Skip to content

Vermed recallZ-0937-2022 · Graphic Controls Acquisition Corporation · initiated Mar 15, 2022

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Graphic Controls Acquisition Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Instructions for Use (IFU) (Instructions for Use) indicate compliance with ANSI/AAMI EC53:2013/(R)2020 ECG trunk cables and patient leadwires. Section 5.3.9 of the standard requires the device to withstand, without breakdown , an application of, 5,000 V d.c. when used with a defibrillator at a voltage above its maximum threshold, Claravue does not meet the specification and delivers an inefficient electrical charge which may result in an ineffective defibrillation.

FDA root cause category: Device Design

What the company is doing

Graphic Controls (d/b/a Nissha Medical Technologies/Vermed/Biomedical Innovations) issued Urgent Field Correction letter via email on 3/4/22. Letter states reason for recall, health risk and action to take: Section 5.1.5 of the standard provides that trunk cables and patient leadwires may be permanently marked to indicate they are not defibrillation proof. Immediately examine your inventory and complete the provided form Claravue IP Set Correction Notice Acknowledgement Form and return to [email protected] If you have product in inventory our team will arrange for it to be returned and exchanged for product labeled correctly once available. In addition, if you may have further distributed this product, please identify your customer and notify them at once of this product correction. Your notification may be enhanced by including a copy of this correction letter. Please complete and return the attached form via e-mail to [email protected] as soon as possible. If you have any questions about this notice, please contact: Ms. Jessica Potter, Vice President Quality and Regulatory Nissha Medical Technologies, 400 Exchange Street, Buffalo, NY 14204, USA Phone: (716) 849-6367 E-Mail: [email protected]

Recall details

Recall number
Z-0937-2022
Status
Terminated
Initiated by the company
Mar 15, 2022
Recalling firm
Graphic Controls Acquisition Corporation
Quantity
19470 units
Product code
DRX
FDA submissions
K101685
Distribution
US Nationwide distribution in the States of AR, CA, CO, FL, IN, KS, MA, MN, NY, OK, VT.

Product description

CLARAVUE Pre-wired Electrode - IP SET M5 RT ADULT-US Model: 32028778 50601-US

Code information (models, lots, serial numbers)

All lots UDI: 03700506307333

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K101685 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 89810).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page