VivAer Stylus recallZ-0808-2025 · Aerin Medical, Inc. · initiated Nov 12, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Aerin Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some units in a single production lot were programmed incorrectly which will result in the VivAer Stylus being incorrectly recognized as RhinAer Stylus when connected to the Aerin Console.
FDA root cause category: Under Investigation by firm
What the company is doing
The recalling firm issued emails on 11/12/2024. It was necessary to notify several consignees by mail with an "URGENT: MEDICAL DEVICE FIELD NOTICE" letter dated 11/20/2024. The email described the product purchased by the consignee and the reason for recall. The consignee is informed the programmed treatment time will be shorter than the recommended 30 second treatment time for VivAer and the use of the device may result in insufficient treatment. Under the heading "Action to be Taken," it says if a defective device is identified or the consignee wishes to return unused devices within this production lot, they are to return them to the recalling firm, and return instructions are provided. Regarding the mailed "URGENT: MEDICAL DEVICE FIELD NOTICE" letters dated 11/20/2024, they informed the consignee that some of the units in the production lot were programmed incorrectly and, in these cases, the programmed treatment time will be 12 seconds of radiofrequency energy delivery with no cooling time, not the 30 second treatment time for VivAer that includes 18 seconds of energy delivery and 12 seconds of cooling time. The letter also contains instructions to share this notification with all device users within their facility and network to ensure they are aware of this issue. This notification should also be shared with any organization where the potentially affected devices have been transferred. If you need any further information or support concerning this notification, please contact us at [email protected] or call 512-221-9956.
Recall details
- Recall number
- Z-0808-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 12, 2024
- Posted by the FDA
- Jan 6, 2025
- Recalling firm
- Aerin Medical, Inc.
- Quantity
- 479 units
- Product code
- GEI
- FDA submissions
- K200300
- Distribution
- US Nationwide Distribution: AL, AZ, CA, CO, GA, ID, IL, IN, KY LA, MI, MN, NY, NE, NY, OH, OR PA, SC, TN, TX, UT, VA, WA, WI, and WV.
Product description
Aerin Medical VivAer Stylus, REF CAT1785 (individual stylus), and REF CAT500 (5-pack), sterile.
Code information (models, lots, serial numbers)
Lot #10235, exp. 2026-05-01; UDI (01)08886479300269(17)260501(10)10235
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VivAer® Stylus FG1787Aerin Medical Inc. · Electrosurgical, Cutting & Coagulation & Accessories1 recall0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K200300 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- VivAer® Stylus FG722Aerin Medical Inc. · Electrosurgical, Cutting & Coagulation & Accessories3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page