THUNDERBEAT Open Extended JawRecalls & reported problems across all models
Olympus Medical Systems Corp. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 81
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls
- 0
We found 2 models of THUNDERBEAT Open Extended Jaw by Olympus Medical Systems Corp. in the FDA's device registry (GUDID). Together they are named in 81 adverse event reports in the FDA's MAUDE database, received between Oct 5, 2016 and Sep 21, 2026 — classified as 9 injuries, 72 malfunctions.
The most frequently reported specific device problem is “Break” (60 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported THUNDERBEAT Open Extended Jaw problems
Device problems
- Break60
- Detachment of Device or Device Component10
- Material Separation5
- Naturally Worn3
- Adverse Event Without Identified Device or Use Problem2
- Degraded2
- Material Fragmentation2
- Material Integrity Problem2
Patient problems
- No Consequences Or Impact To Patient41
- No Clinical Signs, Symptoms or Conditions34
- Foreign Body In Patient2
- Burn, Thermal1
- Full thickness (Third Degree) Burn1
- Hematemesis1
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
THUNDERBEAT Open Extended Jaw models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- THUNDERBEAT Open Extended Jaw TB-0920OEInstrument, ultrasonic surgical63 reports
- THUNDERBEAT Open Extended Jaw TB-0920OEELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES18 reports
Recent THUNDERBEAT Open Extended Jaw adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Naturally Worn
The customer returned the Thunderbeat device to the service center for evaluation related to a separate issue. During the evaluation, a reportable malfunction was identified: the tissue pad was worn out. There was no patient involvement associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841272 ↗MalfunctionSep 1, 2026Material Fragmentation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50546602 ↗MalfunctionMay 20, 2026Material Fragmentation
IT WAS REPORTED THE THUNDERBEAT SURGICAL INSTRUMENT'S PROBE HAD BROKEN EDGES THAT FELL WITHIN THE PATIENT ANATOMY DURING AN ABDOMINAL PERINCOL RESECTION FOR RECTAL CANER. THE DEVICE FRAGMENT WAS RETRIEVED WITH FORCEPS AND RESULTED IN A DELAY TO THE PROCEDURE OF LESS THAN 30 MINUTES. THE PROCEDURE WAS COMPLETED WITH A BACK-UP OLYMPUS DEVICE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 25234752 ↗MalfunctionMay 20, 2026Degraded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL DEVICE TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232259 ↗MalfunctionJan 27, 2026Naturally Worn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL DEVICE TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24189533 ↗MalfunctionNov 4, 2025Device Reprocessing Problem; Material Separation; Detachment of Device or Device Component; Device Reprocessing Problem; Material Separation; Detachment of Device or Device Component
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SUBJECT DEVICE TEFLON PAD WAS DAMAGED AND HAD A MISSING PORTION WHICH WAS EXPOSING THE METAL JAW. THE DISTAL END HAD INDICATIONS OF DRIED RESIDUE ON THE JAW AND PROBES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23472552 ↗MalfunctionSep 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC DEVICE EXHIBITED THAT THE TISSUE PAD IS SPLIT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23102680 ↗MalfunctionJun 9, 2025Degraded; Degraded
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE THUNDERBEAT EXHIBITED WORN OF THE TISSUE PAD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22172096 ↗MalfunctionMar 27, 2025Naturally Worn; Naturally Worn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE REAR END OF THE TISSUE PAD WAS WORN OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21706482 ↗InjuryJan 9, 2025Break; Break
IT WAS REPORTED THAT DURING AN OPEN LIVER SURGERY UNDER SEDATION, THE THUNDERBEAT'S PROBE BROKE AND APPROXIMATELY 1CM GOT STUCK IN THE LIVER. IT WAS RETRIEVED WITH NO DAMAGE TO THE SURROUNDING TISSUE AND THE PROCEDURE WAS COMPLETED USING AN OLYMPUS BACKUP DEVICE. THERE WERE NO REPORTS OF FURTHER PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21119410 ↗MalfunctionNov 8, 2024Material Separation; Material Separation
IT WAS REPORTED, THE ULTRASONIC SURGICAL DEVICE BROKE DURING USE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20642482 ↗InjurySep 17, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED, THE ULTRASONIC SURGICAL DEVICE PROBE BROKE OFF AND FELL INTO THE PATIENT CAVITY DURING A THERAPEUTIC LAPAROSCOPIC OOPHORECTOMY PROCEDURE. THERE WAS A TWO MINUTE DELAY TO RETRIEVE THE PART WITH GRASPING FORCEPS. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20241739 ↗MalfunctionSep 16, 2024Break; Break
IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE JAW PROBE WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20223339 ↗MalfunctionAug 14, 2024Break; Break
IT WAS REPORTED THE ULTRASONIC SURGICAL DEVICE BLADE BROKE OFF IN THE ABDOMINAL AREA. THE BROKEN PART WAS RETRIEVED FROM THE PATIENT BODY, RESULTING IN A SLIGHT DELAY. THE ISSUE OCCURRED DURING A THERAPEUTIC ENDOUROLOGICAL PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19983294 ↗MalfunctionJul 8, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED, WHILE USING THE ULTRASONIC SURGICAL DEVICE, AFTER 2-3 MINUTES OF INTERVENTION ONE OF THE TWO PROBES OF THE TB-SCISSOR BROKE (AND DETACHED FROM THE DEVICE). THE BROKEN PIECE FELL OUTSIDE THE PATIENT'S BODY. THE ISSUE OCCURRED DURING AN UNSPECIFIED THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. THE PATIENT DID NOT REQUIRE ADDITIONAL ANESTHESIA, THERE WAS NO PROLONGED HOSPITALIZATION AND NO OTHER INTERVENTION NECESSARY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19691676 ↗
Other Olympus Medical Systems Corp. products
- EVIS EXERA lll COLONOVIDEOSCOPE16 models · 15,013 reports · 7 recalls
- EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE7 models · 13,448 reports · 7 recalls
- EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE9 models · 8,317 reports · 2 recalls
- EVIS EXERA III BRONCHOVIDEOSCOPE11 models · 6,060 reports · 16 recalls
- EVIS EXERA III XENON LIGHT SOURCE2 models · 6,694 reports
- EVIS EXERA III COLONOVIDEOSCOPE6 models · 6,117 reports · 5 recalls
- EVIS EXERA III DUODENOVIDEOSCOPE OLYMPUS TJF-Q190V2 models · 5,281 reports · 4 recalls
- HD CAMERA HEAD2 models · 5,217 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page