Skip to content

VITEK2Recalls & reported problems across all models

Biomerieux , Inc. · 5 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one VITEK2 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls
2

We found 5 models of VITEK2 by Biomerieux , Inc. in the FDA's device registry (GUDID). Together they are named in 2 adverse event reports in the FDA's MAUDE database, received between Jan 3, 2019 and Oct 3, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

2 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

VITEK2 recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jul 13, 2022Open, Classified

    There are 7 reported software anomalies that may affect use of the device.

    Z-1640-2022 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  2. Jan 28, 2021Terminated

    bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

    Z-1307-2021 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported VITEK2 problems

Device problems

  • Communication or Transmission Problem2

Patient problems

  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

VITEK2 models

5 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent VITEK2 adverse event reports

The 2 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Biomerieux , Inc. products

All Biomerieux , Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page