Skip to content

VITEK 2 COMPACTRecalls & reported problems across all models

Biomerieux , Inc. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
25
Deaths reported
0
Injuries reported
0
Recalls
1

We found 3 models of VITEK 2 COMPACT by Biomerieux , Inc. in the FDA's device registry (GUDID). Together they are named in 25 adverse event reports in the FDA's MAUDE database, received between Sep 17, 2019 and Aug 14, 2026 — classified as 25 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (15 reports).

1 FDA recall record name at least one of these models by its device identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

VITEK 2 COMPACT recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jan 28, 2021Terminated

    bioMrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

    Z-1308-2021 · Root cause: Software design · names 3 of these models · Recall details · FDA record ↗

Most reported VITEK 2 COMPACT problems

Device problems

  • Computer Software Problem15
  • Application Program Problem2
  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Communication or Transmission Problem1
  • Data Problem1
  • Decreased Sensitivity1
  • Failure to Back-Up1
  • High Test Results1

Patient problems

  • No Known Impact Or Consequence To Patient18
  • No Clinical Signs, Symptoms or Conditions7

VITEK 2 COMPACT models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent VITEK 2 COMPACT adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Biomerieux , Inc. products

All Biomerieux , Inc. devices →