ULTRASITERecalls & reported problems across all models
B. Braun Medical Inc. · 12 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one ULTRASITE model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 112
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 12 models of ULTRASITE by B. Braun Medical Inc. in the FDA's device registry (GUDID). Together they are named in 112 adverse event reports in the FDA's MAUDE database, received between Nov 18, 2016 and Jun 9, 2026 — classified as 112 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (51 reports).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
ULTRASITE recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Sep 15, 2022Open, Classified
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.
Z-0090-2023 · Root cause: Process design · Recall details · FDA record ↗
Most reported ULTRASITE problems
Device problems
- Fluid/Blood Leak51
- Device Operates Differently Than Expected30
- Leak/Splash28
- Infusion or Flow Problem2
- Contamination /Decontamination Problem1
Patient problems
- No Consequences Or Impact To Patient52
- No Clinical Signs, Symptoms or Conditions33
- No Known Impact Or Consequence To Patient10
- Blood Loss7
- Chemical Exposure5
- Hemorrhage/Blood Loss/Bleeding4
- Exposure to Body Fluids2
ULTRASITE models
12 models, most reported first. Each model has its own page with its reports, recalls and device details.
- ULTRASITE® 474515Set, administration, intravascular1 recall0 reports
- ULTRASITE® 473438Set, administration, intravascular48 reports
- ULTRASITE® 415110Set, administration, intravascular17 reports
- ULTRASITE® US1921Set, administration, intravascular15 reports
- ULTRASITE® 473444Set, administration, intravascular6 reports
- ULTRASITE® 470021Set, administration, intravascular5 reports
- ULTRASITE® 473447Set, administration, intravascular5 reports
- ULTRASITE® 474921Set, administration, intravascular5 reports
- ULTRASITE® 473443Set, administration, intravascular4 reports
- ULTRASITE® 473448Set, administration, intravascular3 reports
- ULTRASITE® 473992Set, administration, intravascular3 reports
- ULTRASITE® 375100Set, administration, intravascular1 report
Recent ULTRASITE adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2026Leak/Splash
AS REPORTED BY THE USER FACILITY: WHILE IV CONTRAST WAS BEING ADMINISTERED, BRAUN IV TUBE CAP CAME OFF THE IV TUBING. BLOOD AND IV CONTRAST SPILLED ONTO FLOOR.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25449963 ↗MalfunctionMar 19, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: DETAILED INQUIRY DESCRIPTION: WHEN ATTACHING THE ULTRASITE TO THE IV CATHETER AND SECURE IT, BLOOD STARTS POURING OUT OF THE SITE. SO, ULTIMATELY IT IS NOT GETTING A GOOD SEAL. THEY ALL LOOK THE SAME AND WE ARE UNABLE TO TELL IF IT IS OK UNTIL WE TRY TO LOCK IT INTO THE CATHETER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21645334 ↗MalfunctionMar 16, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: DESCRIPTION SUMMARY: RN WAS FLUSHING PATIENT'S CAP (B BRAUN ULTRASITE INJECTION SITE CAP REF: 415110 LOT#0061960991) AND BLOOD WAS LEAKING OUT AT THE CONNECTION WHERE IT SHOULDN'T BE LEAKING OUT OF. ALL CAPS FROM SAME LOT# PULLED. VERIFIED LOT 0061966510 GOES WITH 490102. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21615364 ↗MalfunctionMar 15, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: THE PRODUCT KEEPS BREAKING AND THE PATIENTS ARE SPEWING BLOOD EVERYWHERE. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21613674 ↗MalfunctionFeb 10, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: THERE SEEMS TO BE AN EXTRA PIECE OF PLASTIC ATTACHED AND WHEN YOU USE IT BLOOD COME OUT OF THE PLASTIC ATTACHMENT. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21344167 ↗MalfunctionJan 31, 2025Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: BRIEF INQUIRY DESCRIPTION: LEAKING AT CONNECTION POINT BETWEEN ULTRASITE AND IVC. DETAILED INQUIRY DESCRIPTION: WHEN ATTACHING THE ULTRASITE TO THE IV CATHETER AND SECURE IT, BLOOD STARTS POURING OUT OF THE SITE. SO, ULTIMATELY IT IS NOT GETTING A GOOD SEAL. THEY ALL LOOK THE SAME AND WE ARE UNABLE TO TELL IF IT IS OK UNTIL WE TRY TO LOCK IT INTO THE CATHETER.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21286726 ↗MalfunctionOct 8, 2024Leak/Splash; Leak/Splash
AS REPORTED, THE WHITE CONNECTOR IS NOT STAYING ON THE LUER LOCK END, EVEN AFTER TIGHTENING IT HAS COME LOOSE ON MORE THAN A COUPLE OF OCCASIONS AND HAS CAUSED A CHEMO SPILL ON ONE OCCASION.
Patient problems: Chemical Exposure
MAUDE report 20398503 ↗MalfunctionOct 8, 2024Leak/Splash; Leak/Splash
AS REPORTED, EVENT 1 THE WHITE CONNECTOR IS NOT STAYING ON THE LUER LOCK END, EVEN AFTER TIGHTENING IT HAS COME LOOSE ON MORE THAN A COUPLE OF OCCASIONS AND HAS CAUSED A CHEMO SPILL ON ONE OCCASION.
Patient problems: Chemical Exposure
MAUDE report 20398448 ↗MalfunctionOct 8, 2024Leak/Splash; Leak/Splash
AS REPORTED, EVENT 4: THE WHITE CONNECTOR IS NOT STAYING ON THE LUER LOCK END, EVEN AFTER TIGHTENING IT HAS COME LOOSE ON MORE THAN A COUPLE OF OCCASIONS AND HAS CAUSED A CHEMO SPILL ON ONE OCCASION.
Patient problems: Chemical Exposure
MAUDE report 20398395 ↗MalfunctionOct 8, 2024Leak/Splash; Leak/Splash
AS REPORTED, THE WHITE CONNECTOR IS NOT STAYING ON THE LUER LOCK END, EVEN AFTER TIGHTENING IT HAS COME LOOSE ON MORE THAN A COUPLE OF OCCASIONS AND HAS CAUSED A CHEMO SPILL ON ONE OCCASION.
Patient problems: Chemical Exposure
MAUDE report 20398381 ↗MalfunctionFeb 23, 2024Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 1: BRIEF INQUIRY DESCRIPTION: LEAKING AT THE ULTRASITE AND EXT SET CONNECTION. DETAILED INQUIRY DESCRIPTION: WE HAVE HAD A HANDFUL OF 6.5" EXTENSION SETS RECENTLY THAT ARE LEAKING AND EVEN COMING COMPLETELY APART BETWEEN THE WHITE ULTRASITE INJECTION PORT AND THE DISTAL LUER-LOCK END OF THE TUBING. THIS LEADS TO BACK FLOW OF BLOOD FROM THE IV CATHETER AND POTENTIAL INFECTION RISK. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18766565 ↗MalfunctionFeb 23, 2024Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 2: BRIEF INQUIRY DESCRIPTION: LEAKING AT THE ULTRASITE AND EXT SET CONNECTION. DETAILED INQUIRY DESCRIPTION: WE HAVE HAD A HANDFUL OF 6.5" EXTENSION SETS RECENTLY THAT ARE LEAKING AND EVEN COMING COMPLETELY APART BETWEEN THE WHITE ULTRASITE INJECTION PORT AND THE DISTAL LUER-LOCK END OF THE TUBING. THIS LEADS TO BACK FLOW OF BLOOD FROM THE IV CATHETER AND POTENTIAL INFECTION RISK. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18766564 ↗MalfunctionFeb 23, 2024Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 4: BRIEF INQUIRY DESCRIPTION: LEAKING AT THE ULTRASITE AND EXT SET CONNECTION. DETAILED INQUIRY DESCRIPTION: WE HAVE HAD A HANDFUL OF 6.5" EXTENSION SETS RECENTLY THAT ARE LEAKING AND EVEN COMING COMPLETELY APART BETWEEN THE WHITE ULTRASITE INJECTION PORT AND THE DISTAL LUER-LOCK END OF THE TUBING. THIS LEADS TO BACK FLOW OF BLOOD FROM THE IV CATHETER AND POTENTIAL INFECTION RISK. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18766562 ↗MalfunctionFeb 23, 2024Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 5: BRIEF INQUIRY DESCRIPTION: LEAKING AT THE ULTRASITE AND EXT SET CONNECTION. DETAILED INQUIRY DESCRIPTION: WE HAVE HAD A HANDFUL OF 6.5" EXTENSION SETS RECENTLY THAT ARE LEAKING AND EVEN COMING COMPLETELY APART BETWEEN THE WHITE ULTRASITE INJECTION PORT AND THE DISTAL LUER-LOCK END OF THE TUBING. THIS LEADS TO BACK FLOW OF BLOOD FROM THE IV CATHETER AND POTENTIAL INFECTION RISK. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18766560 ↗MalfunctionFeb 23, 2024Leak/Splash; Leak/Splash
AS REPORTED BY THE USER FACILITY: EVENT 3: BRIEF INQUIRY DESCRIPTION: LEAKING AT THE ULTRASITE AND EXT SET CONNECTION. DETAILED INQUIRY DESCRIPTION: WE HAVE HAD A HANDFUL OF 6.5" EXTENSION SETS RECENTLY THAT ARE LEAKING AND EVEN COMING COMPLETELY APART BETWEEN THE WHITE ULTRASITE INJECTION PORT AND THE DISTAL LUER-LOCK END OF THE TUBING. THIS LEADS TO BACK FLOW OF BLOOD FROM THE IV CATHETER AND POTENTIAL INFECTION RISK. NO INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18766559 ↗
Other B. Braun Medical Inc. products
- Infusomat54 models · 2,783 reports · 30 recalls
- IV Administration Set153 models · 281 reports · 68 recalls
- Hemodialysis Bloodlines9 models · 2,179 reports · 14 recalls
- Outlook25 models · 195 reports · 24 recalls
- Perifix57 models · 224 reports · 20 recalls
- Extension Set41 models · 101 reports · 16 recalls
- Design Options116 models · 132 reports · 14 recalls
- Blood Administration Set24 models · 177 reports · 13 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page