE PATIENT." NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12419044 ↗MalfunctionSep 2, 2021Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 3: END USER REPORTS, "WE ARE HAVING ISSUES WITH OUR T-PORT CONNECTORS FOR IV TUBING. THEY ARE CRACKING AND LEAKING. SAMPLES HAVE BLOOD ON THEM FROM THE LEAKAGE. THIS CAN LEAD TO POTENTIAL INFECTION AND ALSO LOSS OF IVF TO THE PATIENT" NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12419043 ↗MalfunctionMar 17, 2021Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 2: AFTER A 45 MINUTE WAIT, THE PRODUCT WAS NOTICED TO BE LEAKING BLOOD AT THE PROXIMAL END OF THE COLLAR THAT SCREWS ON TO THE IV CATHETER. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11504543 ↗MalfunctionMar 17, 2021Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 1: AFTER A 45 MINUTE WAIT, THE PRODUCT WAS NOTICED TO BE LEAKING BLOOD AT THE PROXIMAL END OF THE COLLAR THAT SCREWS ON TO THE IV CATHETER. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11504536 ↗MalfunctionMar 17, 2021Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 3: AFTER A 45 MINUTE WAIT, THE PRODUCT WAS NOTICED TO BE LEAKING BLOOD AT THE PROXIMAL END OF THE COLLAR THAT SCREWS ON TO THE IV CATHETER. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11504533 ↗MalfunctionMar 17, 2021Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY THE USER FACILITY: EVENT 4: AFTER A 45 MINUTE WAIT, THE PRODUCT WAS NOTICED TO BE LEAKING BLOOD AT THE PROXIMAL END OF THE COLLAR THAT SCREWS ON TO THE IV CATHETER. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11504531 ↗MalfunctionApr 4, 2019Contamination /Decontamination Problem; Contamination /Decontamination Problem
AS REPORTED BY THE USER FACILITY: IT WAS REPORTED THAT THE PHARMACY STAFF HAVE NOTICED THAT WHEN UTILIZING THE 9F90100 SAFELINE INJECTION PORT WITH A NEEDLE, THE INJECTION PORT IS CORING AND A PARTICULATE FORMS IN THE SOLUTION. THE CUSTOMER WAS INFORMED THAT THE SAFELINE 9F90100 IS A NEEDLELESS INJECTION PORT WITH A SPLIT SEPTUM AND THAT THE RECOMMENDED PRODUCT TO USE TO ACCESS THE PORT IS SAFELINE CLIP LOCK CANNULA (NF9200). CUSTOMER WAS INFORMED THAT THE DEVICE IS NOT DESIGNED TO BE USED WITH A NEEDLE AND SHOULD BE PAIRED WITH THE RECOMMENDED DEVICE(S). NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8484464 ↗MalfunctionMar 12, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED BY THE USER FACILITY: CUSTOMER REPORTS, THAT A WHITE PIECE APPEARS TO BE LODGED INSIDE THE IV TUBING RIGHT ABOVE THE LOWER Y PORT. THE NURSE TESTED A SET TO SEE IF THE SET IS BLOCKED AND SHE REPORTED THAT THE IV ADMINISTRATION SET IS NOT OCCLUDED. NO INJURIES. THE SET WAS NOT USED ON A PATIENT.
Patient problems: No Patient Involvement
MAUDE report 7332322 ↗MalfunctionDec 21, 2016Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY USER FACILITY: EVENT 2: ACCESS PORT (FOR NEEDLELESS ENTRY) LEAKING BLOOD WHEN PLACED ON IV ANGIOCATH AND BLOOD BEING DRAWN WITH A LUER-LOCK SYRINGE. NO PRIOR ACCESS FROM ACCESS PORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6196329 ↗MalfunctionDec 21, 2016Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED BY USER FACILITY: ACCESS PORT (FOR NEEDLELESS ENTRY) LEAKING BLOOD WHEN PLACED ON IV ANGIOCATH AND BLOOD BEING DRAWN WITH A LUER-LOCK SYRINGE. NO PRIOR ACCESS FROM ACCESS PORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6196320 ↗MalfunctionNov 23, 2016Device Operates Differently Than Expected; Device Operates Differently Than Expected
AS REPORTED BY USER FACILITY: LEAKING AT SPINLOCK CONNECTOR DURING BLOOD DRAW.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6124564 ↗
Other B. Braun Medical Inc. products
- Infusomat54 models · 2,783 reports · 30 recalls
- IV Administration Set153 models · 281 reports · 68 recalls
- Hemodialysis Bloodlines9 models · 2,179 reports · 14 recalls
- Outlook25 models · 195 reports · 24 recalls
- Perifix57 models · 224 reports · 20 recalls
- Extension Set41 models · 101 reports · 16 recalls
- Design Options116 models · 132 reports · 14 recalls
- Blood Administration Set24 models · 177 reports · 13 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page