Graphic Controls Acquisition Corp · 8 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
133
Deaths reported
0
Injuries reported
133
Recalls
6

We found 8 models of Vermed by Graphic Controls Acquisition Corp in the FDA's device registry (GUDID). Together they are named in 133 adverse event reports in the FDA's MAUDE database, received between May 25, 2018 and Sep 24, 2026 — classified as 133 injuries.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (1 report).

6 FDA recall records name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 12 reportsAug 2025: 0 reportsSep 2025: 6 reportsOct 2025: 4 reportsNov 2025: 8 reportsDec 2025: 19 reportsJan 2026: 18 reportsFeb 2026: 7 reportsMar 2026: 9 reportsApr 2026: 5 reportsMay 2026: 11 reportsJun 2026: 6 reportsJul 2026: 6 reportsAug 2026: 4 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Vermed recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 15, 2022Terminated

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

    Z-0940-2022 · Root cause: Device Design · Recall details · FDA record ↗

  2. Mar 15, 2022Terminated

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

    Z-0939-2022 · Root cause: Device Design · Recall details · FDA record ↗

  3. Mar 15, 2022Terminated

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

    Z-0938-2022 · Root cause: Device Design · Recall details · FDA record ↗

  4. Mar 15, 2022Terminated

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

    Z-0937-2022 · Root cause: Device Design · Recall details · FDA record ↗

  5. Mar 15, 2022Terminated

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

    Z-0936-2022 · Root cause: Device Design · Recall details · FDA record ↗

  6. Mar 15, 2022Terminated

    Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

    Z-0935-2022 · Root cause: Device Design · Recall details · FDA record ↗

Most reported Vermed problems

Device problems

  • Adverse Event Without Identified Device or Use Problem132
  • Patient-Device Incompatibility1

Patient problems

  • Skin Inflammation/ Irritation118
  • Itching Sensation52
  • Blister50
  • Erythema32
  • Burning Sensation27
  • Rash27
  • Burn(s)3
  • Peeling2

Vermed models

8 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Vermed adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.