VivAer StylusRecalls & reported problems across all models
Aerin Medical Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one VivAer Stylus model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 1
We found 2 models of VivAer Stylus by Aerin Medical Inc. in the FDA's device registry (GUDID). Together they are named in 3 adverse event reports in the FDA's MAUDE database, received between Aug 26, 2022 and Mar 5, 2025 — classified as 3 injuries.
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
VivAer Stylus recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Nov 12, 2024Open, Classified
Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.
Z-0808-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported VivAer Stylus problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
Patient problems
- Cartilage damage1
- Deformity/ Disfigurement1
- Epistaxis1
- Perforation1
VivAer Stylus models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- VivAer® Stylus FG1787Electrosurgical, Cutting & Coagulation & Accessories1 recall0 reports
- VivAer® Stylus FG722Electrosurgical, Cutting & Coagulation & Accessories3 reports
Recent VivAer Stylus adverse event reports
The 3 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryMar 5, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT PRESENTED WITH EXTERNAL NASAL VALVE COLLAPSE AND LOWER LATERAL CARTILAGE STENOSIS. THE PHYSICIAN TREATED THE LOWER LATERAL CARTILAGE AND ALAR CARTILAGE WITH THE VIVAER STYLUS AND REPORTED OVERLAPPING SOME TREATMENTS. THE PATIENT RETURNED FOR FOLLOW UP AFTER A PERIOD OF SEVERAL MONTHS WITH ABSORPTION OF A PORTION OF THE LOWER LATERAL CARTILAGE AND WORSENING OF THE EXTERNAL NASAL VALVE COLLAPSE. APPROXIMATELY 14 MONTHS LATER, A REVISION SURGERY WAS PERFORMED, OPEN VALVE NASAL RECONSTRUCTION. IT WAS ALSO REPORTED THAT THE PATIENT HAD TWO PREVIOUS NASAL SURGERIES INCLUDING A RHINOPLASTY. BASED ON THESE CIRCUMSTANCES, INTENTIONAL USE ERROR CAUSED AND CONTRIBUTED TO THE REPORTED EVENT. ADDITIONAL INFORMATION FROM THE PHYSICIAN STATES THEY ARE AWARE OF INDICATIONS FOR USE. AERIN MEDICAL WILL CONTINUE TO MONITOR AND TREND FOR REPORTS OF THIS NATURE.
Patient problems: Cartilage damage; Deformity/ Disfigurement
MAUDE report 21532491 ↗InjuryJan 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT RECEIVED VIVAER TREATMENT OF THE INFERIOR TURBINATES AND SWELL BODIES ON (B)(6) 2023. APPROXIMATELY 1 MONTH POST TREATMENT ((B)(6) 2023) THE PATIENT EXPERIENCED A SEVERE NOSEBLEED AND WENT TO THE EMERGENCY ROOM AND HAD EMBOLISM OF THE SPA DONE TO STOP THE BLEEDING AND SPENT 5 DAYS IN THE HOSPITAL. THE PATIENT REPORTED THAT THEY LOST NEARLY 2 PINTS OF BLOOD AND THAT SHE WAS RECOVERING BUT IN PAIN AND THEIR THROAT WAS IRRITATED FROM THE INTUBATION TUBE. APPROXIMATELY 1 WEEK LATER, THE PATIENT EXPERIENCED A SECOND NOSEBLEED AND RETURNED TO THE EMERGENCY ROOM. A CEREBRAL ANGIOGRAM WAS PERFORMED, AND THE PATIENT HAD EMBOLIZATION OF THE SPA REPEATED TO STOP THE BLEEDING. THE PATIENT IS RECOVERING.
Patient problems: Epistaxis
MAUDE report 18490156 ↗InjuryAug 26, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2022, THE PATIENT WAS TREATED WITH THE VIVAER STYLUS ONLY ON THE RIGHT SIDE. THE SWELL BODY, INTERNAL NASAL VALVE AND THE INFERIOR TURBINATE WERE TREATED. PATIENT HAD FIRST FOLLOW UP VISIT ON (B)(6) 2022 AND AGAIN ON (B)(6) 2022. AT BOTH VISITS, THE PATIENT SHOWED GOOD MUCOSAL HEALING. AT THIRD FOLLOW-UP VISIT ON (B)(6) 2022, MUSCOSA WAS INTACT. PHYSICIAN DISCUSSED SEPTO-TURB OPTION WITH THE PATIENT BECAUSE PATIENT WAS STILL COMPLAINING OF CONGESTION. AT 7-MONTH FOLLOW-UP ON (B)(6) 2022, PATIENT PRESENTED WITH CRUSTING ON SEPTUM AND 1X1 CM2 MUCOSAL PERFORATION IN AREA TREATED WITH VIVAER STYLUS. CARTILAGE WAS INTACT. THE PATIENT REPORTED USING AFRIN OFF AND ON FOR PAST MONTH. PHYSICIAN PLANS TO MOVE FORWARD WITH SEPTOPLASTY PROCEDURE TO ADDRESS PERSISTENT CONGESTION. TOPICAL NASAL DECONGESTANT SPRAYS HAVE BEEN REPORTED TO CAUSE MUCOSAL PERFORATION. THE DETERIORATION OF THE MUCOSA IS PROBABLY RELATED TO THE COMBINATION OF THE RADIOFREQUENCY TREATMENT OF THE TISSUE IN THE AREA FOLLOWED BY USE OF AFRIN.
Patient problems: Perforation
MAUDE report 15307612 ↗
Other Aerin Medical Inc. products
All Aerin Medical Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page