THUNDERBEAT Open Extended Jaw TB-0920OE (UDI 04953170343308)Recalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of THUNDERBEAT Open Extended Jaw (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 04953170343308), received between Jun 14, 2018 and Sep 21, 2026 — classified as 1 injury, 17 malfunctions.
The most frequently reported specific device problem is “Break” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break11
- Degraded2
- Naturally Worn2
- Adverse Event Without Identified Device or Use Problem1
- Material Fragmentation1
- Material Integrity Problem1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient8
- Burn, Thermal1
- Foreign Body In Patient1
- Perforation1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 21, 2026Naturally Worn
The customer returned the Thunderbeat device to the service center for evaluation related to a separate issue. During the evaluation, a reportable malfunction was identified: the tissue pad was worn out. There was no patient involvement associated with this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841272 ↗MalfunctionSep 1, 2026Material Fragmentation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50546602 ↗MalfunctionMay 20, 2026Degraded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL DEVICE TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25232259 ↗MalfunctionJun 9, 2025Degraded; Degraded
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE THUNDERBEAT EXHIBITED WORN OF THE TISSUE PAD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22172096 ↗MalfunctionMar 27, 2025Naturally Worn; Naturally Worn
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE REAR END OF THE TISSUE PAD WAS WORN OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21706482 ↗MalfunctionNov 8, 2024Material Separation; Material Separation
IT WAS REPORTED, THE ULTRASONIC SURGICAL DEVICE BROKE DURING USE. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20642482 ↗MalfunctionSep 21, 2023Break; Break
THE CUSTOMER REPORTED THE JAW OF THE THUNDERBEAT OPEN EXTENDED JAW BROKE OFF DURING AN OPEN THERAPEUTIC HEMICOLECTOMY PROCEDURE. THE CUSTOMER NOTED THAT OTHER THUNDERBEATS WERE USED WITHOUT PROBLEMS. THE DEVICE WAS DISCARDED AT THE SITE BY THE CUSTOMER. THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17792447 ↗MalfunctionApr 9, 2021Break; Break
OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) RECEIVED THE FOLLOWING REPORT FROM THE USER. DURING THE PROCEDURE, THE PROBE OF THE SUBJECT DEVICE WAS BROKEN WITHOUT ANY REASON(FUNCTIONAL MODE: SEAL&CUT MODE 1). THE USER TOOK OUT A BROKEN TIP BY GRASPING IT FROM THE PATIENT BODY. THE PROCEDURE WAS TERMINATED WITH A REPLACEMENT OF THE SAME SET OF EQUIPMENT. BUT THE USER SUSPECTED THE DURABILITY OF THE PRODUCT BECAUSE OF THIS CASE. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11639574 ↗InjuryAug 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
WE RECEIVED THE FOLLOWING REPORT. DURING A HEMI COLECTOMY, THE SUBJECT DEVICE WAS USED. IN THE PROCEDURE, UNINTENDED CONTACT WITH THE DEVICE CAUSED BURN AND PERFORATION. THE PERFORATION WAS RECOGNIZED AT THE TIME AND ADDRESSED ACCORDINGLY. THE INTENDED PROCEDURE WAS COMPLETED WITH AN UNSPECIFIED INTERVENTION PROCEDURE. NO FURTHER INFORMATION WAS PROVIDED. THE SUBJECT DEVICE WAS DISCARDED BY THE USER.
Patient problems: Burn, Thermal; Perforation
MAUDE report 10360458 ↗MalfunctionJul 15, 2020Break; Break
DURING A COLECTOMY, THE SUBJECT DEVICE WAS USED. AFTER 30 MINUTES OF THE START OF THE PROCEDURE, THE PROBE OF THE SUBJECT DEVICE WAS BROKEN OFF AND FELL INSIDE THE PATIENT. THE FRAGMENT WAS RETRIEVED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE BREAKAGE OF THE PROBE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10275503 ↗MalfunctionMar 25, 2020Break; Break
DURING AN UNSPECIFIED PROCEDURE, THE SUBJECT DEVICE WAS USED. IN THE PROCEDURE, THE SUBJECT DEVICE STOPPED WORKING. THE TISSUE PAD WAS MELTED. THERE WAS NO PATIENT INJURY REPORTED. ON MARCH 4, 2020, OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOUND THAT THE TISSUE PAD WAS SEVERELY WORN. THIS IS THE REPORT REGARDING THE SEVERELY WORN TISSUE PAD.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9881460 ↗MalfunctionFeb 28, 2020Break; Material Integrity Problem; Break; Material Integrity Problem
AS REPORTED FOR THIS EVENT, DURING THE PROCEDURE THE DEVICE JAW BROKE OFF. THERE WAS NO ADVERSE IMPACT TO THE PATIENT. THE PROCEDURE WAS COMPLETED, COMPLETION DEVICE UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9769344 ↗MalfunctionSep 30, 2019Break; Break
DURING A HEMICOLECTOMY, THE SUBJECT DEVICE WAS USED. WHEN THE SUBJECT DEVICE WAS CONNECTED WITH THE GENERATOR, PROBE DAMAGE ERROR OCCURRED. THE USER CONTINUOUSLY DISCONNECTED AND CONNECTED THE SUBJECT DEVICE BUT THE GENERATOR ALERTED TO REPLACE THE DEVICE. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. ON (B)(6) 2019, OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOUND THAT THE COATING FOR ELECTRIC INSULATION OF THE GRASPING SECTION WAS PARTIALLY MISSING. THIS IS THE REPORT REGARDING THE MISSING OF THE GRASPING SECTION COATING.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9137741 ↗MalfunctionJul 10, 2019Break; Break
DURING A LOWER ANTERIOR RESECTION, THE SUBJECT DEVICE WAS USED. IN THE PROCEDURE, THE PROBE BROKE OFF INSIDE THE PATIENT. THE FRAGMENT WAS RETRIEVED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE BREAKAGE OF THE PROBE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8779044 ↗MalfunctionMay 10, 2019Break; Break
DURING A KNEE AMPUTATION, THE SUBJECT DEVICE WAS USED. THE BLEEDING COULD NOT BE STOPPED SINCE AN U509 ERROR OCCURRED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8598535 ↗MalfunctionOct 12, 2018Break; Break
OLYMPUS WAS INFORMED THAT THE END OF A VAGINAL HYSTERECTOMY PROCEDURE, THE PROBE AT THE DISTAL TIP OF THE DEVICE BROKE OFF AND FELL INTO THE PATIENT¿S VAGINAL CAVITY. IT WAS REPORTED THAT A PROBE ERROR OCCURRED AND THE DEVICE WAS DISCONNECTED AND RECONNECTED. A SUCCESSFUL PROBE CHECK WAS INITIATED. AFTER A PERIOD OF TIME, THE PROBE DETACHED. THERE WAS NO ADDITIONAL MEDICAL OR SURGICAL INTERVENTION REQUIRED AS THE PROBE WAS QUICKLY RETRIEVED FROM THE PATIENT BY HAND. THERE WAS NO PATIENT INJURY REPORTED. THERE WAS ALSO NO LONGER STAY, ADDITIONAL TREATMENT OR UNEXPECTED BLEEDING TO THE PATIENT REPORTED. IN ADDITION, THE DEVICE WAS INSPECTED PRIOR TO PROCEDURE WITH NO ANOMALIES OBSERVED.
Patient problems: Foreign Body In Patient
MAUDE report 7963260 ↗MalfunctionJun 14, 2018Break; Break
DURING A LOWER ANTERIOR RESECTION, THE SUBJECT DEVICE WAS USED. IN THE PROCEDURE, THE PROBE OF THE SUBJECT DEVICE BROKE OFF AND FELL INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE BREAKAGE OF THE PROBE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7604143 ↗MalfunctionJun 14, 2018Break; Break
DURING A LOWER ANTERIOR RESECTION, THE SUBJECT DEVICE WAS USED. IN THE PROCEDURE, THE PROBE OF THE SUBJECT DEVICE BROKE OFF AND FELL INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED. THE INTENDED PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO PATIENT INJURY REPORTED. THIS IS THE REPORT REGARDING THE BREAKAGE OF THE PROBE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7604139 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page