WATCHDOGRecalls & reported problems across all models
Boston Scientific Corporation · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one WATCHDOG model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 2
We found 2 models of WATCHDOG by Boston Scientific Corporation in the FDA's device registry (GUDID). Together they are named in 5 adverse event reports in the FDA's MAUDE database, received between Jul 14, 2022 and Jun 4, 2025 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Defective Device” (2 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
WATCHDOG recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Apr 5, 2023Open, Classified
The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.
Z-1593-2023 · Root cause: Process control · Recall details · FDA record ↗
- Apr 5, 2023Open, Classified
The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.
Z-1592-2023 · Root cause: Process control · Recall details · FDA record ↗
Most reported WATCHDOG problems
Device problems
- Defective Device2
- Adverse Event Without Identified Device or Use Problem1
- Air/Gas in Device1
- Device-Device Incompatibility1
- Fluid/Blood Leak1
- Loose or Intermittent Connection1
Patient problems
- Air Embolism2
- No Clinical Signs, Symptoms or Conditions2
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
- Non specific EKG/ECG Changes1
- Ventricular Fibrillation1
WATCHDOG models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- WATCHDOG™ H74939343A020Wire, guide, catheter2 recalls2 reports
- WATCHDOG™ H74939343020Adaptor, stopcock, manifold, fitting, cardiopulmonary bypass3 reports
Recent WATCHDOG adverse event reports
The 5 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryJun 4, 2025Defective Device; Defective Device
IT WAS REPORTED THAT VALVE ISSUES OCCURRED CAUSING AIR EMBOLISM AND PATIENT INSTABILITY. A WATCHDOG HEMOSTASIS VALVE WAS SELECTED FOR USE. DURING THE PROCEDURE, VALVE ISSUES WERE ENCOUNTERED, AIR ENTERED THE SYSTEM AND THE VESSEL, WHICH CREATED SOME INITIAL PATIENT INSTABILITY. IT WAS NOTED THAT FLUSHING AND WAITING RESOLVED THE ISSUE. THE REMAINDER OF THE PROCEDURE WAS CARRIED OUT SUCCESSFULLY USING AN ALTERNATE DEVICE AND THE PATIENT WAS FINE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Air Embolism; Insufficient Information
MAUDE report 22146753 ↗InjuryNov 27, 2024Defective Device; Adverse Event Without Identified Device or Use Problem; Defective Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AIR EMBOLISM OCCURRED REQUIRING INTERVENTION. A WATCHDOG HEMOSTASIS VALVE WAS SELECTED FOR A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE. DURING THE PROCEDURE, AN AIR EMBOLISM WAS INTRODUCED VIA THE GUIDE CATHETER. IT WAS NOTED THAT THE CAUSE IS RELATED TO THE WATCHDOG DEVICE. CONSEQUENTLY, THE PATIENT HAD ST ELEVATION AND MAY HAVE GOTTEN A LITTLE SICK BUT WAS ULTIMATELY FINE. CODE STRATEGIES WERE EXECUTED, AND THE PATIENT WAS SUCCESSFULLY REVIVED FROM COMPLICATIONS. THE PATIENT WAS ADMITTED TO THE HOSPITAL BEYOND STANDARD OF CARE. THE PATIENT FULLY RECOVERED, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING ORIGINAL METHOD AND/OR DEVICE. IT WAS FURTHER REPORTED THAT THE PATIENT SUSTAINED AN AIR EMBOLISM THAT RESULTED IN VENTRICULAR FIBRILLATION. THE PATIENT WAS RESUSCITATED SUCCESSFULLY, ADMITTED TO THE HOSPITAL FOR FOLLOW UP CARE AND DISCHARGED FROM THE HOSPITAL IN STABLE CONDITION.
Patient problems: Air Embolism; Non specific EKG/ECG Changes; Ventricular Fibrillation
MAUDE report 20800656 ↗MalfunctionApr 3, 2024Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE DEVICE CAUSED AN AIR INTRODUCTION TO THE SYSTEM. A WATCHDOG HEMOSTASIS VALVE KIT WAS SELECTED FOR USE. DURING THE PREPARATION, IT WAS NOTED THAT THE WATCHDOG VALVE FAILED WHEN ATTACHED TO BOSTON SCIENTIFIC CRYO BALLOON. THE IRRIGATION PORT OF THE VALVE CAME LOOSE WHICH CAUSING AIR INTRODUCTION TO THE SYSTEM. THIS DID NOT IMPACT THE PATIENT AS IT WAS NOT IN THE BODY AT THE TIME. NO PATIENT COMPLICATIONS WERE REPORTED POST PROCEDURE. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19034547 ↗InjuryFeb 23, 2024Loose or Intermittent Connection; Air/Gas in Device
BOSTON SCIENTIFIC WATCHDOG VALVE FAILED WHEN ATTACHED TO BOSTON SCIENTIFIC CRYO BALLOON. THE IRRIGATION PORT OF THE VALVE CAME LOOSE DURING PREP, CAUSING AIR INTRODUCTION TO THE SYSTEM. THIS DID NOT IMPACT THE PATIENT AS IT WAS NOT IN THE BODY AT THE TIME. THIS HAS HAPPENED IN TWO PREVIOUS CASES. THE PLASTIC THREADS WILL STRIP EASILY AND SHOULD NOT BE USED WITH THIS SYSTEM SET UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18782497 ↗MalfunctionJul 14, 2022Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT BLOOD LOSS OCCURRED. A WATCHDOG HEMOSTASIS VALVE WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE WATCHDOG WAS LEAKING AN EXCESSIVE AMOUNT OF BLOOD. THE TUBING WAS REPLACED THREE TIMES AND THE LEAKING WAS NOT RESOLVED. A NEW WATCHDOG WAS SELECTED AND THERE WAS NO LEAKING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 15019143 ↗
Other Boston Scientific Corporation products
- INGEVITY MRI7 models · 16,601 reports
- INGEVITY+3 models · 16,054 reports
- EMBLEM S-ICD4 models · 13,295 reports · 1 recall
- WATCHMAN25 models · 8,024 reports · 6 recalls
- AMS 800 Artificial Urinary Sphincter31 models · 7,515 reports · 2 recalls
- ENDOTAK RELIANCE SG9 models · 6,540 reports · 2 recalls
- RELIANCE 4-FRONT11 models · 5,739 reports
- FINELINE II EZ Sterox6 models · 5,633 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page