WATCHDOG™ H74939343A020Recalls & reported problems

Boston Scientific Corporation · Hemostasis Valve Kit · Class 2

All models of WATCHDOG (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729965794), received between Feb 23, 2024 and Apr 3, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 5, 2023Open, Classified

    The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.

    Z-1593-2023 · Root cause: Process control · Recall details · FDA record ↗

  2. Apr 5, 2023Open, Classified

    The firm received complaints of the Insertion Tool component being occluded, resulting in the inability to pass a guidewire through the tool.

    Z-1592-2023 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Air/Gas in Device1
  • Device-Device Incompatibility1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE DEVICE CAUSED AN AIR INTRODUCTION TO THE SYSTEM. A WATCHDOG HEMOSTASIS VALVE KIT WAS SELECTED FOR USE. DURING THE PREPARATION, IT WAS NOTED THAT THE WATCHDOG VALVE FAILED WHEN ATTACHED TO BOSTON SCIENTIFIC CRYO BALLOON. THE IRRIGATION PORT OF THE VALVE CAME LOOSE WHICH CAUSING AIR INTRODUCTION TO THE SYSTEM. THIS DID NOT IMPACT THE PATIENT AS IT WAS NOT IN THE BODY AT THE TIME. NO PATIENT COMPLICATIONS WERE REPORTED POST PROCEDURE. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19034547 ↗
  • InjuryFeb 23, 2024Loose or Intermittent Connection; Air/Gas in Device

    BOSTON SCIENTIFIC WATCHDOG VALVE FAILED WHEN ATTACHED TO BOSTON SCIENTIFIC CRYO BALLOON. THE IRRIGATION PORT OF THE VALVE CAME LOOSE DURING PREP, CAUSING AIR INTRODUCTION TO THE SYSTEM. THIS DID NOT IMPACT THE PATIENT AS IT WAS NOT IN THE BODY AT THE TIME. THIS HAS HAPPENED IN TWO PREVIOUS CASES. THE PLASTIC THREADS WILL STRIP EASILY AND SHOULD NOT BE USED WITH THIS SYSTEM SET UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18782497 ↗

Other Wire, guide, catheter devices