VISIONIST CRT-PRecalls & reported problems across all models

Boston Scientific Corporation · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name at least one VISIONIST CRT-P model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
271
Deaths reported
2
Injuries reported
129
Recalls
3

We found 2 models of VISIONIST CRT-P by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 271 adverse event reports in the FDA's MAUDE database, received between Oct 10, 2017 and Sep 25, 2026 — classified as 2 deaths, 129 injuries, 140 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (72 reports).

3 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 8 reportsMay 2024: 4 reportsJun 2024: 2 reportsJul 2024: 3 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 5 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 6 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 5 reportsSep 2025: 0 reportsOct 2025: 6 reportsNov 2025: 4 reportsDec 2025: 7 reportsJan 2026: 8 reportsFeb 2026: 3 reportsMar 2026: 4 reportsApr 2026: 14 reportsMay 2026: 4 reportsJun 2026: 9 reportsJul 2026: 3 reportsAug 2026: 5 reportsSep 2026: 12 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

VISIONIST CRT-P recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 20, 2025Open, Classified

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

    Z-0104-2026 · Root cause: Other · Recall details · FDA record ↗

  2. Aug 20, 2025Open, Classified

    Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).

    Z-0103-2026 · Root cause: Other · Recall details · FDA record ↗

  3. Dec 12, 2024Open, Classified

    A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

    Z-0827-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported VISIONIST CRT-P problems

Device problems

  • Premature Discharge of Battery72
  • Adverse Event Without Identified Device or Use Problem49
  • High impedance42
  • Over-Sensing35
  • Signal Artifact/Noise30
  • Pacing Problem27
  • Incorrect, Inadequate or Imprecise Result or Readings24
  • Battery Problem22

Patient problems

  • No Clinical Signs, Symptoms or Conditions161
  • Unspecified Infection41
  • No Consequences Or Impact To Patient16
  • No Known Impact Or Consequence To Patient10
  • Pocket Erosion8
  • Sepsis7
  • Syncope/Fainting7
  • Asystole6

VISIONIST CRT-P models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent VISIONIST CRT-P adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Boston Scientific Corporation products

All Boston Scientific Corporation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page