VISIONIST™ CRT-P U225Recalls & reported problems
Boston Scientific Corporation · Cardiac Resynchronization Therapy Pacemaker · Class 3 · Implant
All models of VISIONIST CRT-P (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Boston Scientific VISIONIST CRT-P is an implanted heart pacemaker whose 241 reports describe early battery drain, sensing and pacing problems, and infections that led to system replacement. The reports include 2 deaths and 120 injuries, and they rose from 24 in the prior 12 months to 58 in the last 12 months. One recall, open since August 2025, covers software meant to prevent a safety mode from starting when battery resistance is high.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 253
- Deaths reported
- 2
- Injuries reported
- 121
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 253 adverse event reports that list this exact device (DI 00802526559426), received between Oct 25, 2018 and Sep 25, 2026 — classified as 2 deaths, 121 injuries, 130 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (69 reports).
1 FDA recall record name this device's identifier, 1 still open.
19 other recalls cover devices cleared under the same FDA submission (P030005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Aug 20, 2025Open, Classified
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Z-0103-2026 · Root cause: Other · Recall details · FDA record ↗
Most reported problems
Device problems
- Premature Discharge of Battery69
- Adverse Event Without Identified Device or Use Problem46
- High impedance39
- Over-Sensing34
- Signal Artifact/Noise27
- Pacing Problem26
- Incorrect, Inadequate or Imprecise Result or Readings22
- Battery Problem21
Patient problems
- No Clinical Signs, Symptoms or Conditions150
- Unspecified Infection39
- No Consequences Or Impact To Patient15
- No Known Impact Or Consequence To Patient9
- Pocket Erosion7
- Sepsis7
- Syncope/Fainting7
- Asystole6
Recent adverse event reports
The 25 most recent of 253 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 25, 2026Failure to Capture; Pacing Problem
t was reported that this cardiac resynchronization therapy pacemaker (CRT-P) and the left ventricular (LV) lead suspended a left ventricular automatic threshold (LVAT) test. Technical Services (TS) reviewed device data and confirmed the LVAT was suspended due to fusion beats. Also, the presenting electrogram (EGM) exhibited loss of capture on the LV lead due to the LV and right ventricular (RV) beats falling on top of each other. TS recommended further evaluation in clinic. This CRT-P remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50899270 ↗InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) was part of a system revision due to infection. There were no additional adverse patient effects reported. The CRT-P was explanted.
Patient problems: Unspecified Infection
MAUDE report 50896912 ↗InjurySep 15, 2026Pacing Problem
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded pacemaker mediated tachycardia (PMT) episode stored in the configured collection period. The Boston Scientific Technical Services consultant discussed that the PMT episodes appeared atrial or sinus driven and presenting electrogram showed that the device was operating as programmed. As of this time, this CRT-P remains in service. No adverse patient effects were reported.Additional information indicated that this CRT-P was explanted and replaced with a different model. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50772778 ↗MalfunctionSep 14, 2026Pacing Problem; Under-Sensing
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) exhibited undersensing and a decrease in biventricular pacing. Technical Services (TS) was consulted and confirmed the undersensing was through the non-Boston Scientific atrial lead. TS informed the health care professional that this allows intrinsic beats to occur and prevents the biventricular pacing. TS reviewed measurements and recommended programming atrial sensitivity to a nominal value. No adverse patient effects were reported. This CRT-P remains in service.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50754468 ↗MalfunctionSep 11, 2026Premature Discharge of Battery
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a Code 1003, indicating the voltage was too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. The device remains in service. No adverse patient effects were reported.Additional information was received indication that the device firmware was successfully updated during a clinic follow-up, and the patient will continue to be monitored by the clinic. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50725356 ↗MalfunctionSep 7, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50656112 ↗InjurySep 7, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50652581 ↗MalfunctionSep 4, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602436 ↗MalfunctionSep 4, 2026Premature Discharge of Battery
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. This device remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602341 ↗MalfunctionSep 4, 2026Signal Artifact/Noise; Over-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602158 ↗InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient with this cardiac resynchronization therapy pacemaker (CRT-P) experienced impaired healing of pocket site. Reportedly, the site was unable to heal since implant. According to the physician, the patient agrees that the pocket site never really healed. Ultimately, the patient's dog jumped on the chest and ripped the site open. There were no additional adverse patient effects reported. The CRT-P was explanted.
Patient problems: Impaired Healing; Pocket Erosion
MAUDE report 50601112 ↗MalfunctionSep 4, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50600881 ↗MalfunctionAug 7, 2026Premature Discharge of Battery
It was reported that this device recorded a Code 1003, indicating the voltage is too low for the projected remaining capacity. A request was made to have data from this device analyzed. Data analysis identified potential high impedance within the battery. Continued monitoring was recommended. The device still remains in service. No adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50181869 ↗InjuryAug 7, 2026Premature Discharge of Battery
It was reported that this cardiac resynchronization therapy pacemaker (CRT-P) recorded a fault code 1003 indicative of a battery voltage too low for the projected remaining capacity. Subsequently, this CRT-P was explanted and replaced. This device was received for investigation. No additional adverse patient effects were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50168830 ↗MalfunctionAug 6, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50165157 ↗MalfunctionAug 6, 2026Over-Sensing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50162193 ↗MalfunctionAug 3, 2026Premature Discharge of Battery
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50119795 ↗MalfunctionJul 30, 2026Signal Artifact/Noise; Premature Discharge of Battery; Use of Device Problem; Data Problem; Interrogation Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50091214 ↗MalfunctionJul 21, 2026Premature Discharge of Battery; Battery Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25892721 ↗MalfunctionJul 9, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) RECORDED A CODE 1003, INDICATING THE VOLTAGE IS TOO LOW FOR THE PROJECTED REMAINING CAPACITY. A REQUEST WAS MADE TO HAVE DATA FROM THIS DEVICE ANALYZED. DATA ANALYSIS IDENTIFIED POTENTIAL HIGH IMPEDANCE WITHIN THE BATTERY. CONTINUED MONITORING WAS RECOMMENDED. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25767040 ↗MalfunctionJun 29, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) RECORDED A CODE 1003, INDICATING THE VOLTAGE IS TOO LOW FOR THE PROJECTED REMAINING CAPACITY. A REQUEST WAS MADE TO HAVE DATA FROM THIS DEVICE ANALYZED. DATA ANALYSIS IDENTIFIED POTENTIAL HIGH IMPEDANCE WITHIN THE BATTERY. CONTINUED MONITORING WAS RECOMMENDED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THIS DEVICE REMAINS IN SERVICE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25659513 ↗InjuryJun 26, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) WAS PART OF A SYSTEM REVISION DUE TO INFECTION. THERE WAS NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED. THIS CRT-P WAS EXPLANTED.
Patient problems: Unspecified Infection
MAUDE report 25638457 ↗MalfunctionJun 17, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) RECORDED A CODE 1003, INDICATING THE VOLTAGE IS TOO LOW FOR THE PROJECTED REMAINING CAPACITY. THIS DEVICE REMAINS IN USE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE PATIENT WILL CONTINUED TO BE MONITORED. THE DEVICE REMAINS IN USE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25538441 ↗MalfunctionJun 16, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) RECORDED A CODE 1003, INDICATING THE VOLTAGE IS TOO LOW FOR THE PROJECTED REMAINING CAPACITY. A REQUEST WAS MADE TO HAVE DATA FROM THIS DEVICE ANALYZED. DATA ANALYSIS IDENTIFIED POTENTIAL HIGH IMPEDANCE WITHIN THE BATTERY. CONTINUED MONITORING WAS RECOMMENDED. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25523841 ↗MalfunctionJun 10, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THIS CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) RECORDED A CODE 1003, INDICATING THE VOLTAGE IS TOO LOW FOR THE PROJECTED REMAINING CAPACITY. A REQUEST WAS MADE TO HAVE DATA FROM THIS DEVICE ANALYZED. DATA ANALYSIS IDENTIFIED POTENTIAL HIGH IMPEDANCE WITHIN THE BATTERY. CONTINUED MONITORING WAS RECOMMENDED. THIS DEVICE REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25463960 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P030005. One submission often covers many models, so check the product and lot details in each record.
- Z-1771-2026 · Mar 19, 2026 · Open, ClassifiedBoston Scientific Cardiac Resynchronization Therapy (CRT)labeled as: 1. VALITUDE CRT-P EL, Model Number U125; 2. VALITUDE CRT-P EL MRI, Model Number U128; 3. VISIONIST CRT-P EL
- Z-0101-2026 · Aug 20, 2025 · Open, ClassifiedModel Number U125, VALITUDE CRT-P EL
- Z-0104-2026 · Aug 20, 2025 · Open, ClassifiedModel Number U226, VISIONIST CRT-P EL
- Z-0105-2026 · Aug 20, 2025 · Open, ClassifiedModel Number U228, VISIONIST CRT-P EL MRI
- Z-0102-2026 · Aug 20, 2025 · Open, ClassifiedModel Number U128, VALITUDE CRT-P EL MRI
- Z-0826-2025 · Dec 12, 2024 · Open, ClassifiedVALITUDE CRT-P, EL (Model Number U125) and EL MRI (Model Number U128)
- Z-0827-2025 · Dec 12, 2024 · Open, ClassifiedVISIONIST CRT-P, EL (Model Number U225 and U226) and EL MRI (Model Number U228)
- Z-2002-2021 · Jun 3, 2021 · Open, ClassifiedVALITUDE CRT-P pacemaker, Models: a) U125 b) U128
- Z-2018-2021 · Jun 3, 2021 · Open, ClassifiedINVIVE CRT-P, Model Numbers: a) V172 b) V173 c) V182 d) V183 e) W172 f) W173
- Z-2019-2021 · Jun 3, 2021 · Open, ClassifiedINTUA CRT-P, Model Numbers: a) V272 b) V273 c) W273
- Z-2020-2021 · Jun 3, 2021 · Open, ClassifiedINLIVEN CRT-P, Model Numbers: a) V284 b) V285 c) W274 d) W275 (added 7/15/2021)
- Z-2003-2021 · Jun 3, 2021 · Open, ClassifiedVISIONIST CRT-P pacemaker, Models: a) U225 b) U226 c) U228
- Z-3258-2018 · Sep 10, 2018 · TerminatedVALITUDE X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)
- Z-3259-2018 · Sep 10, 2018 · TerminatedVISIONIST X4 CRT-P (Cardiac Resynchronization Therapy Pacemaker)
- Z-0360-2018 · Dec 7, 2017 · Open, ClassifiedVALITUDE CRT-P Pacemaker
…and 4 more.
Other Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P) devices
- VALITUDE™ X4 CRT-P U128Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls26 deaths2,073 reports
- Quadra Allure MP™ PM3562St. Jude Medical, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)26 deaths1,180 reports
- Percepta™ Quad CRT-P MRI SureScan™ W4TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall28 deaths897 reports
- VISIONIST™ X4 CRT-P U228Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall4 deaths819 reports
- Percepta™ Quad CRT-P MRI SureScan™ W4TR01Medtronic, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 recall8 deaths703 reports
- INVIVE™ V173Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)1 death670 reports
- VALITUDE™ CRT-P U125Boston Scientific Corporation · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)2 recalls2 deaths558 reports
- QUADRA ALLURE MP™ PM3262St. Jude Medical, Inc. · Pulse generator, pacemaker, implantable, with cardiac resynchronization (CRT-P)27 deaths518 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page