VISIONIST CRT-P recallZ-0827-2025 · Boston Scientific Corporation · initiated Dec 12, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.
FDA root cause category: Process design
What the company is doing
The firm sent an "URGENT MEDICAL DEVICE INFORMATION" letter dated December 2024. The letter describes the product, problem and actions to be taken. The customers were instructed to promptly identify patients, if device enters Safety Mode, perform emergent replacement for patients who are at risk for harm, during replacement of a device in Safety Mode pacing inhibition should be anticipated during electrocautery. General prophylactic replacement is not recommended; schedule device replacement promptly when the longevity remaining reaches four (4) years or if the longevity remaining is less than 4 years, perform system follow-ups in accordance with the instructions for use and update patients records with letter to maintain awareness for the remaining service life of the device. If you have any questions regarding this information or would like to report a clinical event, contact your Boston Scientific representative or Technical Services. Supplement March 26, 2025, High-Level Overview of Recall Strategy: on 03/26/2025, BSC notified only those implanting and following clinicians responsible for the newly identified active advisory devices. The letter sent to these clinicians will be the original December 12, 2024, Product Advisory letter, however, the March 2025 communication included a cover letter stating that new devices were identified that should have been initially categorized in the Advisory population.
Recall details
- Recall number
- Z-0827-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 12, 2024
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 156,329 units (59,937 US, 96,792 OUS) in total
- Product code
- NKE
- FDA submissions
- P030005
- Distribution
- US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.
Product description
VISIONIST CRT-P, EL (Model Number U225 and U226) and EL MRI (Model Number U228)
Code information (models, lots, serial numbers)
Show the full code information (56k characters)
U225 - UDI-DI: 00802526559433, 00802526577116, 00802526577048, 00802526578809, Serial Numbers: 100320 100321 100388 100543 100545 100712 100717 100720 100734 100747 100787 100800 100804 100806 100807 100810 100819 100827 100828 100832 100834 100835 100845 100849 100850 100851 100852 100853 100856 100858 100865 100874 100879 100880 100887 100888 100889 100890 100892 100894 100895 100900 100901 100902 100905 100906 100907 100908 100913 100915 100916 100917 100919 100921 100922 100923 100924 100926 100928 100929 100930 100931 100932 100933 100934 700015 700016 700035 700039 700040 700043 700044 700047 700048 700052 700056 700060 700063 700064 700067 700071 700072 700075 700077 700080 700082 700084 700085 700088 700089 700092 700094 700099 700100 700102 700103 700104 700105 700108 700109 700110 700115 700121 700125 700128 700132 700134 700145 700148 700153 700160 700165 700170 700180 700188 700189 700246 700249 700268 700271 700274 700278 700279 700281 700285 700288 700291 700292 700296 700297 700298 700301 700302 700303 700304 700305 700308 700309 700310 700311 700313 700314 700315 700323 700324 700327 700328 700329 700330 700331 700332 700333 700334 700336 700337 700338 700339 700341 700342 700343 700346 700348 700349 700350 700352 700353 700354 700355 700356 700357 700361 700362 700363 700364 700365 700367 700368 700370 700371 700372 700374 700375 700376 700377 700378 700379 700381 700382 700383 700384 700386 700387 700388 700391 700392 700393 700394 700396 700397 700398 700403 700405 700406 700407 700408 700409 700410 700411 700412 700414 700415 700416 700417 700418 700419 700420 700421 700422 700423 700424 700425 700427 700428 700429 700430 700431 700432 700434 700435 700437 700438 700439 700440 700441 700442 700443 700444 700445 700446 700447 700448 700449 700450 700451 700452 700453 700454 700455 700457 700458 700459 700461 700462 700463 700464 700465 700466 700467 700468 700469 700470 700471 700472 700473 700475 700476 700477 700478 700479 700480 700484 700491 700496 700497 700498 700499 700500 700503 700504 700506 700508 700510 700511 700512 700513 700514 700515 700516 700522 700523 700524 700525 700526 700527 700528 700529 700530 700531 700532 700534 700535 700536 700537 700538 700540 700541 700542 700543 700544 700545 700546 700547 700548 700562 700566 700568 700579 700580 700581 700583 700584 700585 700587 700588 700590 700591 700592 700593 700594 700595 700596 700597 700598 700599 700600 700603 700604 700606 700607 700609 700610 700611 700612 700613 700614 700617 700620 700622 700623 700624 700625 700626 700627 700628 700629 700631 700632 700633 700635 700637 700640 700641 700642 700643 700644 700645 700647 700648 700652 700653 700654 700658 700660 700663 700664 700665 700666 700667 700668 700670 700671 700673 700674 700676 700677 700679 700680 700681 700682 700683 700684 700693 700694 700695 700696 700698 700699 700704 700705 700710 700711 700712 700714 700715 700716 700717 700718 700719 700720 700722 700723 700724 700725 700727 700728 700729 700730 700731 700732 700733 700734 700736 700737 700739 700740 700741 700742 700743 700745 700748 700749 700750 700751 700752 700753 700754 700755 700756 700757 700758 700760 700761 700764 700766 700767 700768 700772 700773 700774 700775 700776 700778 700779 700780 700781 700783 700785 700788 700789 700790 700791 700793 700794 700795 700797 700799 700803 700804 700807 700808 700809 700810 700812 700813 700815 700817 700819 700821 700822 700824 700825 700826 700827 700829 700830 700835 700836 700840 700841 700842 700844 700845 700849 700855 700857 700862 700864 700865 700866 700868 700870 700871 700872 700877 700881 700885 700886 700887 700888 700889 700890 700891 700892 700893 700894 700895 700896 700897 700899 700900 700901 700902 700903 700904 700905 700907 700908 700909 700910 700911 700912 700913 700914 700915 700916 700917 700918 700919 700920 700921 700922 700923 700924 700925 700926 700927 700929 700933 700934 700935 700936 700937 700938 700939 700940 700941 700942 700945 700947 700948 700949 700950 700951 700953 700954 700955 700957 700958 700959 700962 700963 700964 700966 700967 700968 700969 700970 700971 700972 700973 700974 700975 700976 700977 700979 700980 700981 700983 700986 700987 700988 700989 700990 700991 700992 700993 700995 700996 700997 700998 701000 701001 701004 701006 701007 701009 701010 701011 701013 701016 701017 701018 701019 701020 701021 701022 701025 701026 701027 701028 701031 701035 701037 701038 701039 701040 701042 701044 701045 701046 701047 701049 701051 701052 701053 701054 701057 701059 701061 701063 701066 701069 701070 701071 701073 701074 701075 701076 701078 701079 701080 701083 701085 701086 701087 701088 701089 701090 701092 701093 701094 701095 701096 701097 701098 701099 701100 701101 701106 701107 701108 701109 701111 701112 701113 701116 701118 701119 701120 701121 701122 701123 701124 701126 701127 701128 701129 701130 701131 701132 701133 701134 701135 701138 701139 701140 701141 701142 701143 701145 701146 701148 701149 701151 701152 701154 701155 701158 701159 701161 701162 701163 701164 701166 701167 701169 701170 701172 701173 701174 701175 701176 701177 701199 701202 701203 701204 701205 701206 701207 701209 701210 701233 701234 701235 701237 701238 701433 701487 701488 701489 701490 701491 701492 701493 701494 701515 701516 701518 701521 701522 701523 701524 701526 701527 701528 701530 701532 701544 701546 701547 701548 701549 701550 701551 701552 701553 701554 701555 701556 701557 701558 701567 701721 701736 701753 701785 701786 701792 701796 701797 701800 701801 701802 701805 701809 701811 701812 701813 701814 701815 701819 701821 701824 701825 701826 701830 701831 701832 701835 701836 701838 701839 701843 701846 701849 701853 701855 701859 701860 701861 701862 701863 701864 701866 701868 701875 701878 701884 701888 701890 701892 701893 701894 701898 701899 701900 701901 701903 701905 701907 701911 701914 701918 701920 701923 701924 701926 701929 701930 701935 701936 701937 701938 701939 701940 701941 701942 701944 701946 701947 701948 701949 701952 701953 701955 701957 701959 701960 701961 701962 701968 701977 701984 701996 702007 702008 702041 702063 702070 702073 702135 702136 702147 702154 702155 702158 702159 702216 702217 702218 702239 702241 702243 702244 702252 702253 702254 702257 702258 702259 702286 702287 702290 702294 702295 702296 702297 702298 702301 702302 702303 702304 702327 702330 702334 702366 702368 702370 702371 702372 702373 702374 702377 702382 702415 702421 702423 702424 702428 702429 702437 702442 702443 702444 702448 702454 702462 702463 702464 702465 702471 702477 702483 702501 702509 702510 702514 702521 702522 702523 702527 702531 702536 702545 702546 702547 702549 702552 702555 702557 702558 702561 702562 702566 702570 702573 702576 702579 702580 702581 702582 702584 702586 702590 702591 702593 702595 702600 702601 702603 702604 702606 702610 702615 702617 702618 702623 702624 702626 702631 702632 702636 702637 702641 702643 702651 702653 702654 702655 702658 702660 702662 702664 702666 702667 702668 702669 702672 702673 702678 702682 702685 702689 702690 702698 702699 702700 702701 702703 702705 702706 702708 702709 702711 702712 702713 702714 702717 702719 702721 702722 702723 702725 702726 702727 702728 702730 702733 702735 702745 702752 702753 702754 702755 702756 702764 702765 702770 702773 702781 702782 702784 702786 702789 702790 702793 702794 702796 702797 702799 702800 702805 702807 702811 702815 702816 702818 702819 702822 702823 702824 702825 702826 702829 702831 702834 702835 702837 702838 702841 702843 702844 702846 702847 702848 702849 702850 702854 702860 702861 702862 702864 702865 702867 702869 702871 702873 702874 702875 702878 702882 702883 702884 702885 702886 702887 702888 702890 702905 702907 702908 702910 702926 702928 702932 702933 702935 702936 702937 702938 702939 702940 702945 702948 702949 702951 702955 702956 702958 702959 702962 702963 702965 702969 702975 702979 702980 702981 702983 702988 702989 702994 702997 703006 703007 703018 703019 703023 703024 703026 703035 703036 703038 703040 703042 703043 703045 703046 703049 703053 703055 703058 703064 703068 703070 703100 703104 703108 703111 703115 703124 703125 703137 703161 703162 703174 703186 703188 703190 703192 703196 703201 703202 703203 703205 703210 703216 703217 703221 703224 703232 703237 703238 703242 703246 703258 703262 703263 703265 703269 703270 703271 703275 703276 703278 703280 703283 703286 703292 703296 703298 703299 703300 703302 703303 703306 703308 703309 703310 703311 703312 703316 703318 703321 703332 703339 703347 703354 703358 703363 703368 703370 703372 703373 703377 703378 703382 703389 703393 703394 703395 703396 703397 703398 703402 703403 703407 703408 703412 703413 703415 703416 703419 703426 703429 703430 703431 703432 703433 703436 703437 703439 703444 703445 703448 703452 703454 703456 703464 703465 703466 703467 703471 703472 703473 703476 703477 703480 703482 703483 703484 703486 703493 703496 703502 703504 703508 703509 703510 703511 703512 703514 703515 703516 703518 703519 703520 703525 703526 703527 703528 703529 703530 703535 703536 703538 703539 703540 703541 703543 703545 703547 703548 703550 703551 703552 703554 703555 703556 703557 703558 703560 703563 703564 703565 703566 703567 703568 703569 703570 703571 703572 703573 703574 703575 703576 703578 703579 703580 703581 703582 703583 703584 703585 703586 703587 703589 703591 703592 703593 703594 703596 703599 703600 703601 703602 703603 703606 703608 703609 703610 703611 703615 703618 703619 703620 703621 703622 703623 703625 703626 703627 703628 703629 703630 703631 703633 703634 703635 703636 703637 703638 703640 703641 703645 703646 703647 703648 703650 703651 703652 703655 703658 703659 703662 703663 703664 703666 703667 703668 703669 703672 703673 703674 703676 703678 703682 703685 703686 703688 703694 703695 703696 703697 703698 703699 703700 703703 703704 703707 703710 703719 703722 703723 703729 703730 703732 703739 703740 703741 703747 703750 703755 703765 703778 703779 703784 703790 703795 703808 703810 703823 703826 703827 703847 703850 703869 703871 703872 703873 703875 703876 703879 703880 703881 703882 703883 703886 703888 703890 703894 703895 703900 703907 703912 703913 703916 703923 703925 703927 703930 703932 703935 703939 703951 703953 703954 703955 703968 703970 703975 703977 703981 703983 703985 703998 704008 704011 704013 704019 704022 704025 704033 704034 704040 704046 704049 704053 704057 704062 704063 704065 704066 704068 704070 704071 704073 704079 704081 704082 704083 704084 704087 704212 704216 704240 704242 704243 704247 704249 704254 704258 704259 704261 704264 704266 704267 704270 704271 704274 704275 704308 704310 704313 704317 704318 704319 704322 704328 704329 704331 704334 704335 704336 704338 704343 704346 704348 704350 704353 704356 704358 704359 704360 704361 704362 704363 704364 704366 704367 704370 704371 704372 704373 704374 704375 704376 704377 704379 704380 704381 704382 704383 704384 704385 704386 704387 704388 704389 704390 704391 704392 704393 704395 704397 704399 704400 704401 704403 704404 704405 704406 704407 704408 704418 704419 704420 704421 704422 704423 704424 704426 704427 704431 704435 704436 704440 704446 704451 704452 704454 704455 704457 704486 704600 704601 704606 704609 704613 704618 704619 704640 704642 704644 704654 704765 704766 704767 704789 704795 704797 704798 704799 704807 704810 704812 704817 704818 704823 704825 704827 704832 704850 704853 704854 704878 704879 704929 705000 705003 705023 705025 705090 705098 705099 705101 705104 705105 705106 705119 705127 705129 705131 705133 705135 705138 705139 705141 705142 705144 705147 705149 705150 705151 705153 705154 705157 705158 705176 705178 705179 705181 705182 705183 705184 705185 705186 705188 705189 705192 705194 705195 705196 705198 705199 705200 705201 705202 705203 705204 705205 705206 705207 705208 705209 705213 705214 705215 705216 705217 705218 705219 705222 705223 705224 705228 705229 705231 705232 705233 705234 705235 705237 705238 705240 705242 705243 705244 705245 705246 705249 705261 705262 705263 705264 705265 705266 705267 705269 705272 705278 705279 705282 705283 705285 705289 705290 705293 705294 705295 705297 705298 705300 705303 705305 705308 705310 705312 705314 705315 705316 705317 705318 705319 705323 705324 705325 705326 705327 705329 705339 705463 705465 705466 705491 705496 705503 705505 705506 705507 705512 705513 705515 705516 705517 705518 705519 705527 705529 705534 705536 705539 705542 705544 705546 705549 705555 705556 705561 705565 705566 705578 705584 705585 705586 705591 705592 705593 705600 705601 705602 705643 705652 705664 705665 705668 705670 705671 705673 705675 705677 705678 705679 705682 705689 705695 705697 705698 705704 705706 705710 705717 705751 705756 705767 705771 705772 705774 705776 705777 705778 705780 705791 705795 705798 705801 705802 705807 705810 705813 705814 705815 705823 705850 705853 705864 705869 705870 705872 705881 705906 705908 705939 705946 706000 706004 706005 706008 706009 706012 706013 706015 706017 706018 706024 706031 706032 706034 706035 706036 706037 706038 706039 706044 706050 706053 706057 706058 706060 706061 706062 706064 706065 706068 706069 706071 706072 706073 706074 706077 706079 706080 706081 706085 706086 706087 706091 706094 706098 706106 706108 706122 706123 706126 706127 706150 706151 706152 706153 706154 706156 706157 706158 706159 706160 706161 706162 706163 706164 706165 706167 706168 706169 706172 706176 706177 706178 706179 706181 706182 706183 706184 706186 706187 706188 706192 706194 706195 706196 706203 706204 706205 706210 706213 706218 706219 706222 706223 706224 706225 706226 706227 706228 706233 706234 706235 706238 706239 706240 706241 706242 706246 706247 706249 706256 706270 706271 706272 706274 706276 706277 706279 706280 706290 706296 706298 706308 706311 706312 706315 706317 706318 706325 706330 706338 706346 706356 706357 706359 706364 706365 706366 706368 706369 706370 706378 706379 706417 706426 706427 706430 706437 706441 706443 706451 706462 706468 706470 706478 706487 706491 706492 706496 706497 706520 706528 706550 706556 706558 706561 706563 706566 706568 706569 706571 706575 706579 706580 706582 706584 706594 706606 706611 706638 706639 706644 706812 706813 706908 707251 707261 707262 707264 707267 707270 707273 707278 707298 707327 707329 707336 707337 707338 707342 707344 707346 707347 707348 707354 707356 707362 707366 707397 707501 707502 707506 707507 707509 707510 707521 707522 707523 707524 707525 707542 707543 707544 707545 707546 707548 707550 707551 707553 707554 707558 707560 707569 707572 707576 707577 707578 707580 707583 707584 707606 707609 707613 707615 707644 707659 707768 707769 707778 707779 707808 707810; U226 - UDI-DI: 00802526559457, 00802526559464, 00802526577062, 00802526577123, Serial Numbers: 700333 700460 700461 700484 700490 700498 700587 700591 100052 100053 100054 100055 100058 100059 100060 100061 100062 100063 100066 100067 100068 100069 100070 100107 100108 100109 100111 100112 100113 100116 100117 100118 100120 100123 100124 100126 100127 100128 100129 100132 100133 100134 100135 100136 700002 700003 700004 700006 700007 700008 700009 700010 700011 700012 700013 700014 700015 700018 700019 700020 700021 700023 700024 700025 700026 700030 700031 700032 700033 700034 700035 700036 700037 700039 700040 700042 700043 700044 700045 700048 700050 700051 700052 700087 700088 700089 700091 700092 700107 700108 700109 700110 700111 700113 700114 700118 700120 700121 700122 700123 700124 700125 700126 700132 700135 700137 700139 700140 700141 700142 700143 700145 700146 700147 700149 700151 700153 700154 700155 700156 700159 700172 700183 700184 700186 700187 700191 700192 700193 700194 700198 700199 700201 700205 700207 700209 700210 700211 700212 700213 700214 700215 700216 700218 700219 700220 700222 700225 700226 700227 700229 700230 700231 700232 700236 700237 700239 700240 700241 700245 700246 700247 700249 700264 700265 700266 700267 700268 700269 700270 700275 700276 700277 700278 700279 700280 700282 700284 700286 700287 700288 700289 700290 700291 700292 700293 700296 700298 700299 700300 700301 700302 700304 700305 700306 700309 700310 700311 700312 700313 700314 700316 700318 700319 700321 700323 700325 700326 700327 700449 700453 700454 700467 700470 700471 700473 700474 700475 700476 700477 700478 700479 700480 700482 700485 700487 700566 700575 700576 700579 700580 700581 700584 700585 700588 700590 700593 700594 700595 700596 700597 700599; U228 - UDI-DI: 00802526577130, 00802526559488, 00802526578830, Serial Numbers: 702970 709765 100745 100815 100824 100836 100851 100868 100877 100890 100891 100892 100894 100901 100902 100904 100919 100920 100923 100926 100928 100942 100944 100950 100951 100963 100976 100986 101029 101030 101031 101033 101034 101036 101068 101071 101085 101105 101150 101151 101152 101153 101155 101156 101191 101192 101197 101198 101209 101215 101216 101217 101218 101252 101262 101270 101286 101287 101288 101339 101340 101342 101346 101349 101351 101368 101379 101380 101382 101383 101385 101386 101388 101394 101396 101397 101399 101400 101405 101406 101408 101409 101439 101440 101441 101443 101444 101447 101453 101456 101458 101461 101483 101490 101643 101684 101685 101686 101688 101876 101899 101900 101901 101902 101903 101905 101906 101908 101910 101912 101913 101914 101915 101916 101918 101920 101922 101923 101925 101926 101927 101928 101929 101931 101933 101936 101937 101938 101939 101940 101941 101942 101943 101945 101946 101947 101949 101952 101953 101956 101957 101958 101959 101960 101961 101967 101969 101971 101973 101975 101976 101978 101980 101990 101991 101994 101995 101996 101997 101999 102034 102041 102046 102080 102083 102084 102086 102094 102095 102096 102098 102105 102122 102124 102125 102130 102133 102164 102181 102183 102184 102187 102188 102192 102193 102241 102256 102259 102264 102269 102270 102303 102304 102307 102310 102321 102322 102323 102326 102331 102332 102335 102336 102338 102339 102341 102343 102344 102345 102347 102348 102357 102358 102360 102361 102362 102364 102365 102368 102369 102370 102372 102373 102375 102376 102377 102378 102379 102385 102391 102392 102393 102394 102395 102396 102402 102406 102408 102438 102454 102456 102460 102479 102482 102483 700001 700002 700004 700005 700006 700008 700010 700011 700012 700014 700017 700018 700019 700020 700021 700024 700025 700026 700027 700028 700029 700030 700031 700032 700033 700035 700036 700037 700038 700040 700041 700042 700043 700046 700047 700048 700049 700050 700051 700052 700053 700054 700055 700056 700057 700058 700059 700060 700061 700062 700063 700064 700067 700068 700069 700072 700073 700074 700075 700076 700077 700078 700079 700080 700081 700082 700083 700085 700087 700089 700092 700094 700095 700096 700098 700100 700101 700102 700104 700106 700107 700108 700109 700111 700113 700114 700115 700116 700117 700118 700119 700120 700121 700122 700123 700124 700125 700127 700128 700129 700131 700133 700134 700135 700137 700138 700139 700140 700141 700142 700143 700145 700146 700147 700148 700149 700150 700151 700152 700153 700154 700155 700156 700157 700158 700161 700162 700163 700164 700165 700167 700170 700171 700172 700173 700174 700176 700178 700179 700181 700182 700183 700184 700198 700201 700202 700204 700210 700211 700214 700215 700216 700218 700220 700221 700224 700225 700226 700228 700229 700232 700234 700235 700236 700237 700238 700239 700240 700243 700246 700247 700255 700259 700276 700282 700284 700317 700333 700335 700338 700342 700347 700368 700375 700376 700377 700382 700383 700384 700385 700386 700406 700414 700419 700420 700423 700430 700431 700442 700444 700462 700463 700464 700468 700469 700470 700471 700472 700473 700475 700476 700478 700479 700482 700483 700484 700485 700488 700489 700492 700493 700494 700496 700497 700498 700499 700502 700503 700505 700506 700509 700510 700511 700512 700513 700514 700515 700516 700517 700518 700521 700522 700523 700525 700526 700527 700528 700529 700533 700534 700535 700537 700541 700543 700544 700545 700551 700552 700553 700554 700556 700557 700558 700560 700567 700575 700576 700578 700631 700644 700646 700663 700664 700667 700670 700698 700701 700703 700704 700710 700712 700725 700812 700819 700836 700838 700899 700908 700974 700990 701015 701017 701033 701034 701037 701040 701043 701048 701055 701060 701082 701085 701086 701089 701090 701093 701100 701103 701104 701107 701118 701120 701122 701126 701128 701148 701170 701184 701191 701192 701197 701213 701220 701222 701229 701238 701239 701243 701252 701257 701265 701269 701270 701298 701301 701319 701346 701351 701353 701368 701371 701374 701375 701377 701378 701382 701384 701385 701399 701427 701432 701444 701446 701451 701454 701457 701461 701466 701468 701475 701476 701477 701484 701485 701487 701488 701491 701494 701496 701500 701517 701523 701527 701531 701533 701538 701539 701544 701550 701555 701560 701580 701584 701594 701601 701605 701608 701613 701616 701618 701630 701633 701649 701650 701653 701658 701659 701671 701672 701674 701677 701682 701690 701691 701701 701705 701706 701709 701711 701719 701721 701726 701730 701734 701735 701737 701742 701743 701748 701749 701757 701778 701780 701783 701790 701794 701795 701798 701799 701800 701801 701803 701804 701805 701806 701807 701810 701813 701814 701818 701823 701830 701835 701848 701889 701890 701891 701896 701897 701903 701926 701928 701929 701933 701935 701937 701938 701958 701959 701960 701967 701970 701972 702043 702047 702088 702100 702102 702127 702129 702130 702133 702135 702137 702138 702141 702142 702146 702150 702151 702157 702158 702160 702162 702164 702165 702178 702184 702188 702190 702195 702197 702222 702223 702224 702232 702235 702247 702249 702263 702267 702279 702280 702300 702303 702304 702305 702306 702307 702308 702323 702324 702327 702330 702353 702354 702358 702367 702369 702370 702374 702375 702379 702380 702384 702417 702424 702436 702439 702472 702511 702513 702517 702518 702520 702521 702524 702527 702529 702537 702571 702577 702578 702583 702585 702591 702592 702594 702597 702602 702604 702609 702615 702617 702625 702629 702631 702634 702635 702636 702639 702642 702643 702651 702654 702661 702664 702666 702669 702673 702675 702682 702687 702688 702689 702692 702699 702700 702707 702711 702712 702720 702723 702724 702725 702730 702734 702749 702750 702752 702755 702756 702764 702773 702774 702776 702780 702783 702788 702794 702795 702797 702799 702805 702808 702811 702816 702819 702820 702821 702826 702835 702839 702841 702842 702843 702847 702849 702850 702851 702852 702856 702859 702865 702867 702868 702869 702870 702872 702880 702890 702926 702934 702943 702951 702953 702964 702966 702972 702977 702983 703016 703020 703022 703023 703036 703040 703043 703045 703066 703069 703071 703074 703078 703081 703083 703088 703089 703093 703095 703097 703100 703102 703103 703107 703113 703121 703123 703125 703127 703144 703145 703149 703153 703154 703155 703156 703160 703163 703172 703174 703178 703179 703181 703183 703187 703189 703190 703191 703193 703194 703197 703199 703200 703203 703206 703208 703209 703211 703212 703213 703214 703216 703217 703218 703220 703223 703230 703233 703234 703235 703236 703238 703239 703240 703241 703242 703243 703244 703248 703251 703254 703257 703259 703261 703263 703264 703266 703267 703270 703271 703272 703273 703275 703276 703277 703278 703280 703282 703284 703285 703286 703287 703288 703290 703293 703295 703298 703299 703300 703301 703302 703303 703304 703305 703306 703307 703308 703309 703310 703311 703312 703314 703316 703317 703318 703319 703320 703321 703322 703325 703326 703329 703330 703332 703333 703334 703335 703336 703337 703342 703343 703346 703348 703350 703353 703354 703358 703359 703360 703361 703362 703363 703364 703366 703367 703368 703369 703370 703371 703375 703380 703381 703382 703383 703386 703387 703388 703389 703390 703391 703392 703393 703394 703395 703396 703398 703401 703403 703406 703408 703409 703410 703414 703415 703416 703420 703428 703429 703430 703433 703438 703439 703441 703447 703448 703449 703450 703466 703469 703470 703472 703474 703476 703487 703489 703499 703500 703501 703502 703509 703518 703519 703525 703531 703532 703534 703547 703550 703557 703558 703559 703568 703573 703574 703579 703581 703582 703583 703584 703586 703608 703611 703612 703613 703614 703615 703616 703617 703620 703624 703627 703628 703630 703631 703632 703633 703635 703636 703638 703642 703643 703645 703646 703652 703653 703654 703657 703659 703664 703668 703672 703674 703675 703676 703678 703682 703683 703684 703686 703692 703693 703694 703696 703698 703699 703701 703702 703704 703707 703709 703710 703711 703713 703714 703715 703716 703717 703721 703723 703727 703728 703729 703732 703734 703735 703736 703738 703739 703743 703744 703747 703748 703751 703754 703756 703757 703758 703762 703764 703765 703766 703768 703771 703778 703780 703782 703786 703789 703790 703794 703795 703803 703805 703806 703812 703813 703817 703820 703821 703822 703824 703825 703833 703836 703837 703841 703843 703845 703848 703849 703850 703855 703856 703857 703860 703861 703863 703864 703868 703869 703872 703873 703875 703876 703877 703879 703881 703906 703908 703911 703912 703916 703917 703918 703919 703920 703922 703927 703929 703930 703931 703935 703936 703941 703942 703943 703945 703948 703949 703952 703954 703957 703959 703960 703962 703963 703964 703971 703972 703974 703975 703982 703983 703984 703985 703986 703987 703988 703989 703990 703992 703993 703994 703995 703996 704001 704004 704006 704011 704014 704015 704016 704017 704019 704020 704021 704022 704023 704024 704025 704027 704030 704035 704036 704037 704038 704039 704040 704042 704045 704047 704048 704050 704051 704054 704056 704057 704060 704062 704063 704065 704068 704075 704076 704079 704083 704085 704087 704089 704090 704092 704093 704094 704095 704096 704098 704100 704102 704103 704104 704105 704107 704108 704109 704113 704114 704115 704117 704118 704120 704121 704123 704124 704127 704129 704132 704133 704134 704138 704140 704142 704143 704144 704145 704146 704149 704150 704151 704152 704154 704156 704159 704161 704164 704165 704166 704168 704171 704174 704175 704176 704177 704178 704179 704181 704182 704183 704185 704186 704187 704190 704192 704193 704195 704196 704197 704198 704200 704202 704205 704209 704222 704234 704237 704260 704262 704277 704307 704355 704358 704359 704365 704366 704387 704388 704389 704391 704398 704403 704432 704592 704606 704608 704615 704616 704619 704620 704623 704625 704633 704634 704635 704636 704637 704639 704648 704650 704667 704671 704672 704674 704678 704686 704689 704690 704691 704697 704705 704708 704709 704711 704713 704716 704717 704718 704719 704723 704724 704728 704751 704752 704759 704767 704771 704772 704775 704782 704786 704789 704790 704791 704808 704819 704833 704838 704842 704846 704847 704848 704849 704851 704855 704868 704873 704878 704892 704894 704925 704962 704963 704966 704969 704970 704975 704982 704983 704999 705004 705023 705026 705027 705028 705029 705047 705050 705074 705075 705077 705080 705083 705613 705617 706134 706203 706229 706231 706235 706598 706606 706608 706610 706615 706616 706619 706622 706623 706624 706625 706628 706629 706630 706631 706632 706634 706635 706636 706638 706641 706642 706643 706645 706647 706649 706651 706653 706654 706655 706657 706658 706659 706661 706663 706665 706667 706668 706669 706670 706671 706674 706675 706676 706677 706678 706680 706681 706684 706685 706686 706688 706689 706690 706691 706693 706694 706696 706698 706701 706704 706705 706708 706709 706710 706711 706713 706717 706718 706720 706722 706727 706728 706730 706731 706733 706738 706740 706742 706743 706744 706748 706749 706750 706751 706752 706755 706756 706758 706786 706787 706790 706791 706795 706798 706799 706800 706806 706810 706811 706814 706815 706817 706819 706820 706822 706823 706824 706825 706830 706836 706837 706838 706840 706843 706846 706848 706849 706850 706851 706852 706854 706855 706857 706859 706860 706864 706866 706867 706868 706870 706871 706872 706874 706875 706877 706879 706880 706887 706888 706897 706900 706901 706903 706904 706905 706906 706910 706911 706912 706913 706914 706922 706923 706924 706925 706941 706942 706944 706945 706959 706960 706962 706963 706966 706972 706973 706975 706978 706980 706981 706983 706984 706988 706989 706990 706993 706995 706996 706997 706998 707000 707002 707006 707012 707013 707016 707017 707018 707032 707033 707034 707036 707039 707042 707043 707044 707045 707047 707050 707051 707053 707055 707059 707061 707062 707063 707064 707066 707067 707068 707069 707070 707073 707075 707076 707082 707106 707109 707111 707112 707123 707128 707130 707140 707142 707143 707144 707151 707152 707153 707154 707175 707176 707178 707179 707180 707298 707300 707302 707306 707316 707317 707320 707321 707323 707326 707346 707349 707350 707352 707355 707362 707365 707367 707368 707370 707372 707374 707375 707379 707380 707382 707388 707393 707397 707401 707402 707404 707406 707408 707410 707413 707414 707415 707416 707417 707419 707421 707422 707423 707425 707427 707428 707432 707433 707435 707436 707438 707442 707444 707449 707456 707459 707460 707462 707464 707465 707467 707470 707473 707475 707476 707481 707482 707483 707484 707488 707489 707490 707491 707493 707495 707499 707505 707523 707526 707529 707533 707536 707540 707544 707629 707641 707645 707647 707673 707683 707685 707686 707689 707691 707695 707704 707705 707716 707718 707721 707723 707729 707732 707740 707741 707745 707756 707760 707765 707766 707769 707772 707775 707776 707783 707795 707821 707829 707834 707841 707843 707873 707883 707884 707885 707886 707887 707890 707952 707983 707984 707985 707988 707989 707990 707992 707998 708002 708024 708025 708033 708035 708036 708039 708054 708063 708072 708073 708081 708083 708084 708085 708086 708087 708088 708090 708091 708092 708093 708097 708098 708099 708100 708101 708103 708107 708113 708114 708119 708120 708121 708123 708125 708126 708127 708128 708129 708131 708133 708134 708135 708137 708138 708139 708141 708143 708144 708148 708149 708150 708155 708157 708158 708160 708161 708163 708170 708171 708172 708173 708178 708179 708180 708181 708189 708194 708200 708201 708206 708209 708212 708214 708217 708219 708220 708222 708223 708229 708233 708315 708318 708319 708320 708321 708322 708323 708336 708345 708350 708352 708527 708529 708607 708615 708628 708631 708635 708637 708639 708644 708645 708646 708648 708651 708652 708653 708654 708664 708669 708680 708686 708690 708691 708692 708693 708695 708697 708701 708703 708704 708705 708710 708712 708713 708716 708718 708759 708760 708762 708772 708774 708775 708778 708792 708800 708803 708804 708806 708811 708814 708815 708816 708817 708819 708827 708830 708832 708838 708840 708844 708846 708848 708854 708855 708856 708864 708877 708878 708879 708889 708892 708898 708905 708908 708913 708919 708920 708923 708934 708939 708945 708950 708952 708953 708955 708958 708966 708968 708969 708971 708973 708974 708976 708980 708981 708982 708984 708986 708991 708992 708995 708997 708999 709000 709010 709011 709012 709016 709018 709019 709021 709022 709023 709027 709028 709031 709032 709033 709035 709037 709038 709039 709041 709042 709043 709045 709048 709049 709050 709053 709054 709057 709059 709063 709064 709066 709067 709069 709070 709072 709073 709075 709076 709077 709078 709079 709080 709082 709083 709085 709086 709087 709090 709091 709092 709095 709099 709101 709107 709109 709115 709116 709117 709119 709123 709125 709130 709140 709146 709150 709151 709157 709158 709161 709162 709166 709168 709169 709171 709172 709173 709176 709178 709180 709182 709183 709184 709186 709188 709193 709194 709196 709197 709198 709199 709202 709204 709205 709206 709207 709209 709210 709212 709213 709216 709218 709219 709220 709221 709226 709227 709230 709231 709232 709233 709241 709248 709249 709253 709255 709260 709261 709264 709265 709266 709267 709268 709270 709272 709274 709279 709281 709284 709287 709290 709292 709293 709298 709300 709306 709311 709316 709321 709325 709327 709328 709332 709335 709337 709340 709341 709349 709354 709360 709361 709362 709368 709369 709371 709372 709373 709380 709386 709387 709390 709393 709395 709397 709399 709400 709404 709416 709418 709425 709430 709433 709437 709443 709458 709459 709461 709464 709468 709474 709475 709490 709493 709494 709497 709498 709501 709502 709503 709510 709564 709565 709566 709567 709568 709572 709574 709578 709587 709589 709593 709609 709618 709635 709655 709656 709657 709658 709659 709662 709663 709665 709667 709668 709669 709670 709671 709673 709674 709684 709685 709687 709688 709693 709695 709714 709716 709717 709719 709720 709722 709723 709750 709758 709759 709760 709761 709762 709763 709764 709765 709766 709767 709768 709772 709774 709776 709781 709782 709784 709786 709789 709790 709791 709792 709794 709797 709803 709804 709805 709806 709807 709808 709809 709812 709813 709814 709815 709816 709818 709819 709820 709823 709824 709826 709827 709828 709829 709830 709831 709832 709833 709836 709837 709838 709844 709845 709846 709847 709848 709849 709850 709851 709852 709853 709854 709855 709856 709857 709859 709861 709862 709863 709864 709867 709868 709869 709870 709871 709872 709873 709876 709877 709878 709880 709881 709882 709883 709884 709885 709892 709893 709894 709895 709897 709898 709899 709900 709904 709905 709906 709907 709908 709909 709912 709913 709914 709916 709917 709918 709919 709920 709921 709922 709923 709924 709927 709929 709930 709931 709932 709933 709934 709935 709936 709937 709938 709939 709941 709942 709943 709944 709945 709946 709948 709949 709950 709951 709952 709953 709954 709955 709957 709958 709960 709961 709962 709963 709964 709965 709966 709967 709969 709970 709971 709972 709973 709974 709975 709977 709978 709979 709980 709982 709983 709984 709985 709986 709990 709992 709993 709995 709996 709998 710000 710001 710003 710005 710006 710008 710009 710010 710011 710014 710015 710016 710017 710018 710020 710021 710022 710023 710025 710027 710028 710030 710032 710034 710036 710037 710042 710043 710044 710046 710048 710049 710051 710052 710054 710055 710057 710058 710059 710061 710062 710063 710064 710066 710068 710071 710073 710074 710075 710076 710078 710079 710080 710081 710083 710089 710092 710093 710095 710096 710097 710098 710100 710101 710103 710104 710105 710106 710107 710109 710110 710111 710112 710113 710114 710115 710116 710117 710118 710119 710120 710123 710124 710126 710136 710137 710138 710139 710140 710141 710143 710144 710145 710146 710148 710149 710174 710175 710176 710177 710179 710180 710182 710194 710203 710204 710205 710209 710213 710219 710220 710221 710222 710224 710226 710227 710229 710231 710238 710244 710245 710247 710248 710249 710250 710252 710254 710258 710260 710262 710265 710267 710269 710272 710276 710278 710280 710282 710285 710291 710294 710296 710297 710298 710299 710301 710305 710307 710311 710312 710315 710319 710321 710322 710324 710327 710328 710334 710342 710344 710347 710348 710349 710352 710353 710362 710366 710367 710369 710370 710372 710375 710381 710386 710392 710415 710416 710422 710429 710433 710440 710467 710475 710483 710754 710766 710867 710901 710927 710931 710969 710981 710982 710983 710990 710998 711003 711006 711008 711017 711031 711032 711034 711036 711040 711041 711043 711045 711049 711055 711057 711061 711070 711071 711072 711078 711085 711088 711098 711104 711110 711114 711119 711130 711142 711143 711146 711156 711158 711162 711169 711173 711175 711176 711178 711181 711182 711184 711185 711186 711189 711194 711195 711198 711205 711210 711211 711212 711216 711223 711225 711229 711232 711236 711237 711239 711240 711242 711246 711252 711255 711258 711260 711261 711262 711263 711267 711268 711269 711270 711271 711274 711277 711279 711283 711289 711293 711294 711295 711296 711298 711302 711303 711306 711307 711308 711310 711311 711314 711318 711323 711332 711338 711340 711342 711344 711345 711347 711348 711349 711350 711352 711356 711359 711361 711364 711372 711373 711377 711382 711383 711385 711390 711396 711397 711398 711399 711407 711408 711417 711421 711425 711434 711435 711436 711437 711438 711445 711446 711448 711454 711456 711457 711459 711460 711461 711465 711467 711468 711470 711473 711474 711479 711483 711486 711488 711492 711493 711494 711496 711497 711500 711507 711508 711518 711519 711520 711522 711523 711526 711527 711528 711531 711532 711534 711535 711536 711539 711540 711541 711542 711543 711544 711545 711552 711553 711554 711559 711560 711562 711569 711570 711573 711575 711587 711591 711592 711596 711599 711600 711601 711602 711603 711606 711607 711608 711609 711611 711612 711614 711616 711618 711622 711624 711625 711626 711627 711628 711630 711632 711634 711635 711636 711637 711639 711640 711642 711645 711646 711649 711650 711651 711652 711654 711655 711656 711659 711662 711663 711669 711673 711678 711680 711685 711686 711687 711688 711689 711691 711692 711693 711694 711695 711696 711697 711698 711699 711700 711702 711703 711705 711706 711707 711710 711711 711714 711715 711726 711729 711732 711734 711737 711739 711740 711741 711742 711744 711745 711746 711747 711749 711750 711751 711755 711757 711760 711762 711766 711768 711771 711772 711778 711779 711783 711784 711791 711793 711794 711795 711796 711797 711798 711799 711801 711802 711803 711805 711806 711807 711808 711809 711810 711813 711815 711816 711818 711819 711820 711822 711823 711829 711834 711835 711838 711839 711840 711841 711842 711843 711848 711852 711853 711854 711855 711856 711857 711858 711859 711860 711861 711862 711863 711864 711866 711867 711868 711869 711870 711871 711872 711873 711874 711875 711876 711878 711880 711881 711883 711887 711888 711890 711891 711893 711895 711896 711899 711901 711903 711904 711906 711908 711910 711913 711915 711916 711917 711919 711920 711922 711926 711931 711933 711934 711935 711936 711937 711938 711939 711940 711948 711951 711952 711954 711956 711957 711958 711959 711960 711964 711965 711966 711967 711970 711971 711974 711975 711976 711977 711978 711979 711980 711981 711983 711988 711990 711999 712002 712010 712012 712014 712015 712017 712018 712025 712033 712039 712040 712044 712045 712061 712062 712063 712066 712067 712070 712072 712078 712082 712085 712086 712087 712089 712092 712093 712099 712118 712123 712124 712128 712129 712133 712135 712140 712156 712170 712177 712184 712186 712188 712192 712193 712194 712200 712204 712206 712207 712212 712213 712214 712216 712221 712223 712224 712225 712226 712228 712234 712238 712241 712242 712245 712247 712250 712251 712255 712263 712264 712266 712269 712270 712276 712285 712293 712295 712296 712304 712307 712308 712309 712313 712315 712316 712318 712320 712321 712322 712323 712326 712327 712328 712329 712331 712333 712334 712338 712340 712342 712345 712346 712347 712348 712349 712350 712355 712356 712358 712359 712363 712365 712367 712369 712370 712373 712374 712375 712377 712378 712379 712380 712381 712382 712385 712391 712393 712399 712402 712403 712407 712413 712417 712424 712431 712432 712434 712448 712450 712453 712454 712455 712457 712458 712460 712461 712465 712466 712467 712469 712470 712480 712514 712518 712519 712526 712527 712529 712531 712536 712539 712550 712557 712584 712591 712600 712606 712607 712610 712612 712616 712627 712631 712636 712640 712641 712642 712643 712644 712645 712646 712647 712648 712651 712653 712654 712656 712657 712659 712660 712666 712668 712669 712670 712671 712675 712681 712683 712723 712728 712729 712730 712731 712733 712736 712749 712758 712764 712769 712773 712781 712788 712810 712813 712814 712818 712819 712839 712842 712843 712845 712848 712849 712851 712852 712858 712859 712862 712868 712872 712874 712876 712881 712884 712885 712887 712888 712894 712897 712900 712902 712903 712904 712905 712908 712912 712916 712919 712921 712922 712926 712930 712940 712943 712945 712952 712954 713050 713051 713053 713063 713067 713071 713081 713088 713090 713091 713092 713094 713096 713097 713098 713100 713102 713103 713106 713118 713127 713136 713137 713138 713141 713142 713145 713151 713152 713155 713159 713160 713164 713169 713172 713173 713176 713177 713180 713184 713188 713189 713196 713197 713198 713202 713206 713208 713210 713215 713226 713232 713235 713238 713239 713240 713246 713247 713253 713255 713258 713263 713267 713268 713269 713274 713275 713277 713278 713279 713289 713290 713291 713292 713293 713295 713296 713297 713299 713301 713302 713303 713306 713307 713313 713315 713318 713319 713320 713326 713328 713329 713330 713331 713332 713338 713345 713348 713349 713350 713351 713355 713356 713360 713362 713363 713372 713373 713375 713377 713383 713387 713390 713391 713395 713396 713398 713399 713400 713403 713404 713405 713406 713407 713410 713411 713414 713416 713417 713418 713419 713420 713421 713423 713424 713425 713427 713429 713431 713433 713434 713435 713436 713437 713438 713443 713444 713445 713446 713447 713458 713477 713479 713484 713487 713488 713489 713493 713495 713500 713503 713547 713550 713554 713603 713604 713605 713622 713658 713666 713674 713675 713681 713692 713696 713701 713702 713703 713708 713714 713717 713721 713724 713728 713732 713734 713737 713739 713741 713742 713743 713744 713745 713747 713748 713749 713751 713752 713753 713754 713755 713756 713760 713765 713766 713767 713770 713772 713774 713775 713777 713778 713779 713781 713782 713783 713784 713785 713787 713788 713789 713790 713791 713792 713793 713794 713797 713798 713801 713803 713809 713813 713816 713817 713822 713825 713827 713828 713832 713834 713837 713838 713839 713840 713841 713842 713843 713848 713854 713855 713858 713860 713861 713862 713864 713867 713868 713869 713870 713872 713873 713874 713879 713880 713881 713882 713885 713886 713890 713891 713892 713893 713894 713895 713896 713897 713899 713900 713902 713904 713905 713906 713909 713910 713911 713912 713913 713914 713918 713925 713926 713928 713936 713940 713941 713942 713943 713945 713947 713951 713952 713957 713960 713961 713962 713963 713967 713972 713975 713980 713994 713995 713996 713997 714000 714001 714005 714008 714044 714047 714076 714083 714091 714092 714093 714094 714096 714097 714102 714106 714107 714124 714133 714134 714137 714138 714140 714141 714149 714150 714155 714161 714171 714174 714176 714177 714178 714179 714180 714186 714189 714190 714201 714202 714204 714206 714207 714213 714221 714224 714229 714231 714232 714236 714238 714240 714241 714243 714244 714246 714247 714251 714252 714253 714254 714256 714260 714261 714262 714263 714264 714265 714267 714268 714275 714277 714279 714281 714282 714283 714284 714288 714289 714290 714301 714305 714318 714320 714322 714328 714331 714332 714333 714334 714336 714337 714353 714375 714405 714418 714419 714444 714461 714467 714471 714472 714482 714485 714498 714501 714507 714509 714510 714566 714599 714627 714645 714651 714652 714656 714659 714661 714662 714663 714665 714666 714684 714691 714696 714700 714705 714706 714711 714715 714716 714719 714727 714738 714739 714744 714747 714749 714760 714762 714763 714764 714766 714771 714773 714775 714776 714778 714781 714783 714787 714790 714791 714792 714795 714796 714797 714798 714813 714815 714816 714830 714836 714879 714882 714885 714887 714888 715103 715104 715107 715110 715111 715112 715113 715114 715115 715117 715211 715212 715214 715215 715218 715224 715253 715289 715415 715424 715435 715440 715442 715461 715467 715470 715476 715477 715481 715482 715489 715496 715513 715518 715519 715528 715541 715543 715544 715545 715546 715547 715549 715560 715565 715568 715569 715573 715575 715577 715582 715583 715585 715588 715589 715593 715594 715598 715600 715605 715616 715619 715631 715635 715636 715638 715641 715642 715648 715649 715650 715651 715652 715653 715654 715656 715657 715660 715661 715663 715672 715673 715674 715675 715676 715677 715679 715680 715681 715682 715683 715684 715685 715686 715687 715688 715689 715690 715712 715713 715731 715732 715733 715734 715735 715746 715748 715749 715750 715753 715754 715755 715756 715757 715759 715760 715764 715781 715783 715785 715786 715787 715788 715789 715793 715794 715798 715801 715812 715813 715814 715816 715818 715819 715820 715821 715822 715824 715825 715826 715828 715831 715832 715833 715834 715835 715836 715837 715838 715839 715840 715843 715844 715848 715850 715852 715856 715857 715858 715859 715862 715863 715864 715865 715867 715870 715873 715877 715882 715883 715884 715885 715889 715890 715895 715897 715899 715905 715907 715916 715921 715927 715935 715937 715938 715940 715943 715972 715983 715984 715987 715991 715992 715993 715996 715997 716008 716010 716044 716049 716050 716073 716077 716078 716083 716084 716088 716089 716090 716094 716109 716114 716118 716119 716121 716122 716141 716145 716150 716151 716152 716153 716157 716158 716160 716161 716162 716163 716164 716165 716173 716175 716176 716182 716189 716190 716199 716211 716218 716222 716228 716229 716230 716231 716271 716272 716274 716301 716327 716402 716403 716404 716405 716406 716407 716408 716409 716410 716442 716456 716457 716458 716459 716472 716473 716476 716479 716480 716482 716490 716491 716496 716498 716501 716503 716506 716507 716508 716510 716511 716512 716514 716515 716516 716519 716574 716576 716634 716638 716639 716640 716642 716644 716650 716660 716664 716670 716692 716694 716753 716754 716755 716756 716757 716880 716948 717007 717014 717015 717020 717021 717025 717148 717266 717267 717272 717274 717275 717276 717283 717301 717309 717311 717316 717317 717334 717336 717338 717339 717369 717378 717386 717388 717415 717417 717419 717445 717446 717474 717475 717476 717477 717480 717483 717504 717505 717510 717520 717550 717553 717569 717579 717603 717624 717626 717646 717649 717650 717651 717652 717657 717659 717683 717722 717773 717774 717788 717790 717794 717796 717797 717806 717807 717809 717810 717824 717825 717827 717830 717842 717843 717847 717850 717853 717876 717894 717895 717900 717915 717941 717952 717958 717959 718019 718031 718037 718049 718069 718070 718074 718077 718102 718105 718107 718155 718157 718158 718159 718223 718224 718228 718233 718237 718243 718249 718250 718254 718260 718284 718295 718299 718302 718304 718307 718328 718332 718336 718337 718351 718353 718358 718361 718366 718371 718375 718383 718385 718386 718393 718403 718422 718423 718426 718428 718433 718436 718438 718451 718452 718453 718454 718455 718456 718457 718458 718459 718460 718461 718468 718469 718470 718471 718472 718473 718474 718475 718477 718478 718481 718484 718487 718489 718490 718491 718492 718494 718496 718497 718499 718503 718504 718505 718506 718508 718509 718510 718511 718513 718514 718515 718517 718518 718519 718520 718521 718522 718523 718524 718525 718527 718528 718529 718532 718534 718539 718540 718542 718544 718547 718549 718550 718551 718554 718557 718558 718561 718562 718566 718567 718569 718570 718572 718576 718577 718587 718593 718599 718608 718612 718622 718624 718625 718626 718628 718629 718630 718631 718633 718634 718635 718638 718641 718645 718650 718651 718653 718654 718655 718656 718657 718659 718660 718664 718669 718671 718672 718674 718675 718678 718681 718682 718684 718685 718686 718687 718688 718690 718691 718695 718696 718697 718699 718700 718702 718704 718707 718708 718709 718711 718712 718714 718717 718721 718722 718723 718724 718727 718730 718731 718732 718733 718734 718736 718738 718739 718741 718742 718743 718744 718745 718746 718747 718748 718749 718751 718752 718753 718754 718755 718756 718757 718758 718759 718760 718762 718764 718766 718768 718769 718770 718771 718772 718773 718774 718775 718776 718777 718778 718780 718781 718783 718787 718792 718793 718794 718797 718798 718799 718801 718803 718805 718806 718807 718808 718809 718812 718813 718814 718815 718817 718818 718819 718820 718823 718825 718831 718832 718833 718834 718835 718836 718838 718841 718842 718843 718844 718846 718847 718850 718851 718852 718853 718854 718855 718856 718858 718859 718860 718861 718870 718877 718881 718882 718884 718885 718886 718887 718888 718889 718890 718892 718894 718895 718896 718898 718899 718901 718902 718903 718905 718906 718908 718915 718916 718924 718951 718952 718953 718954 718955 718958 718959 718960 718962 718968 718972 718973 718976 718985 718993 718999 719003 719004 719007 719010 719013 719016 719017 719022 719024 719025 719029 719031 719035 719039 719040 719044 719047 719048 719049 719050 719071 719078 719080 719081 719083 719086 719088 719091 719106 719108 719109 719112 719113 719124 719125 719126 719137 719145 719148 719149 719150 719151 719152 719153 719154 719179 719180 719189 719494 719648 719661 719663 719664 719666 719667 719668 719670 719731 719734 719735 719744 719753 719754 719755 719757 719764 719767 719769 719770 719771 719773 719774 719776 719779 719781 719782 719791 719792 719793 719794 719797 719798 719799 719801 719802 719808 719810 719813 719814 719819 719820 719821 719823 719824 719826 719827 719841 719845 719858 719861 719862 719863 719865 719881 719885 719894 719901 719904 719906 719910 719939 719942 719963 719971 719972 719973 719975 719977 719979 719980 719983 719984 719986 719988 720009 720016 720021 720022 720023 720025 720026 720031 720032 720041 720042 720044 720045 720047 720048 720049 720050 720064 720066 720067 720068 720071 720072 720073 720074 720075 720076 720084 720085 720093 720094 720095 720101 720105 720106 720107 720111 720114 720116 720117 720119 720120 720121 720123 720126 720133 720138 720140 720148 720152 720161 720172 720175 720178 720180 720182 720188 720189 720190 720201 720211 720234 720237 720297 720300 720304 720306 720307 720308 720310 720314 720317 720321 720323 720325 720330 720331 720333 720334 720336 720337 720338 720339 720340 720341 720342 720343 720344 720345 720346 720348 720351 720355 720358 720360 720361 720364 720367 720370 720371 720372 720375 720376 720378 720379 720381 720394 720399 720401 720405 720430 720436 720439 720441 720442 720443 720450 720463 720469 720472 720474 720477 720485 720497 720500 720504 720511 720514 720517 720523 720524 720551 720552 720555 720573 720574 720597 720600 720605 720607 720608 720610 720613 720614 720624 720625 720647 720656 720659 720662 720663 720668 720674 720676 720677 720682 720683 720685 720687 720726 720728 720729 720733 720741 720750 720751 720752 720758 720762 720763 720764 720765 720767 720772 720773 720784 720786 720788 720789 720797 720798 720802 720811 720823 720829 720835 720836 720855 720856 720863 720865 720871 720874 720951 720952 720953 720957 720980 721018 721033 721034 721035 721051 721061 721120 721124 721133 721146 721247 721249 721355 721501 721528 721529 721549 721552 721569 721570 721576 721578 721580 721581 721584 721586 721597 721602 721603 721604 721611 721628 721634 721639 721640 721649 721651 721652 721656 721657 721658 721660 721663 721665 721678 721681 721684 721686 721687 721688 721689 721690 721691 721692 721695 721697 721698 721700 721701 721702 721705 721706 721707 721708 721714 721727 721729 721730 721731 721732 721733 721734 721737 721739 721744 721748 721749 721751 721752 721753 721754 721755 721757 721758 721759 721762 721764 721765 721768 721775 721776 721777 721778 721779 721780 721781 721782 721783 721784 721785 721790 721792 721796 721800 721802 721804 721807 721809 721810 721811 721812 721813 721815 721816 721818 721819 721820 721824 721825 721827 721828 721829 721837 721838 721839 721842 721843 721844 721848 721850 721851 721852 721853 721856 721858 721859 721865 721868 721873 721875 721881 721882 721884 721898 721899 721900 721905 721907 721914 721928 721931 721933 721937 721944 721946 721948 721950 721957 721970 721971 721976 721979 721980 721982 721983 721984 721985 721986 721988 721989 721990 721991 721992 721994 721995 722038 722039 722043 722048 722049 722050 722051 722052 722053 722054 722055 722056 722057 722058 722059 722062 722063 722065 722066 722067 722068 722069 722070 722074 722075 722076 722079 722080 722083 722085 722086 722087 722088 722089 722092 722093 722094 722097 722102 722103 722104 722105 722106 722107 722108 722109 722110 722111 722113 722114 722116 722119 722120 722121 722122 722123 722124 722125 722136 722137 722138 722139 722141 722142 722144 722145 722146 722150 722151 722152 722153 722154 722155 722156 722162 722164 722167 722169 722188 722189 722190 722195 722200 722202 722224 722228 722230 722236 722241 722251 722252 722254 722323 722336 722338 722342 722346 722351 722354 722355 722362 722363 722365 722366 722367 722368 722370 722371 722373 722375 722376 722378 722383 722384 722387 722392 722395 722399 722410 722411 722412 722414 722415 722422 722426 722428 722431 722438 722442 722443 722447 722449 722450 722451 722453 722459 722461 722462 722463 722464 722481 722485 722486 722487 722489 722491 722494 722495 722497 722504 722506 722515 722529 722543 722544 722560 722563 722564 722566 722567 722595 722605 722610 722611 722634 722636 722638 722640 722641 722642 722662 722663 722664 722665 722666 722668 722669 722671 722673 722686 722687 722690 722693 722697 722702 722717 722725 722728 722731 722733 722747 722769 722774 722776 722782 722786 722791 722796 722801 722803 722808 722811 722816 722820 722828 722841 722843 722847 722848 722850 722851 722854 722864 722869 722882 722885 722894 722896 722901 722907 722940 722953 722961 722962 722970 722971 722973 722974 722993 723001 723008 723014 723015 723016 723019 723026 723034 723035 723043 723047 723055 723057 723061 723069 723071 723072 723078 723091 723095 723098 723104 723105 723106 723109 723113 723116 723121 723133 723145 723147 723151 723152 723159 723167 723177 723181 723190 723191 723193 723194 723201 723203 723205 723283 723289 723332 723398 723401 723416 723417 723421 723423 723424 723425 723426 723428 723437 723440 723481 723482 723511 723513 723514 723516 723526 723545 723548 723584 723600 723601 723629 723630 723632 723663 723689 723693 723699 723706 723719 723720 723724 723725 723753 723781 723812 723835 723836 723840 723841 723843 723845 723852 723868 723869 723883 723892 723963 723989 723993 724001 724005 724007 724008 724010 724012 724013 724199 724205 724268 724404 724407 725272 725321 725327 725443 725469 725477 725495 725589 725590 725591 725592 725993 725997 726002 726008 726012 726013 726016 726020 726021 726022 726024 726027 726028 726034 726046 726059 726073 726074 726084 726092 726097 726126 726127 726128 726130 726132 726133 726138 726141 726169 726181 726184 726193 726195 726219 726223 726225 726232 726235 726240 726279 726285 726286 726298 726300 726301 726302 726303 726304 726306 726307 726309 726311 726325 726331 726333 726335 726337 726355 726358 726361 726388 726571 726572 726578 726580 726583 726584 726585 726588 726627 726628 726630 726664 726669 726670 726695 726703 726706 726721 726727 726728 726731 726734 726735 726749 726758 726759 726768 726770 726789 726792 726793 726794 726797 726816 726896 726899 726902 726940 726942 726949 726951 726953 726960 726961 726966 726967 726988 727021 727022 727025 727026 727027 727033 727034 727035 727040 727064 727066 727067 727079 727084 727085 727087 727097 727099 727105 727135 727136 727168 727170 727171 727174 727183 727185 727194 727202 727205 727211 727215 727216 727217 727218 727222 727228 727234 727237 727238 727246 727262 727272 727275 727276 727278 727279 727283 727287 727292 727293 727294 727295 727296 727297 727298 727301 727304 727305 727306 727316 727317 727320 727322 727329 727330 727331 727332 727333 727336 727340 727341 727343 727346 727353 727354 727380 727384 727386 727388 727389 727391 727394 727396 727397 727399 727403 727433 727443 727444 727450 727451 727454 727455 727456 727457 727460 727462 727465 727466 727468 727469 727471 727472 727478 727479 727480 727482 727490 727498 727526 727564 727565 727566 727567 727570 727574 727576 727577 727579 727580 727581 727590 727592 727593 727618 727619 727622 727623 727626 727627 727648 727649 727651 727656 727856 727857 727864 727866 727869 727974 727979 727980 727981
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VISIONIST™ CRT-P U226Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls18 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under P030005 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VALITUDE™ X4 CRT-P U128Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls26 deaths2,073 reports
- VISIONIST™ X4 CRT-P U228Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall4 deaths819 reports
- INVIVE™ V173Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death670 reports
- VALITUDE™ CRT-P U125Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls2 deaths558 reports
- VISIONIST™ CRT-P U225Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall2 deaths253 reports
- INTUA™ V273Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)264 reports
- INTUA™ V272Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)23 reports
- INVIVE™ V172Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95969).
- Z-0822-2025 · Open, ClassifiedACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)
- Z-0823-2025 · Open, ClassifiedPROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221), and EL MRI (Model Number L231).
- Z-0824-2025 · Open, ClassifiedESSENTIO DR Pacemaker, SL (Model Number L101, SL MRI (Model Number L111), EL (Model Number L121), and EL MRI (Model Number L131)
- Z-0826-2025 · Open, ClassifiedVALITUDE CRT-P, EL (Model Number U125) and EL MRI (Model Number U128)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page