Skip to content

WATCHDOG recallZ-1593-2023 · Boston Scientific Corporation · initiated Apr 5, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.

FDA root cause category: Process control

What the company is doing

Boston Scientific issued an URGENT MEDICAL DEVICE PRODUCT REMOVAL notice on 04/05/2023 by overnight mail. The notice explained the issue and the hazard and requested the consignee segregate the devices pending return. Medical facilities who distributed the devices and distributors were directed to send the notice to their customers.

Recall details

Recall number
Z-1593-2023
Status
Open, Classified
Initiated by the company
Apr 5, 2023
Posted by the FDA
May 16, 2023
Recalling firm
Boston Scientific Corporation
Quantity
10833 units
Product code
DTL
FDA submissions
K172453
Distribution
Worldwide

Product description

WATCHDOG ACCESSORY KIT (20-PACK), UPN H74939343A021 (outer package), UPN H74939343A020 (inner package)

Code information (models, lots, serial numbers)

GTIN 08714729965770 (outer package), GTIN 08714729965794 (inner package), Batch numbers: 30319836,30359588,30368418,30392647,30409465,30421836, 30442123,30522919,30564343,30577971,30593836,30600837, 30600839,30601021,30616553,30620297,30625219,30638465, 30638467,30638469,30638471,30665570,30666996,30681966, 30681968,30681970,30712588,30725194,30780665,30801487, 30801489,30801491,30811488,30845828,30845830,30845832, 30885087,30890263,30890265,30914070,30922072,30929565, 30935495,30935497,30935499,30981820

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 92072).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page